Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090539

    Validate with FDA (Live)

    Date Cleared
    2009-10-30

    (245 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BiPAP AutoSV Advanced is intended to provide non-invasive ventilatory support to treat adult patients with OSA and Respiratory Insufficiency caused by central and/or mixed apneas and periodic breathing.

    Device Description

    The Respironics BiPAP AutoSV Advanced is a microprocessor controlled blower based Bi-level positive pressure system that delivers two positive pressure levels (IPAP/EPAP). The dual pressure levels provide a more natural means of delivering pressure support therapy to the patient resulting in improved patient comfort. A flow sensor and redundant pressure sensors in the patient airway feed data on measured flow and pressure into a microprocessor controller, which in turn regulates the blower assembly. A user interface displays clinical data and enables the operator to set and adjust certain clinical parameters.

    The BiPAP AutoSV Advanced pressure control contains various which are used to configure positive pressure therapies. With these controls, the device delivers minimum pressure support determined by the PSmin control. The device may automatically provide additional pressure support with inspiratory pressures between PSmin and PSmax to normalize patient ventilation during sleep disordered breathing events. Note: When PSmin = PSmax, this is equivalent to traditional bi-level therapy.

    The BiPAP AutoSV Advanced is fitted with alarms to alert the user to changes that will affect the treatment. Some of the alarms are pre-set (fixed), others are user adjustable.

    The BiPAP AutoSV Advanced Ventilatory Support System is intended for use with a patient circuit that is used to connect the device to the patient interface device (mask). A typical patient circuit consists of a six-foot disposable or reusable smooth lumen 22mm tubing and a method of venting exhaled gases.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria and a study proving the device meets them in the format requested. The document is a 510(k) summary for the BiPAP AutoSV Advanced, focusing on its substantial equivalence to predicate devices and general performance modifications.

    However, based on the text, I can extract the following relevant information:

    Key Takeaways from the Text:

    • Clinical Validation: The manufacturer performed "clinical validation to ensure the efficacy of the algorithm was not affected" due to modifications to the AutoSV algorithm.
    • Bench Data: Other device modifications were validated using "bench data" including waveform performance, triggering data, and overall event detection and control data.
    • Equivalence Claim: The testing "confirmed that the BiPAP AutoSV Advanced performs equivalently to the device predicate BiPAP AutoSV (K063540). All tests were verified to meet the required acceptance criteria."

    Given the limited information, I cannot accurately fill out many of the requested fields. Here's a structured response based on what is available and what cannot be determined:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "All tests were verified to meet the required acceptance criteria" and that the device "performs equivalently to the device predicate BiPAP AutoSV (K063540)." However, it does not provide a specific table of these criteria or quantitative reported performance values for each criterion.

    Acceptance Criteria (Example - Not explicitly stated in document)Reported Device Performance (Example - Not explicitly stated in document)
    Event detection accuracy (e.g., % sensitivity/specificity for apneas)"Improved event detection" compared to predicate. Equivalent to predicate.
    Pressure control stability/accuracy"Use of EPAP Control for pressure control when central events are detected." Equivalent to predicate.
    Response time of backup rate"Improving the response time of the backup rate." Equivalent to predicate.
    Waveform performanceMet required acceptance criteria; equivalent to predicate.
    Triggering performanceMet required acceptance criteria; equivalent to predicate.
    Overall event detection and controlMet required acceptance criteria; equivalent to predicate.

    2. Sample Size and Data Provenance:

    • Sample Size for Test Set: Not specified. The document mentions "clinical validation" but does not give the number of patients or cases.
    • Data Provenance: Not specified. It's unclear if the validation data was prospective or retrospective, or its country of origin.

    3. Number of Experts and Qualifications for Ground Truth:

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method:

    • Adjudication Method: Not specified.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • MRMC Study Done?: No information provided. The document focuses on performance equivalence to a predicate device, not on human reader improvement with AI assistance.

    6. Standalone Performance Study:

    • Standalone Study Done?: Likely yes, as implied by "bench data" and "clinical validation to ensure the efficacy of the algorithm was not affected." However, specific metrics for standalone performance (e.g., sensitivity, specificity for event detection) are not provided. The comparison is made against a predicate device.

    7. Type of Ground Truth Used:

    • Type of Ground Truth: Not specified explicitly. For clinical validation of a ventilator algorithm, it would typically involve physiological measurements and potentially expert interpretation of physiological signals (e.g., polysomnography data for sleep disordered breathing events). For bench testing, it would be against known physical models or simulated conditions.

    8. Sample Size for Training Set:

    • Sample Size: Not specified. The document mentions modifications to an existing algorithm ("AutoSV algorithm") but does not detail a training set for a de novo algorithm development.

    9. How Ground Truth for Training Set was Established:

    • How Ground Truth Established: Not specified. As it's an update to an existing algorithm, it's possible that historical data and expert-defined criteria from the predicate device's development were used, but this is conjecture.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1