Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K092387

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-05

    (243 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medivators DSD Edge Endoscope Reprocessing System tests, washes, disinfects and rinses flexible endoscopes, such as fiberoptic and video endoscopes between patient uses. The DSD Edge system is indicated to provide high level disinfection, using Rapicide PA High Level Disinfectant, of heat sensitive semi-critical endoscopes. Manual cleaning of endoscopes is required prior to placement in the DSD Edge system.

    Rapicide PA contact conditions in the DSD Edge

    • 5 Minutes 30℃ 850 ppm peracetic acid .
    Device Description

    The DSD Edge system is an electro-mechanical system intended to test, wash and high level disinfect flexible fiberoptic and video endoscopes between uses. The DSD Edge is capable of asynchronously reprocessing two scopes at a time.

    The DSD Edge system uses the peracetic acid based Rapicide PA High Level Disinfectant (K082988). The DSD Edge is a single use system in that it mixes Part A and Part B of Rapicide PA with water immediately prior to reprocessing and the disinfectant is not reused.

    Endoscopes must be pre-cleaned and manually cleaned to SGNA and facility guidelines prior to being placed in the system for reprocessing.

    After the endoscopes are connected to the DSD Edge the system tests the scopes for leaks in the outer skin. If the endoscopes pass the leak test the system will proceeds to a scope flush or an optional wash cycle followed by a flush.

    For the disinfection cycle, the incoming water is mixed with the two part germicide in the basin. The temperature of the incoming water is monitored to ensure that the water temperature is within the operating constraints (30℃) required for disinfection after the water and germicide are mixed together. A sample of the germicide is retained to test for MRC. Following disinfection, the endoscopes are rinsed and dried by the machine, either by filtered air or an optional alcohol rinse, and are then removed from the machine for the next use.

    The machine has built in safety features which stop the cycle and alarm when certain conditions exist which could indicate that disinfection might be compromised. These alarms and causes are defined in the directions for use for the product.

    The DSD Edge also prints records indicating the results of testing, disinfection, etc. which are required for permanent records.

    AI/ML Overview

    The document describes the Medivators DSD Edge Endoscope Reprocessing System, an electro-mechanical system designed to test, wash, and high-level disinfect flexible endoscopes. The system uses Rapicide PA High Level Disinfectant. The following information details the acceptance criteria and the study that proves the device meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria for performance; instead, it indicates that the device must "perform as required" for various functions. The reported device performance is based on successful completion of these tests.

    Acceptance Criteria (Implicit)Reported Device Performance
    Material compatibility with disinfectantSystem materials showed no significant deterioration over time; filters were compatible with disinfectant.
    Biocompatibility (disinfectant residue)Any remaining disinfectant residues would not have an effect on patients or users.
    Proper functioning of leak checkDevice performs leak checks correctly.
    Proper functioning of washing cycleDevice performs washing cycles correctly.
    Proper functioning of disinfection cycleDevice performs disinfection cycles correctly.
    Proper functioning of rinse cyclesDevice performs rinse cycles correctly.
    Proper functioning of drying cyclesDevice performs drying cycles correctly.
    Proper functioning of error messages/alarmsError messages were tested and function properly to notify users of failure modes.
    Disinfectant temperature maintenanceDisinfectant remained at its required temperature (30℃) for the required length of time (5 minutes).
    Proper functioning of self-disinfection cycleSelf-disinfection cycle works properly, disinfecting all areas including the water filtration system.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a quantitative sample size for the test set (e.g., number of endoscopes, number of cycles). The testing appears to be conducted by the manufacturer, Medivators, and the data provenance is not explicitly stated as retrospective or prospective, nor does it specify country of origin beyond "Medivators has provided testing." Typically, such premarket submissions involve prospective testing conducted at the manufacturer's facilities or a designated testing laboratory.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of device (endoscope reprocessor) does not typically involve human expert assessment for establishing "ground truth" in the same way an AI diagnostic tool would. Instead, performance is validated against established engineering and biological standards (e.g., microbial reduction, temperature control, leak detection accuracy). Therefore, the concept of "experts establishing ground truth" in the context of radiologists or similar medical professionals does not apply here. The "experts" would be the engineers and microbiologists conducting the tests and verifying the device's functionality against pre-defined specifications. The document does not specify the number or qualifications of these internal experts.

    4. Adjudication Method for the Test Set

    Adjudication methods like 2+1 or 3+1 consensus are not applicable for the type of testing described (material compatibility, functional performance, residue levels). The validation relies on objective measurements and compliance with technical specifications and safety standards defined in FDA guidance documents (e.g., FDA's Guidance on Premarket [510(k)] Submissions for Automated Endoscope Washers, Washer/Disinfectors, and Disinfectors Intended for Use in Health Care Facilities).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC comparative effectiveness study is not relevant for this device. This study type is typically used for diagnostic or screening tools where a human reader's interpretation is central to the outcome and performance needs to be measured with and without AI assistance across multiple cases and readers. The Medivators DSD Edge is an automated reprocessing system, not a diagnostic tool requiring human interpretation of medical images or data.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The device itself is a standalone automated system. Its performance metrics (e.g., effective disinfection, leak detection, proper rinsing) are evaluated independently of a human operator making real-time decisions that affect the disinfection process. While manual pre-cleaning is required, the "reprocessing" cycle within the DSD Edge is fully automated. Therefore, the performance described reflects the standalone capabilities of the device in performing its indicated functions.

    7. The Type of Ground Truth Used

    The "ground truth" for verifying the performance of the DSD Edge Reprocessing System is based on:

    • Engineering Specifications: Ensuring mechanical and electrical components (e.g., pumps, sensors, heating elements) function as designed.
    • Chemical/Physical Measurements: Verifying parameters like disinfectant concentration (850 ppm peracetic acid), temperature (30℃), contact time (5 minutes), and residue levels.
    • Microbiological Standards: Although not explicitly detailed in this summary, the "high-level disinfection" claim inherently relies on demonstrating the device's ability to achieve a specified reduction in microbial load (e.g., vegetative bacteria, mycobacteria, fungi, viruses) under the specified contact conditions. This is typically proven through standardized inactivation studies.
    • Material Science: Verifying material compatibility to ensure integrity and longevity of the device and associated filters.

    8. The Sample Size for the Training Set

    This document describes a medical device (an endoscope reprocessor), not an AI/machine learning algorithm. Therefore, the concept of a "training set" for an algorithm to learn from does not apply. The device's operation is based on pre-programmed logic, physical and chemical principles, and validated engineering design, not on a machine learning model trained on a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, there is no "training set" in the context of machine learning for this device. The 'ground truth' for the design and validation of this electro-mechanical system is established through engineering design principles, recognized industry standards, scientific understanding of disinfection processes, and regulatory guidance documents.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1