Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K083234

    Validate with FDA (Live)

    Date Cleared
    2009-03-25

    (142 days)

    Product Code
    Regulation Number
    878.4370
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kimberly-Clark Corporation intends to market the sterile KC100 Surgical Drapes as devices made of natural or synthetic material intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. Kimberly-Clark intends to market the sterile KC100 Surgical Equipment Covers which are protective barrier covers that are intended to be used to cover surgical equipment and provide a protective barrier for that equipment.

    Device Description

    The surgical drapes and equipment covers described in this 510(k) submission are identical in all specifications to the predicate device models identified in K080629 except for minor variations in the widths and lengths of three models.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    BiocompatibilityMet acceptance criteria
    Barrier propertiesMet acceptance criteria
    Tensile strengthMet acceptance criteria
    Tear strengthMet acceptance criteria
    Alcohol repellencyMet acceptance criteria
    FlammabilityMet acceptance criteria

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for the test set for each individual criterion. It only states that "All results of testing met acceptance criteria."

    The data provenance is not explicitly mentioned. However, since this is a 510(k) submission to the FDA, it is highly likely that the testing was conducted by or on behalf of the manufacturer (Kimberly-Clark Corporation) to demonstrate compliance with US regulations. The nature of the tests (biocompatibility, barrier properties, strength, etc.) suggests these would be laboratory-based studies, likely prospective in nature (i.e., conducted specifically for this submission).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not provided in the document. For device performance testing (like material properties), "ground truth" is typically established by standardized testing methods and analytical instruments, not by expert consensus in the same way it would be for diagnostic image interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore not provided. Adjudication methods (like 2+1, 3+1) are relevant for studies where human interpretation or judgment is involved, especially in medical image analysis or clinical trials to resolve discrepancies. For material property testing, the results are typically objectively measured against predefined standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable and therefore not provided. The device (surgical drapes and equipment covers) is a medical device, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human readers with and without AI assistance would not be relevant.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This information is not applicable and therefore not provided. As mentioned above, the device is a physical sterile barrier product, not a software algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for this testing would be based on objective measurements and predefined material specifications/standards. For example:

    • Biocompatibility: Established by reference to ISO 10993 standards and their associated test methods and acceptable limits (e.g., cell viability percentages, irritation scores).
    • Barrier properties: Measured against standardized test methods for fluid penetration, microbial penetration, etc., with defined acceptable levels.
    • Tensile and tear strength: Measured using force-testing equipment against specified minimum strength requirements.
    • Alcohol repellency & Flammability: Measured according to relevant industry standards and acceptable criteria.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. As the device is a physical product and not an AI/machine learning model, there is no "training set" in the context of algorithm development. The testing conducted is to verify the physical properties of the manufactured product.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1