    K Number
    K091374

    Validate with FDA (Live)

    Date Cleared
    2009-05-26

    (15 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The subject modified device is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal/OB; Abdominal (GYN & Urology); Pediatric; Small Organ (breast, testes, thyroid); Neonatal and Adult Cephalic; Cardiac (adult & pediatric); Peripheral Vascular; Intra-operative (abdominal, thoracic and PV), Musculo-skeletal Conventional & Superficial, Transrectal and Transvaginal, and Thoracic/Pleural for motion/sliding and fluid detection.

    Device Description

    The GE Compact Ultrasound is a very compact and portable diagnostic ultrasound system having three variations: LOGIQ e and Vivid e, each with options and features suited for its market niche. It has an integrated keyboard, LCD display and several interchangeable electronic-array transducers with an approximate size of 34 cm wide, 29 cm deep and 6 cm high in transport configuration and provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, an intuitive layout of specialized controls, color GUI display and Doppler audio.

    AI/ML Overview

    This is a 510(k) premarket notification for a diagnostic ultrasound system (GE LOGIQ i/e, Vivid e Ultrasound) and does not contain detailed acceptance criteria and performance study results in the way a clinical trial submission for a novel device might.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not specify quantitative acceptance criteria or detailed reported device performance metrics (e.g., sensitivity, specificity, accuracy, image quality scores). Instead, it relies on a "substantial equivalence" argument to predicate devices. The "performance" is implicitly deemed equivalent to the predicate devices.

    Acceptance Criteria (Not explicitly stated/quantified)Reported Device Performance (Implicitly equivalent to predicate)
    Safety:Conforms with applicable medical device safety standards.
    - Acoustic outputEvaluated and conforms.
    - BiocompatibilityEvaluated and conforms.
    - Cleaning and disinfection effectivenessEvaluated and conforms.
    - Electromagnetic compatibilityEvaluated and conforms.
    - Thermal, electrical, and mechanical safetyEvaluated and conforms.
    Effectiveness:
    - Overall performanceComparable to predicate GE LOGIQ e and Venue 40.
    - Key safety and effectiveness featuresSame as predicate devices.
    - Physical overall construction and materialsSame as predicate devices.
    - Technological characteristicsEquivalent to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    No specific test set or clinical data is mentioned. The submission states: "Clinical Tests: None required." Therefore, there is no sample size for a test set and no data provenance (country of origin, retrospective/prospective). The rationale is that "Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. Diagnostic ultrasound has accumulated a long history of safe and effective performance."

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Since no clinical tests or test sets were used, no experts were involved in establishing ground truth for a test set.

    4. Adjudication Method

    Not applicable, as no test set or clinical data requiring adjudication was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission explicitly states "Clinical Tests: None required." The device is cleared based on demonstrating substantial equivalence to predicate devices, not on proving improved effectiveness through head-to-head clinical trials with human readers.

    6. Standalone Performance Study

    No standalone performance study (algorithm only, without human-in-the-loop) was conducted or reported in this 510(k) submission. This device is a diagnostic ultrasound system, not an AI algorithm requiring standalone performance evaluation in the typical sense for medical imaging AI. Its performance is inherent in its imaging capabilities and user-controlled operation.

    7. Type of Ground Truth Used

    The ground truth for the device's performance, as argued in this 510(k), is essentially the established safe and effective performance of its predicate devices in clinical practice, along with compliance to recognized standards. There isn't a "ground truth" derived from expert consensus, pathology, or outcomes data for this specific submission's evaluation.

    8. Sample Size for the Training Set

    Not applicable. The GE LOGIQ i/e and Vivid e Ultrasound system is a conventional diagnostic ultrasound device, not an AI/machine learning model in the modern sense that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this conventional diagnostic ultrasound device.