Found 1 results

510(k) Data Aggregation

    K Number
    K081299

    Validate with FDA (Live)

    Date Cleared
    2008-08-08

    (92 days)

    Product Code
    Regulation Number
    866.5880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TRF method is an in vitro diagnostic test for the quantitative measurement of transferrin in human serum, heparinized plasma, EDTA plasma or urine on the Dimension Vista® System. Measurements of transferrin aid in the diagnosis of malnutrition, acute inflammation, infection, and iron deficiency anemia.

    Protein 1 Calibrator is an in vitro diagnostic product for the calibration of the Dimension Vista® System for: α₁-Acid Glycoprotein (A1AG), α₁-Antitrypsin (A1AT), β₂-Microglobulin (B2MIC), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT), Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG) [serum/plasma] and (IGG-C) [cerebrospinal fluid], Immunoglobulin G Subclass 1 (IGG1), Immunoglobulin G Subclass 2 (IGG2), Immunoglobulin G Subclass 3 (IGG3), Immunoglobulin G Subclass 4 (IGG4), Immunoglobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), soluble Transferrin Receptor (STFR), Transferrin (TRF) [serum/plasma] and (TRF-U) [urine].

    PROT3 CON is an assayed, low level intralaboratory quality control for assessment of precision and analytical bias on the Dimension Vista® System in the determination of α -Microglobulin (A1MIC), specialty Albumin (sALB*), Immunoglobulin G (IGG -C*), Microalbumin (MALB) and Transferrin (TRF-U**). * For Cerebrospinal fluid (CSF) ** For urine

    Device Description

    Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.

    Protein 1 Calibrator is a multi-analyte, liquid human serum based product containing q acid glycoprotein, a-antitrypsin, ß2-microglobulin, C3 complement, C4 complement, ceruloplasmin, haptoglobin, hemopexin, homocystine, immunoglobulin A, immunoglobulin E, immunoglobulin G, immunoglobulin G subclass 1, immunoglobulin G subclass 2, immunoglobulin G subclass 3, immunoglobulin G subclass 4, immunoglobulin M, prealbumin, retinol binding protein, soluble transferrin receptor and transferrin.

    Protrein 3 Control is a multi-analyte, lyophilized, polygeline and rabbit plasma albumin based product containing a - Microglobulin, albumin, immunoglobulin G and transferrin.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Siemens Healthcare Diagnostics Inc. 510(k) Notification for the Dimension Vista® System TRF Flex® reagent cartridge, Protein 1 Calibrator, and Protein 3 Control, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary primarily focuses on establishing substantial equivalence through a method comparison study. It does not explicitly state pre-defined acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy against a gold standard, or specific numerical targets for regression analysis. Instead, the "acceptance criteria" can be inferred as achieving a strong correlation and suitable regression statistics (slope, intercept, correlation coefficient) when compared to the legally marketed predicate device.

    Acceptance Criteria (Inferred)Reported Device Performance
    Strong correlation with predicate device (Dade Behring N Antisera to Human Transferrin on BN ProSpec® System) for Transferrin (TRF) measurements in urine.Regression Analysis: Equation provided: (Specific regression equation not fully legible in the provided text, but mentioned as having been performed).
    Suitable regression statistics (slope, intercept, correlation coefficient) demonstrating substantial equivalence.Demonstrated substantial equivalence based on correlation studies and provided information.
    The modified Dimension Vista® System TRF assay, Protein 1 Calibrator, and Protein 3 Control are substantially equivalent to the legally marketed devices.Conclusion states that the devices are substantially equivalent based on the correlation studies.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The study evaluated urine samples with concentrations ranging from 2.0 to 24.4 mg/L. The exact number of samples is not explicitly stated in the provided text.
    • Data Provenance: The text does not specify the country of origin of the data. The study appears to be prospective in nature, comparing the new device against an existing one in a method comparison.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" here is not established by expert consensus on clinical interpretation, but rather by the measurements obtained from the legally marketed predicate device. This is a measurement device comparison, not a diagnostic interpretation device.

    4. Adjudication Method for the Test Set

    This is not applicable as the study is a method comparison of quantitative measurements, not a diagnostic interpretation that typically involves adjudication of reader disagreements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device is an in vitro diagnostic (IVD) reagent and calibrator/control system for quantitative measurement, not an AI-assisted diagnostic imaging or interpretation tool that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The device itself is a standalone measurement system (reagent, calibrator, control used on an automated analyzer). The comparison study implicitly tests the standalone performance of the Dimension Vista® System TRF assay against the predicate device. The performance data presented directly reflects the algorithmic (or analytical) output of the device.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the measurements obtained from the legally marketed predicate device: the Dade Behring N Antisera to Human Transferrin assay on the BN ProSpec® System. This is a form of comparative analytical performance against an established method.

    8. The Sample Size for the Training Set

    This is not applicable. This submission describes an in vitro diagnostic test system (reagent, calibrator, control). These types of devices do not typically involve "training sets" in the same way machine learning algorithms do. The development of the assay would involve internal optimization and validation, but not a distinct "training set" for an algorithm in the sense of AI.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, the concept of a "training set" and associated ground truth is not applicable to this type of IVD device submission.