Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K091928

    Validate with FDA (Live)

    Device Name
    10 X 10 VISION
    Manufacturer
    Date Cleared
    2010-06-16

    (351 days)

    Product Code
    Regulation Number
    892.1680
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Bioptics 10 X 10 VISION is a digital specimen radiography add-on upgrade to previously released cabinet x-ray systems. The 10 X 10 Vision is used for digital specimen radiography to provide rapid verification that the correct tissue has been excised during excisional or percutaneous biopsy. The 10 X 10 Vision is excluded from any use in in-vivo diagnostic or screening mammography applications.

    Device Description

    The 10 X 10 VISION will be marketed as a digital specimen radiography add-on upgrade for standard film-based cabinet x-ray systems. using high-performance, flat panel CMOS detector, and high-resolution Gd202S:Tb (Gadolinium Oxysulfide) scintillator technologies. Bioptics digital x-ray imagers are packaged in a thin (<14 mm) form factor similar to standard film cassette, include power supply, cables and needed mechanical accessories for convenient and cost-efficient upgrade of existing film-based cabinet x-ray systems. The x-ray imager assembly is installed permanently in the detector/cassette housing compartment or at the base of the host cabinet x-ray unit. The passive x-ray detection technology does not emit radiation and its low power (<5W) consumption does not require additional cooling. The 10 X 10 VISION employs the use of Bioptics Vision image acquisition software. The Bioptics Vision software handles digital x-ray image acquisition, calibration, image display, image analysis and manipulation, patient database, image archiving, and transmittal. Bioptics Vision software is the central part of this system. Bioptics Vision software is compliant with Digital Imaging and Communications in Medicine (DICOM) 3.0, and comes with DICOM Print, Store and Modality Work List (MWL).

    AI/ML Overview

    The submission describes a digital specimen radiography add-on upgrade called "10 X 10 VISION" intended for use with existing film-based cabinet x-ray systems. The device provides digital x-ray imaging for rapid verification of excised tissue during biopsy procedures.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Explicitly Stated or Implied)Reported Device Performance
    Compliance with 21 CFR 1020.30 & 1020.40Validated that use of 10 X 10 VISION will not affect compliance with 21 CFR 1020.31 upon installation. Tests performed to assure radiation emissions and leakage current are in compliance.
    High sensitivity, high resolution, dynamic rangeGd2O2S(Tb) scintillator and 50 micron pixels provide high sensitivity with high resolution and dynamic range. Resolution (contact mode): 10 lp/mm (50 microns).
    Easy to use, intuitive software controlsDevice is easy to use, has "easy to learn (intuitive) software controls."
    Repeatable results with automatic exposureAllows the user to select automatic exposure techniques, providing for repeatable results.
    DICOM 3.0 ComplianceBioptics Vision software is compliant with Digital Imaging and Communications in Medicine (DICOM) 3.0, and comes with DICOM Print, Store and Modality Work List (MWL). DICOM V3.0 standards and protocol structure. Provides DICOM networking interface.
    Rapid image transfer to Radiology/PathologyBioptics Vision software transfers images to Radiology and Pathology within seconds through DICOM interface.
    Adherence to design specifications and requirements"The output of these two performance data records [software validation and operational environment tests] documents that 10 X 10 VISION met its required requirements and design specifications as intended."
    Safety Features"Confirmation that all identified hazards have been adequately addressed by software functionality, the user interface, documentation or user SOP."
    Functionality (image acquisition, display, analysis, storage, etc.)Software controls complete image acquisition and processing. Handles digital x-ray image acquisition, calibration, image display, image analysis and manipulation, patient database, image archiving.
    Minimum Image Display Time4 seconds.
    Substantial Equivalence to predicate devicePerformance testing and validation studies document that 10 X 10 VISION is substantially equivalent to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify a sample size for a test set in the context of clinical or image-based performance evaluation. The testing described primarily focuses on engineering verification and validation of the device's technical specifications and software functionality.

    The data provenance is for engineering verification and validation data, not patient data. It is internal company data generated during device development ("Extremely controlled and detailed design inputs and outputs define Bioptics product development activities").

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts or the establishment of ground truth by experts for a test set. This type of evaluation is typically associated with clinical performance studies, which were explicitly stated as not being conducted or required. The testing performed was technical (software validation, compliance with radiation standards).

    4. Adjudication Method for the Test Set

    As there is no mention of a test set involving human interpretation of images or clinical outcomes, there is no adjudication method described. The "review" of performance data was done by "appropriate management" within Bioptics, Inc.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submitter explicitly states: "To the submitters knowledge, the predicate device...did not provide or reference any clinical tests submitted in compliance with 807.92(b)(2), therefore the submitter believes such clinical testing is not appropriate or required by FDA and has not made or provided any summary of such testing."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done

    The primary studies described are related to the standalone performance of the device's technical attributes and software functionality, but not in the context of an "algorithm only" with clinical outcomes without human interaction. This device is an imaging system, not an AI diagnostic algorithm. The "standalone" testing refers to the system meeting its design specifications and processing images as intended.

    7. The Type of Ground Truth Used

    For the technical and software validation, the "ground truth" was essentially the device's own design specifications and requirements, along with compliance standards (e.g., DICOM, 21 CFR 1020.30/1020.40). The testing confirmed that the device operated according to these pre-defined technical benchmarks. There was no "expert consensus," "pathology," or "outcomes data" ground truth established for the purpose of proving diagnostic accuracy.

    8. The Sample Size for the Training Set

    The document does not specify a sample size for a training set. This is expected as the device is a digital imaging system, not an AI model that typically requires a large training dataset for learning. The software "handles digital x-ray image acquisition, calibration, image display," etc., implying standard image processing and display functionalities, not machine learning model training based on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a training set in the context of an AI/ML model, the concept of establishing ground truth for a training set is not applicable to this submission. The validation focused on the system's engineering and software adherence to specifications.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1