Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093755

    Validate with FDA (Live)

    Date Cleared
    2011-07-15

    (585 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AG-695 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample from the fingertip.. The AG-695 Single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The AG-695 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-695 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use.

    This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a qlucose test.

    The AGS-1000 single Blood Glucose Test Strips are for use with the AG-695 Single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample.

    The AG-695 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample from the fingertip.. The AG-695 MULT! Blood Glucose Monitoring System is intended for testing outside the body ( in vitro diagnostic use) and is intended to be used by healthcare professionals for multiple patients in a professional healthcare setting as an aid in monitoring the effectiveness of diabetes control.

    The AG-695 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use.

    This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a alucose test.

    The AGS-1000 MULTI Blood Glucose Test Strips are for use with the AG-695 MULTI Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample.

    This system should only be used with single-use, auto-disabling lancing devices.

    The AG-696 Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of qlucose (sugar) in fresh capillary whole blood sample from the fingertip. The AG-696 Single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The AG-696 Single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-696 Single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use.

    This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

    The AGS-1000 single Blood Glucose Test Strips are for use with the AG-696 Single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample.

    The AG-696 MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample from the fingertip. The AG-696 MULTI Blood Glucose Monitoring System is intended for testing outside the body ( in vitro diagnostic use) and is intended to be used by healthcare professionals for multiple patients in a professional healthcare setting as an aid in monitoring the effectiveness of diabetes control.

    The AG-696 MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use.

    This system contains a speaking function that provides audible test results for users with impaired vision. The audible function does not provide complete instructions for all functions of the meter or for performing a glucose test.

    The AGS-1000 MULTI Blood Glucose Test Strips are for use with the AG-696 MULT! Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood sample.

    This system should only be used with single-use, auto-disabling lancing devices.

    Device Description

    The four blood Glucose Monitoring system AG-695 Single, AG-695 MULTI, AG-696 Single and AG-696 MILTI are all based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds. The control solution available is used to test the performance of the device. All of them use the same technological characteristics for testing but the appearance is different from their predicate device. More over, all of the four systems have a voice function.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the AG-695/AG-696 Blood Glucose Monitoring Systems:

    The provided document is a 510(k) Summary for a medical device, which typically focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report of accuracy, sensitivity, and specificity in the way one might for a diagnostic imaging AI. For blood glucose monitoring systems (BGMS), the primary acceptance criteria revolve around the accuracy of glucose measurements compared to a laboratory reference method, as defined by standards like ISO 15197.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document references compliance with ISO 15197 and FDA Draft Guidance Document "Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems: October 24, 2006." These standards set forth the acceptance criteria for accuracy. While the document mentions compliance, it does not explicitly present a table of acceptance criteria or detailed device performance statistics (e.g., mean absolute relative difference, percentage of results within specific error margins) in the provided text.

    However, based on typical ISO 15197 (2003) guidelines (the likely version for a 2011 submission), common acceptance criteria for accuracy are:

    Acceptance Criteria (Typical ISO 15197:2003)Reported Device Performance
    For glucose concentrations < 75 mg/dL (4.2 mmol/L): All results within ±15 mg/dL (±0.83 mmol/L) of the reference method.Not explicitly detailed in the provided text. The document states the device "conform to ... ISO 15197".
    For glucose concentrations ≥ 75 mg/dL (4.2 mmol/L): All results within ±20% of the reference method.Not explicitly detailed in the provided text. The document states the device "conform to ... ISO 15197".
    All glucose concentrations: At least 95% of results within specific error margins (e.g., ±15 mg/dL or ±20%).Not explicitly detailed in the provided text. The document states the device "conform to ... ISO 15197".

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for the Test Set: Not explicitly stated in the provided text. The document only mentions "the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness."
    • Data Provenance: Not explicitly stated. Given that Andon Medical Co., Ltd. is based in Tianjin, China, and the application date is 2009, it's highly probable the testing was conducted in China, though this is not confirmed. The document does not specify if the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable and not typically found in 510(k) submissions for blood glucose monitoring systems. The "ground truth" for these devices is established by a laboratory reference method, not by expert consensus (like radiologists for imaging studies).

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used when multiple human readers interpret data to establish a ground truth or resolve discrepancies. For blood glucose monitors, the reference method provides the objective truth, and there is no human adjudication process involved in comparing device readings to the reference.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a standalone blood glucose monitoring system, not an AI-powered diagnostic imaging tool or an AI assistant for human readers. Therefore, MRMC studies and "human reader improvement with AI" are not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies would have been conducted as a standalone performance assessment of the device (meter and test strips) against a laboratory reference method. The device's electrochemical biosensor technology and capillary action principle are essentially the "algorithm" that converts the electrochemical signal into a glucose reading. The performance assessment outlined by ISO 15197 and FDA guidance focuses on this standalone accuracy without human interpretation influencing the numerical result.

    7. The Type of Ground Truth Used

    The ground truth used for blood glucose monitoring systems is typically established by a laboratory reference method for glucose measurement (e.g., hexokinase method, glucose oxidase method) performed on venous plasma or serum samples. This is considered the gold standard for glucose concentration.

    8. The Sample Size for the Training Set

    Not applicable in the context of this device and report. This is a traditional electrochemical biosensor device, not a machine learning or AI algorithm that requires a "training set" in the computational sense. The device's "training" and calibration would involve manufacturing processes to ensure sensor consistency and calibration against known glucose concentrations.

    9. How the Ground Truth for the Training Set was Established

    Not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1