    K Number
    K093677

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-03-05

    (98 days)

    Product Code
    Regulation Number
    888.1100
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stryker Arthroscopes are endoscopic devices introduced into a patient to provide an internal view or image of the interior of a joint for examination, diagnosis, and/or therapy. Arthroscopes are indicated for use in arthroscopic procedures performed in the hip, knee, shoulder, wrist (carpel tunnel syndrome), temporal mandibular joint, ankle, elbow, and feet (plantar fascia release).

    Device Description

    Stryker Arthroscopes are endoscopic devices introduced into a patient to provide an internal view or image of the interior of a joint for examination, diagnosis, and/or therapy. The Stryker Arthroscope is a long tube containing a series of lenses. At the distal end, an objective lens captures the image of the object. Lens along the rod relay the image. At the proximal end, a proximal coupling lens relays the image to a CCD (Camera).

    The Stryker Arthroscopes come in various diameters including 1.9mm, 2.7mm, and 4.0mm. Larger size arthroscopes are used for general viewing, while smaller diameter arthroscopes are used for restricted surgical sites. The Stryker Arthroscopes come in several directions of view including, 0°, 30°,45°, 70°. The direction of view enables viewing of different parts. Materials of the Arthroscope include stainless steel, titanium, PEEK, Glass, and Sapphire.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Stryker Arthroscope, focusing on establishing substantial equivalence to a predicate device rather than presenting a study to prove the device meets specific performance acceptance criteria based on clinical outcomes or diagnostic accuracy.

    Therefore, many of the requested details about acceptance criteria, study design, and ground truth are not applicable or cannot be extracted from this document, as it concerns a medical device (an arthroscope) that's primarily a viewing instrument for internal examination. The "performance testing" mentioned refers to engineering and safety standards, not clinical performance in terms of diagnostic accuracy or reader improvement.

    Here's a breakdown of what can be extracted based on the provided text, and what cannot:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not specify quantitative "acceptance criteria" in terms of clinical performance metrics (e.g., sensitivity, specificity, AUC) or diagnostic accuracy. Instead, it demonstrates compliance with voluntary safety and performance standards and highlights technological characteristics compared to a predicate device.

    CategoryAcceptance Criteria (Implied by Standards & Equivalence)Reported Device Performance
    BiocompatibilityConformity to ISO 10993-1 (Evaluation and Testing), 10993-10 (Irritation and Delayed-Type Hypersensitivity). Implied acceptance: non-toxic, non-irritating materials.Subjected to and passed biocompatibility testing requirements. Patient-contacting materials are identical to predicate device (Henke Sass Wolf Arthroscope K080560).
    Electrical SafetyConformity to IEC 60601-1 (General Requirements for Basic Safety and Essential Performance), IEC 60601-2-18 (Safety of endoscopic equipment). Implied acceptance: safe electrical operation.Subjected to and passed electrical safety testing requirements.
    SterilizationConformity to TIR 12 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing) and ISO 14937 (Sterilization of Health Care Products - General Requirements). Implied acceptance: device can be safely and effectively sterilized for reuse.Subjected to and passed sterilization testing requirements.
    Optical PerformanceConformity to ISO 8600-1 (General requirements for medical endoscopes), 8600-3 (Determination of field of view and direction of view), and 8600-5 (Determination of Optical Resolution). Implied acceptance: meets specified optical properties for visualization.Met all specified design and performance requirements. Specific FOV values are 105°, 80°, 65°. Specific direction of view values are 0°, 30°, 45°, 70°. Outer Diameters: 4mm, 2.7mm, 2.3mm, 1.9mm. Working Lengths: 165mm, 140mm, 120mm, 75mm, 72mm, 58mm.
    Overall DesignSubstantial equivalence to predicate device (Henke Sass Wolf Arthroscopes K080560) in construction, materials, and technological characteristics (FOV, Direction of View, Outer Diameter, Working Length, Reusability, Light Guide End Adapter) such that no new questions of safety or effectiveness are raised.The Stryker Arthroscopes are substantially equivalent to the predicate device. Differences in Field of View and Working Length do not affect safety and efficacy, as they are within acceptable ranges or do not pose new risks.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable. The document discusses performance testing related to engineering standards (biocompatibility, electrical safety, sterilization, optics) and a comparison of technological characteristics to a predicate device. It does not describe a clinical "test set" in the context of diagnostic accuracy or clinical outcomes.
    • Data Provenance: Not applicable for clinical performance data. The data presented is from internal engineering tests and a comparison to a predicate device's specifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. This type of information is relevant for studies involving human interpretation or clinical judgment. The 510(k) pertains to the physical and functional characteristics of an arthroscope, not its interpretative output.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There is no "test set" of cases requiring adjudication by experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • MRMC Study: No, an MRMC study was not done. The device is an arthroscope, a visually guided surgical instrument, not an AI-powered diagnostic or interpretive tool.
    • Effect Size of Human Reader Improvement: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance Study: No, a standalone performance study in the context of an algorithm's diagnostic accuracy was not done. The Stryker Arthroscope is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Not applicable in the context of clinical outcomes or diagnostic accuracy. For the engineering performance tests, the "ground truth" would be defined by the specifications in the relevant ISO and IEC standards (e.g., a material is biocompatible if it passes ISO 10993 tests; an arthroscope meets optical resolution if it conforms to ISO 8600-5). The substantial equivalence claim is based on direct comparison of design and material specifications between the proposed device and the predicate device.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. There is no "training set" as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable.