Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090429

    Validate with FDA (Live)

    Date Cleared
    2009-09-18

    (211 days)

    Product Code
    Regulation Number
    884.6160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IVF ICSI dish is intended for holding oocytes and sperm during fertilization via intracytoplasmic sperm injection (ICSI).

    Device Description

    The IVF ICSI Dish is an injection molded polystyrene dish with a single dish-sized well. The dish has overall outside dimensions of 1.999" (50.77 mm) diameter and 0.340" (8.64 mm) height without the lid. Addition of the lid increases overall height to 0.384" (9.75 mm). The dish has a fluid capacity of 12.5 ml, if filled to the top of the sidewall; however, individual IVF clinics typically use a small fraction of that potential volume. The polystyrene used for the dish and lid are virgin crystal-grade polystyrene, which has successfully passed the USP 32 <88> Class VI test for in vivo cytotoxicity and USP 32 <87> for in vitro cytotoxicity. The dish is designed in such a way that when the lid is mounted on the dish, dishes can be stacked. The lid can be removed with one hand. The IVF ICSI Dish is packed in strips of 3 dishes with lids and 40 strips in a box for a total of 120 units. The IVF ICSI Dish is terminally sterilized by gamma irradiation to achieve an SAL of 10 °. The dish is non-pyrogenic as tested by LAL, and non-embryo toxic as tested by one-cell mouse embryo assay (MEA), The IVF ICSI dish is disposable and intended and labeled for single use.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the IVF ICSI Dish:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission functions as a 510(k) Premarket Notification, which primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a performance study with explicit numerical acceptance criteria and results. Therefore, the device performance is described in terms of meeting established standards and comparison to the predicate, rather than measured against specific quantitative acceptance metrics.

    Acceptance Criteria (Inferred from 510(k) Text)Reported Device Performance (IVF ICSI Dish)
    Sterility Assurance Level (SAL): 10⁻⁶Sterile (SAL 10⁻⁶), via gamma radiation
    Non-pyrogenic: Negative LAL testTested non-pyrogenic by LAL
    Non-embryotoxic: Supports embryonic developmentPassed 1-cell mouse embryo test at ≥80% expanded blastocysts at 96 hours
    Biocompatibility (in vivo): USP Class VIPolystyrene material successfully passed USP 32 <88> Class VI test for in vivo cytotoxicity
    Biocompatibility (in vitro): USP <87>Polystyrene material successfully passed USP 32 <87> for in vitro cytotoxicity
    Material Composition: Virgin crystal-grade polystyreneVirgin crystal-grade polystyrene
    Physical Design: Single well, stackable, one-hand lid removal, optically clear flat bottomSingle dish-sized well, lid can be removed with one hand, dishes can be stacked, flat and optically clear bottom.
    Manufacturing Process: Injection moldedInjection molded
    Disposability: Single useDisposable and intended and labeled for single use

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not detail specific "test set" sample sizes in the context of a clinical trial or performance study with human subjects. The evaluations related to the device's material and biological safety (sterility, pyrogenicity, embryotoxicity, cytotoxicity) are laboratory-based tests.

    • Mouse Embryo Assay (MEA): This is a biological assay. The standard practice for MEA involves a statistically significant number of mouse embryos, typically in the range of 20-30 embryos per group (test and control) across multiple replicates, to achieve statistical power for the a ≥80% expanded blastocyst rate. The document doesn't specify the exact number of mouse embryos used.
    • USP Cytotoxicity Tests: These tests follow established protocols, which involve specific cell culture setups and replicates. Sample sizes are specified by the USP monographs, but not detailed here.
    • LAL Test: This test is performed on extracts from the device, not on a "sample set" of devices in the same way clinical data is collected.

    Data Provenance: The data primarily comes from laboratory testing performed by the manufacturer (Thermo Fisher Scientific) or qualified contract laboratories. There is no mention of country of origin for the data or whether it's retrospective or prospective, as it pertains to internal product quality and safety testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to the technical and biological safety tests described for this device. The "ground truth" for these tests (e.g., whether a device is sterile, non-pyrogenic, or non-embryotoxic) is established by the well-defined protocols and acceptance criteria of the respective assays (e.g., USP standards, MEA protocols). Expert interpretation is involved in executing and verifying these lab tests, but it's not a consensus-based scoring of a "test set" in the way it would be for diagnostic imaging.

    4. Adjudication Method for the Test Set

    Not applicable. The tests performed (sterility, LAL, MEA, cytotoxicity) have objective endpoints and established analytical methods, not subjective assessments requiring adjudication by multiple readers or experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI Improvement

    Not applicable. This is a medical device (labware), not an AI/software-as-a-medical-device (SaMD) that assists human readers in making diagnostic decisions. Therefore, no MRMC study or AI-related comparative effectiveness was performed or is relevant.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable, as this device is a piece of labware, not an algorithm or AI.

    7. The Type of Ground Truth Used

    The ground truth for the various tests is established by:

    • Established biological assay standards: For embryotoxicity (MEA) and pyrogenicity (LAL), the ground truth is defined by the performance of control groups and the established threshold for acceptable biological response (e.g., ≥80% blastocyst development, absence of pyrogenic response).
    • Chemical/physical material standards: For material composition and cytotoxicity, the ground truth is based on the chemical properties of the polystyrene and the results of standardized USP tests.
    • Microbiological standards: For sterility, the ground truth is defined by the absence of microbial growth as detected by sterility testing, demonstrating the SAL.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning algorithm; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.