Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K082731

    Validate with FDA (Live)

    Date Cleared
    2008-12-17

    (90 days)

    Product Code
    Regulation Number
    878.4630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ML24000 UVA-1 Phototherapy Unit is a medical ultraviolet light source, which is intended for use by or under the direction of a licensed physician for the treatment of atopic dermatitis (eczema) on all skin types (I - VI).

    The ML24000 UVA-1 Phototherapy Unit is a full body medical ultraviolet device, which is intended for the treatment of atopic dermatitis (eczema) on all skin types (I - VI).

    Device Description

    The ML24000 UVA-1 Phototherapy Unit is a microprocessor controlled full body ultraviolet light source, with spectral output at peak wavelengths of 370-390 nm. It is intended for use by or under the direction of a physician, for the treatment of atopic dermatitis (eczema). The desired dose is selected using the operator interface located on the front panel of the device. The ML24000 UVA-1 Phototherapy Unit delivers full body phototherapy, whereby Philips CLEO HPA 1018 Medium Pressure Lamps, which surround the patient, deliver the specified dose of UVA-1.

    AI/ML Overview

    The provided document is a 510(k) summary for the ML24000 UVA-1 Phototherapy Unit. It states that the device is substantially equivalent to a legally commercialized predicate device. This type of submission relies on demonstrating similarity to an already approved device rather than conducting new clinical studies with detailed acceptance criteria and performance statistics as would be found in a PMA (Pre-Market Approval) application.

    Therefore, many of the specific details requested regarding acceptance criteria, study design, sample sizes, ground truth establishment, expert qualifications, and MRMC studies are not present in this document because they are not typically required for a 510(k) submission that demonstrates substantial equivalence to a predicate device for this type of medical device.

    Here's a breakdown based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    There is no explicit table of acceptance criteria or reported device performance metrics in the provided 510(k) summary. The submission focuses on demonstrating "substantial equivalence" to a predicate device, meaning its performance is considered comparable enough that specific new performance metrics are not required.

    2. Sample Size for the Test Set and Data Provenance:

    No specific test set or clinical study data is provided to determine sample size or data provenance. The substantial equivalence argument relies on the design, materials, and intended use being similar to the predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    Not applicable. No new clinical study data with ground truth established by experts is mentioned or required for this type of 510(k) submission.

    4. Adjudication Method for the Test Set:

    Not applicable. No new clinical study or test set requiring adjudication is mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a phototherapy unit, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to human reader performance with or without AI assistance is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a medical device for treatment, not an algorithm.

    7. The type of ground truth used:

    Not applicable. The basis for approval is substantial equivalence to a predicate device, not a new clinical study generating ground truth data.

    8. The Sample Size for the Training Set:

    Not applicable. This document refers to a phototherapy unit, not an AI/ML algorithm that would require a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for an AI/ML algorithm.

    Summary of the Device's Justification for Marketing based on Substantial Equivalence:

    The ML24000 UVA-1 Phototherapy Unit is considered substantially equivalent to its predicate devices (Daavlin Distributing Company 3 Series PC & SP Phototherapy Cabinet Ultraviolet Phototherapy Cabinet K063621 and Flex Controlled Phototherapy Equipment 3 Series X Ultraviolet Phototherapy Cabinet K0050695) based on the following:

    • Design Configuration: Constructed in the same design configuration.
    • Energy Sources: Utilizes similar energy sources (UV lamps).
    • Materials: Uses materials of similar and/or identical composition.
    • Intended Use: The intended use, general and specific indications for use (treatment of atopic dermatitis (eczema) on all skin types I-VI), mode of operation, labeling, treatment area, and general operating principles are the same or similar.
    • Performance Data: The performance data is stated to be "the same as or very similar to that of the claimed predicate devices." The UV lamps and cabinet construction are also noted as similar.

    The primary difference noted is the peak wavelength of the UV lamps: the ML24000 has a peak wavelength of 365nm (output spectrum 340nm to 400nm), while the predicate devices had a peak wavelength of 350nm (output spectrum 320nm to 400nm). This change was deemed not to impact the substantial equivalence for its intended use.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1