Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093458

    Validate with FDA (Live)

    Date Cleared
    2009-12-04

    (28 days)

    Product Code
    Regulation Number
    862.1225
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CR-E reagent, when used in conjunction with SYNCHRON® Systems, UniCel® DxC Systems and SYNCHRON® Systems AQUA CAL 1 and 2 and SYNCHRON CX® Calibrator Level 1 and 2, is intended for the quantitative determination of creatinine (CR-E) concentration in human serum, plasma or urine (urine is not available on the SYNCHRON CX PRO Systems).

    Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

    Device Description

    CR-E reagent is used to measure the creatinine concentration by an enzymatic method. This enzymatic creatinine method utilizes a multi-step approach ending with a photometric end-point reaction.

    The SYNCHRON Enzymatic Creatinine (CR-E) Reagent is designed for optimal performance on the SYNCHRON LX, UniCel® DxC 600/800, and SYNCHRON CX® PRO Clinical Systems. The reagent kit contains two 200-test cartridges that are packaged separately from the associated calibrator.

    AI/ML Overview

    Here's an analysis of the provided text regarding the SYNCHRON® Systems Enzymatic Creatinine (CR-E) Reagent, structured according to your request:

    Acceptance Criteria and Device Performance Study for SYNCHRON® Systems Enzymatic Creatinine (CR-E) Reagent

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Analyte: CreatinineAnalyte: Creatinine
    Matrix (Serum): Limit of Detection (Analytical Sensitivity) = 0.1 mg/dL (Extended from 0.2 mg/dL)Matrix (Serum): Limit of Detection (Analytical Sensitivity) = 0.1 mg/dL
    Matrix (Urine): Limit of Detection (Analytical Sensitivity) = 10 mg/dL (Unchanged)Matrix (Urine): Limit of Detection (Analytical Sensitivity) = 10 mg/dL
    Intended Use: Quantitative determination of creatinine in human serum, plasma, or urine for diagnosis/treatment of renal diseases, monitoring renal dialysis, and as a calculation basis for other urine analytes.Intended Use: Device performs as intended for quantitative determination of creatinine in human serum, plasma, or urine for diagnosis/treatment of renal diseases, monitoring renal dialysis, and as a calculation basis for other urine analytes. (Implied by substantial equivalence and performance data supporting equivalency).

    Note: The primary acceptance criterion highlighted in the document is the successful extension of the Limit of Detection (Analytical Sensitivity) for serum samples to 0.1 mg/dL, replacing the previous 0.2 mg/dL. The urine sample LoD remained unchanged and met the existing criterion. The overall performance is implicitly tied to demonstrating "substantial equivalence" to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Performance data from validation testing supports equivalency." However, it does not explicitly provide the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data). This information is typically found in the detailed validation reports submitted to the FDA, but is not included in this summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used or their qualifications to establish ground truth. For in vitro diagnostic devices like this, ground truth is typically established by comparing the device's results to a recognized reference method or to results from a legally marketed predicate device, rather than expert consensus on individual cases.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for a test set. This type of method is usually applied in studies involving subjective interpretation (e.g., imaging studies) rather than quantitative chemical assays. For this device, the "ground truth" would be the result from a reference method or the predicate device, and the comparison would be statistical.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices where human readers interpret observations (e.g., radiology AI). The SYNCHRON® Systems Enzymatic Creatinine Reagent is an automated in vitro diagnostic device, where the measurement is quantitative and performed by the instrument.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone study was done. The SYNCHRON® Systems Enzymatic Creatinine (CR-E) Reagent is an automated assay run on the SYNCHRON LX, UniCel® DxC 600/800, and SYNCHRON CX® PRO Clinical Systems. The "performance data from validation testing" that "supports equivalency" would inherently be a standalone evaluation of the reagent's analytical performance on these instruments. There is no human interpretation component where the results are subsequently re-evaluated by a human to influence the device's performance.

    7. Type of Ground Truth Used

    The type of ground truth used is comparison to a legally marketed predicate device (K091742) and, implicitly, to established analytical methods for creatinine measurement. The validation testing would involve comparing the CR-E reagent's results against a gold standard or a well-characterized reference method, likely by linearity, precision, and accuracy studies, or by demonstrating concordance with the predicate device. The primary change here is an extended Limit of Detection for serum, which would have been validated against samples with known low creatinine concentrations.

    8. Sample Size for the Training Set

    The document does not specify the sample size for the training set. For an in vitro diagnostic reagent, there isn't typically a "training set" in the machine learning sense. Instead, the reagent's formulation and assay parameters are developed and optimized through extensive R&D and early validation studies using various concentrations of known analytes and patient samples. The "training" is part of the product development and optimization process, not a distinct "training set" as understood in AI/ML validation studies.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, the concept of a "training set ground truth" isn't directly applicable in the same way it would be for an AI/ML algorithm. However, throughout the development and optimization of the reagent, the "ground truth" for determining assay parameters and performance would have been established by:

    • Reference Methods: Using highly accurate and precise reference methods (e.g., isotope dilution mass spectrometry) to determine the true creatinine concentration in control materials and patient samples.
    • Known Concentrations: Preparing solutions with precisely known concentrations of creatinine.
    • Comparison to Predicate: Benchmarking performance against the legally marketed predicate device (Beckman Coulter SYNCHRON Systems Enzymatic Creatinine (CR-E) reagent, K091742) during development and optimization.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1