    K Number
    K090322

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-03-17

    (36 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    These Digital Radiographic Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

    Device Description

    Easy Moving Plus is a mobile x-ray unit that covers, with a wide range of models, all the specific needs of any radiographic examination at the patient's bed, first aid, and emergency, orthopedics, pediatric, and operating theater. The cordless Easy Moving "B" (battery powered version) combines stand alone feature for exposures with battery assisted motor drive for the greatest ease in imaging.

    AI/ML Overview

    The provided text is a 510(k) summary for the Sedecal Easy Moving Plus Mobile X-Ray System. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria for a novel device. Therefore, many of the requested categories about specific performance metrics, ground truth, and expert studies are not applicable or explicitly stated in this type of submission.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission for substantial equivalence to predicate devices, there isn't a table of specific acceptance criteria in the sense of quantitative performance targets for a new therapeutic or diagnostic algorithm. Instead, the "acceptance criteria" are implied by demonstrating that the new device's characteristics are either identical to or non-inferior to the predicate devices, particularly regarding safety and effectiveness.

    The document directly compares the Sedecal Easy Moving Plus to its predicate devices, K012663 (Easy Moving film) and K043002 (Easy Moving Digital). The performance is effectively assessed by establishing that the new device shares the same intended use, configuration, relevant performance standards, and similar technical specifications (e.g., generator type, collimator, electrical safety).

    CharacteristicPredicate Device K012663 (Easy Moving film)Predicate Device K043002 (Easy Moving Digital)Sedecal Easy Moving Plus (K090322)
    Intended UseSame as indicated belowSame as indicated belowIntended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
    ConfigurationBattery operated mobileBattery operated mobileBattery operated mobile
    Performance Standard21 CFR 1020.3021 CFR 1020.3021 CFR 1020.30
    GeneratorHigh frequency made by SedecalHigh frequency made by SedecalHigh frequency made by Sedecal
    Generator Power20 or 30 kw (2 models)16 to 50 kw (4 models)20 to 50 kw (4 models)
    CollimatorRalco R221 DHHSRalco R221 DHHSRalco R221 DHHS
    Image AcquisitionFilmDigital CANON 50GCANON 50G (Film or digital models available)
    Electrical SafetyElectrical Safety per IEC-60601. UL listedElectrical Safety per IEC-60601. UL listedElectrical Safety per IEC-60601. UL listed

    2. Sample Size Used for the Test Set and Data Provenance

    This is not a study involving a "test set" in the context of an AI algorithm or a diagnostic test with patient data. The submission relies on bench and external laboratory testing to applicable standards, as stated in Section 5 and 7. The data provenance is not specified beyond "bench and external laboratory results," implying controlled engineering and electrical testing rather than clinical data from specific countries or patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The study is not assessing a diagnostic or predictive device requiring expert-established ground truth from clinical cases. It's about the safety and performance of an X-ray system itself.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no clinical "test set" requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

    Not applicable. This is a medical device clearance for an X-ray system, not a clinical study evaluating the effectiveness of human readers with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is an X-ray system, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission are the applicable performance standards (e.g., 21 CFR 1020.30 for performance standards for diagnostic X-ray systems and their major components; IEC-60601 for electrical safety). The device's compliance with these standards (through bench and laboratory testing) serves as the "ground truth" for its safety and effectiveness.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.