510(k) Data Aggregation

    K Number
    K093406

    Validate with FDA (Live)

    Date Cleared
    2010-03-25

    (143 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AngioDynamics, Inc. Morpheus® SMART PICC CT and Procedure Kit is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injections of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapy use a 4 French or larger catheter. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

    Device Description

    The AngioDynamics 6F Triple Lumen Morpheus® SMART PICC CT and Procedure Kits device description is as follows:
    • 6 French outside diameter, reversed tapered shaft design
    • Catheter usable length is 55 cm
    • Shaft inner lumen is a triple lumen design
    • Catheter shaft tubing is marked with depth indicators
    • The catheter has one power injectable lumen
    • The product labeling warns against power injection procedures through the two small lumens, which are clearly identified as non CT.
    • Three extension legs to facilitate injection through each lumen of the catheter shaft.

    AI/ML Overview

    The AngioDynamics, Inc. Morpheus® SMART PICC CT and Procedure Kit is a medical device. The provided text describes the device, its intended use, and the verification activities undertaken to demonstrate its safety and effectiveness, primarily through comparison to predicate devices.

    Here's the detailed breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestProposed Device Acceptance CriteriaProposed 6F Triple Catheter (K093406) T=0 (Initial)Proposed 6F Triple Catheter (K093406) T=1 (After Accelerated Aging/Sterilization)
    Tip WhipTip "whipping" must be less than or equal to predicate devicesPassPass
    Tip DisplacementTip displacement must be less than or equal to predicate devicesPassPass
    CT Injection - Catheter Flow RateFlow Rate = 5mL/sec ± 0.5mL/sec with 95% confidence that 95% of the population meets specification(Value not explicitly stated, but assumed to pass as no failure indicated)(Value not explicitly stated, but assumed to pass as no failure indicated)
    CT Injection - CT Lumen IntegrityCT Lumen must withstand a minimum of 10 injections at a minimum flow rate of 5mL/sec with a 95% confidence that 80% of the population meets specificationPassPass
    AspirationMinimum aspiration rate to be 3cc/min using a 10cc syringe without total collapse (all three catheter lumens) with 95% confidence that 95% of the population meets specificationPassPass
    Gravity Flow RateMinimum gravity flow rate to be 750 ml/hr for CT lumen and 182 ml/hr for non-CT lumens with 95% confidence that 95% of the population meets specificationPassPass
    Stylet Withdrawal TestingStylet withdrawal force must be less than 2 lbs with 95% confidence that 95% of the population meets specificationPassPass
    Wire Withdrawal TestingGuidewire withdrawal force must be less than 2 lbs with 95% confidence that 95% of the population meets specificationPassPass
    Static BurstCatheter burst pressure must be greater than 150 psi with 95% confidence that 95% of the population meets the specificationPassPass
    Dynamic BurstCatheter dynamic burst pressure must be greater than 300 psi with 95% confidence that 95% of the population meets specificationPassPass
    Extension Leg to Natural Color Hub - Tensile TestingTensile strength must be greater than 5 lbs with 95% confidence that 99% of the population meets the specificationPassPass
    Non-CT Extension Leg to Trifurcate - Tensile TestingTensile strength must be greater than 5 lbs with 95% confidence that 99% of the population meets the specificationPassPass
    CT Extension Leg to Trifurcate - Tensile TestingTensile strength must be greater than 5 lbs with 95% confidence that 99% of the population meets the specificationPassPass
    Shaft to Trifurcate - Tensile TestingTensile strength must be greater than 5 lbs with 95% confidence that 99% of the population meets the specificationPassPass
    Catheter Shaft (at nominal OD) - Tensile TestingTensile strength must be greater than 5 lbs with 95% confidence that 99% of the population meets the specificationPassPass
    Pressure MonitoringNo difference observed between SVS and catheter in mean pressure, systolic pressure and diastolic pressure as compared to predicate devicesPassPass
    Natural FrequencyThe natural frequency must be equal to or greater than the predicate devicePassPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific number of devices (sample size) used for each individual test. Instead, acceptance criteria are generally expressed with statistical confidence levels (e.g., "95% confidence that 95% of the population meets specification" or "95% confidence that 80% of the population meets specification"), implying that a sufficient sample size was used to achieve these statistical guarantees.

    The data provenance is not specified regarding country of origin or specific details of test conditions beyond the type of tests. The studies are described as "verification/validation activities completed on the proposed device," which are typically laboratory-based engineering and performance tests rather than clinical trials with patient data. Thus, the data would be prospective in the sense that the tests were specifically designed and executed for this device submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to the studies described. The "ground truth" for these tests (e.g., a catheter's flow rate, tensile strength, or burst pressure) is determined by objective measurements and engineering standards, not by expert consensus or interpretation of clinical findings.

    4. Adjudication Method for the Test Set

    This is not applicable. The tests are objective measurements against defined acceptance criteria, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on the engineering and performance characteristics of the device, primarily comparing it to predicate devices based on objective technical specifications. It's a physical medical device, not an AI or imaging diagnostic tool that would typically involve MRMC studies.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    This is not applicable. The device is a physical medical catheter, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on engineering specifications and objective measurements of the device's physical and functional properties, and comparison to the performance of legally marketed predicate devices. This includes:

    • Physical durability (tensile strength, burst pressure)
    • Flow rates (CT injection, gravity flow, aspiration)
    • Mechanical behavior (tip whipping, tip displacement, stylet/wire withdrawal force)
    • Biocompatibility (tested per ISO 10993)

    8. The Sample Size for the Training Set

    This is not applicable. The studies described are performance and safety verification tests for a physical medical device, not a machine learning model that would require a training set.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set for a physical medical device.