Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093603

    Validate with FDA (Live)

    Device Name
    STORM-BASE 2.0
    Date Cleared
    2010-03-18

    (118 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StorM-Base 2.0 is intended to be used by physicians and other qualified medical personnel for the documentation, communication, display, printing and storage of medical diagnostic data in the field of Lithotripsy. When installed with a Storm-Touch system, it is also an image archiving system based on the DICOM standard capable of capturing and annotating lithotripsy procedures for documentation purposes. Images imported and distributed by StorM-Base 2.0 are for viewing and reference purposes only and are not intended for primary diagnosis.

    Device Description

    The Strorz StorM-Base 2.0 is a Windows XP based software application that stores patient data, examination results including the number, type and location of stones, treatment images and data. The application may be installed on a standalone PC or on the StroM-Touch MODULITH Lithotripter (K070579). When installed on the StroM-Touch MODULITH Lithotripter, the system can acquire the images and treatment data directly from the lithotripter via an import function and be networked to the Storz Controller Buss.

    AI/ML Overview

    The Karl Storz Endoscopy StorM-Base 2.0 is a software application intended for documentation, communication, display, printing, and storage of medical diagnostic data in lithotripsy. It is explicitly stated that "Images imported and distributed by StorM-Base 2.0 are for viewing and reference purposes only and are not intended for primary diagnosis."

    Given this intended use, the submission primarily focuses on software verification and validation, compliance with standards (DICOM), and comparison to a predicate device regarding similar technical characteristics and safety. There are no clinical performance acceptance criteria or studies involving human readers or ground truth for diagnostic accuracy, as the device is not intended for primary diagnosis.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance related to diagnostic accuracy, nor can I answer questions about sample sizes for test sets, expert-established ground truth, adjudication methods, MRMC studies, or standalone performance. The document only addresses the software's functional and safety aspects.

    Here's a summary of the information available based on the provided text, primarily focusing on the software's development and validation:

    1. Table of Acceptance Criteria and Reported Device Performance (Operational and Safety Focus):

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with software verification and validation methodsDeveloped in compliance with software verification and validation methods.
    Implementation of appropriate risk control measuresAppropriate risk control measures identified and implemented through a risk management process meeting ISO 14971.
    Proper implementation of requirementsVerification and validation testing performed to ensure requirements are properly implemented.
    Compliance with DICOM standardVerified to be in compliance with the DICOM standard.
    System functionality (storage, display, printing, etc.)Technical characteristics similar to predicate (Windows-based, stores patient info, complies with HL7/DICOM, archives data, imports/distributes images).
    No manipulation of original PACS or HIS dataDoes not manipulate original PACS or HIS data.
    Images not for diagnostic useStored images are not intended for diagnostic use.

    Since the device is not intended for primary diagnosis, the following points address what is NOT applicable or available in the provided text:

    1. Sample size used for the test set and the data provenance: Not applicable as no clinical diagnostic performance study was conducted or reported. The "test set" would refer to software validation tests, not clinical data sets for diagnostic accuracy.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as there is no ground truth for diagnostic accuracy established for this device's intended use.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is not an AI for diagnostic assistance.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable for diagnostic performance. The device is a data management system.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for diagnostic accuracy. Ground truth for software validation would be based on functional requirements.
    7. The sample size for the training set: Not applicable as this is not a machine learning or AI device requiring a "training set" in the context of diagnostic data.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1