510(k) Data Aggregation

    K Number
    K082604

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-11-21

    (74 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This digital radiographic system is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremitics, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

    Device Description

    This digital diagnostic x-ray system consists of a tubehead/collimator assembly mounted on a ceiling suspension OR a U-Arm, along with a generator, generator control, and an elevating x-ray table. Power ratings for the available generators are in the rage of 50 kw to 80 kW. Exposure voltage range varies from 40 - 125 K V or 40 - 150 k V with current of 300 - 100 mA. Exposure time is 1 ms - 10 s. Models: Model DR3000 (U-arm single detector) and Model DR4000 (dual detector-ceiling suspension and table and upright bucky) and DR1000 single detector (standing bucky and cciling suspension. The digital subsystem is the SmartRad (CMT Medical Technologies) K003438.

    AI/ML Overview

    This 510(k) summary (K082604) for the Viztek DR Series Digital Diagnostic Digital X-Ray Systems (Multiple Models) is a submission for substantial equivalence to predicate devices, not a study presenting novel performance data. Therefore, many of the requested elements regarding acceptance criteria and study details are not directly applicable.

    Here's an analysis based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in terms of clinical performance metrics (e.g., sensitivity, specificity, accuracy) because it is a submission for substantial equivalence, not a direct performance study against a predefined clinical benchmark. The acceptance criteria essentially revolve around demonstrating equivalence to legally marketed predicate devices in terms of technical characteristics and intended use.

    CharacteristicAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Intended UseSame as URS LP Digital, Millennium, Radpro (Sedecal) K042876"SAME" - Intended for diagnostic radiographic exposures of skull, spine, chest, abdomen, extremities, and other body parts in adults and pediatrics, with patient sitting, standing, or lying prone/supine.
    ConfigurationU-Arm mount or Ceiling Suspension (same as predicate)"SAME" - U-Arm mount or Ceiling Suspension.
    Performance StandardCompliance with 21 CFR 1020.30 (same as predicate)"SAME" - Compliance with 21 CFR 1020.30.
    GeneratorHigh frequency made by Sedecal (same as predicate)"Uses same generator made by Sedecal".
    Electrical SafetyElectrical Safety per IEC-60601, UL listed (same as predicate)"SAME" - Electrical Safety per IEC-60601, UL listed.
    Safety and EffectivenessBench and laboratory test results indicate safety and effectiveness comparable to predicate devices."The results of bench and test laboratory indicates that the new device is as safe and effective as the predicate devices."

    Note on "Digital Panel" and "Number of panels": While these are present in the comparison chart, they highlight differences rather than direct "acceptance criteria" for performance. The Viztek device uses a Pixium 4600 panel and can have one or two panels, while the predicate uses a CANON CXDI 50G and has one. The implication is that these differences do not raise new questions of safety or effectiveness.

    2. Sample size used for the test set and the data provenance

    Not applicable. This is a 510(k) submission for substantial equivalence, not a clinical study involving a test set of patient data for performance evaluation. The "tests" mentioned are bench and laboratory tests, not clinical evaluations with a patient sample.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No clinical ground truth establishment or expert review of a test set of images is described for this type of submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a digital X-ray system, not an AI-powered diagnostic tool, and no MRMC study is described. The submission focuses on the hardware and its equivalence.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a digital X-ray system, and there is no "algorithm only" performance reported in this context. The device's primary function is to acquire X-ray images.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the clinical sense. The "ground truth" for this submission is adherence to recognized performance standards (e.g., 21 CFR 1020.30, IEC-60601 for electrical safety), and the technical specifications and intended use of the legally marketed predicate devices.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable. No training set is involved.