Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100011

    Validate with FDA (Live)

    Date Cleared
    2010-10-08

    (277 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The F&P Bubble CPAP System is intended to provide CPAP to spontaneously breathing neonates and infants who require breathing support due to conditions associated with prematurity (such as Respiratory Distress Syndrome) or other conditions where CPAP is required or desired and is prescribed by a physician.

    The Bubble CPAP System is for use in the hospital clinical environment such as the NICU (Neonatal Intensive Care Unit) and PICU (Pediatric Intensive Care Unit).

    Device Description

    The Fisher & Paykel Healthcare (F&P) Bubble CPAP (continuous positive airway pressure) System provides respiratory support to spontaneously breathing infants. The F&P Bubble CPAP System delivers heated and humidified respiratory gas through an inspiratory breathing circuit to the infant via a nasal interface. An expiratory circuit connects to a water column threshold resistor which pressurizes the circuit. The F&P Bubble CPAP System consists of an in-line pressure relief valve, a humidification chamber, a heated breathing circuit, patient nasal interface and CPAP generator (water column threshold resistor). The System is intended to be operated at input gas flows of 4 - 15 L/min with available CPAP levels of 3 - 10 cmH2O.

    AI/ML Overview

    The Fisher & Paykel Healthcare (F&P) Bubble CPAP System provides respiratory support to spontaneously breathing infants.

    Here's an analysis of the provided information, focusing on acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly present a table of numerical "acceptance criteria" for the device's main function (CPAP delivery) or a direct comparison to specific threshold values. Instead, it relies on demonstrating substantial equivalence to a predicate device.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial Equivalence in CPAP Delivery (Bench Test)Bench tests show that the F&P Bubble CPAP System's water column threshold resistor and the predicate Dräger device's electronically regulated valve provide substantially equivalent CPAP. The CPAP level delivered to the patient by the F&P system is "very close to the intended pressure set at the CPAP generator."
    Substantial Equivalence in Work of Breathing (Bench Test)The work of breathing of the F&P Bubble CPAP System is "significantly less than the predicate device in all measurements." While not directly "equivalent," being better than the predicate device supports substantial equivalence in functionality.
    Compliance with ISO 8185:2007 (E) (Respiratory Humidifiers)The single heated breathing circuit meets requirements for Surface Temperature, Humidification System Output, and Protection Against Hazardous Output.
    Compliance with ISO 5367:2000 (E) (Breathing Tubes)The breathing circuit meets requirements for Occlusion Resistance.
    Biocompatibility (ISO 10993-1)Assessment and testing show that all materials contacting the patient directly or indirectly are suitable for patient contact.
    Stability of CPAP over timeThe auto-leveling mechanism of the CPAP generator compensates for accumulated condensate and allows the mean CPAP level to remain stable over time.
    Pressure Oscillations EffectGas bubbling from the water threshold resistor creates pressure oscillations, but these "do not adversely affect the performance or safety of the device." This is confirmed by clinical evidence.
    Clinical Efficacy and SafetyEight peer-reviewed and published reports (including three randomized controlled trials) using the F&P Bubble CPAP System demonstrate it has been "widely and successfully used around the world." Comparisons with other forms of respiratory support have "positive outcomes towards the F&P Bubble CPAP System."

    Study Details Proving Acceptance Criteria:

    The submission outlines a combination of bench testing and a summary of existing clinical literature to establish substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Bench Tests: The specific sample sizes for the various bench tests are not explicitly stated in the provided summary.
    • Clinical Performance: "Eight peer reviewed and published reports using the F&P Bubble CPAP System, including three randomized controlled trials."
      • Data Provenance: "Used around the world" and "compared with other forms of respiratory support in a variety of settings." These are retrospective studies (already published). The countries of origin are not specified beyond "around the world."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Bench Tests: The summary does not specify the number or qualifications of experts involved in establishing ground truth for the bench tests. Ground truth would likely be based on established engineering principles and measurement standards.
    • Clinical Performance: For the eight peer-reviewed and published reports, the "ground truth" (i.e., clinical outcomes, efficacy, safety) would have been established by the clinicians and researchers involved in those studies. The number and specific qualifications of these experts are not detailed in this 510(k) summary. These would be implicit in the peer-reviewed publications.

    4. Adjudication Method for the Test Set

    • Bench Tests: No adjudication method is described. Bench tests typically rely on direct measurements against predefined standards.
    • Clinical Performance: For the summarized clinical reports, the adjudication method would have been specific to each individual study and subject to the peer-review process of the journal in which they were published. This summary does not detail these methods (e.g., 2+1, 3+1).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not performed or referenced. The device is a medical device (CPAP system), not an imaging or diagnostic AI tool that would typically involve human readers interpreting cases. The comparison is between the F&P Bubble CPAP System and a predicate CPAP device, primarily through bench testing of physical performance and a review of existing clinical literature on the F&P system.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    • The F&P Bubble CPAP System is a medical device that delivers a therapy, not an algorithm in the sense of a software-only tool. Its "standalone performance" is understood through its function of generating and delivering CPAP. The bench tests described (CPAP level delivered, work of breathing, compliance with ISO standards) directly assess its standalone physical performance.
    • There isn't an "algorithm only" component in the context of typical AI/software medical devices. The device's operation is its standalone performance.

    7. Type of Ground Truth Used

    • Bench Tests: Engineering measurements, adherence to ISO standards, and comparison to the predicate device's measured performance.
    • Clinical Performance: Clinical outcomes, safety data, and efficacy measures reported in peer-reviewed medical literature.

    8. Sample Size for the Training Set

    • The F&P Bubble CPAP System is a physical medical device, not a machine learning model. Therefore, there is no "training set" in the context of artificial intelligence or machine learning. The device's design is based on engineering principles and knowledge, validated through testing.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no training set for an AI/ML model, this question is not applicable. The "ground truth" for the device's development would be established engineering specifications, medical requirements for CPAP delivery, and safety standards, which are met through design and validation testing.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1