Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K081008

    Validate with FDA (Live)

    Date Cleared
    2008-12-22

    (256 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lin-Zhi International (LZI) Buprenorphine Enzyme Immunoassay, when used in conjunction with Beckman Coulter® Synchron LX®, CX®, and UniCel® DxC automated clinical system analyzers, is intended for the qualitative and semi-quantitative determination of norbuprenorphine (buprenorphine metabolite) in human urine, at a cutoff value of 10 ng/mL.

    The Norbuprenorphine Drugs of Abuse (DAU) Calibrators are for use as calibrators in the qualitative and semi-quantitative calibration of the Lin-Zhi International (LZI) Buprenorphine Enzyme Immunoassay with Beckman Coulter® Synchron LX®, CX®, and UniCel® DxC automated clinical system analyzers.

    The Norbuprenorphine Drugs of Abuse (DAU) Controls are for use as assayed quality control materials to monitor the precision of the Lin-Zhi International (LZI) Buprenorphine Enzyme Immunoassay with Beckman Coulter® Synchron LX®, CX®, and UniCel® DxC automated clinical system analyzers.

    The assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Chromatography/mass spectrometry (GC/MS or LC/MS) is the preferred confirmatory method (1,2). Clinical consideration and professional judgement should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

    Device Description

    The LZI Buprenorphine assay is a homogeneous enzyme immunoassay with ready-to-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, buprenorphine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody would bind to free drug, the unbound buprenorphine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at 340 nm

    AI/ML Overview

    Here's an analysis of the provided text regarding the Lin-Zhi International (LZI) Buprenorphine Enzyme Immunoassay, Calibrators and Controls:


    1. Table of Acceptance Criteria and Reported Device Performance

    For an in vitro diagnostic (IVD) device like this immunoassay, acceptance criteria typically revolve around precision (reproducibility) and method comparison (agreement with a gold standard). The provided document presents detailed precision data. While explicit "acceptance criteria" for these values aren't stated as pass/fail thresholds in the summary, the detailed data demonstrates the device's performance characteristics.

    The "Method comparison against GC/MS confirmation device" section provides the most direct performance against a ground truth.

    Acceptance Criteria (Implied) and Reported Device Performance:

    Performance MetricImplied Acceptance Criterion (Typical for IVD)Reported Device Performance (LZI Buprenorphine Enzyme Immunoassay)
    Precision (Semi-Quantitative)
       Within-Run CV% (various concentrations)Low coefficient of variation (CV%) to indicate high reproducibility within a single run. (e.g., typically <10-20% for low concentrations, <5-10% for higher concentrations, though exact criteria vary by analyte and intended use)CX4CE Analyzer: CV% ranged from 0.0% (at 0 ng/mL) to 11.2% (at 2.5 ng/mL). LX20 Pro Analyzer: CV% ranged from n/a (at 0 ng/mL) to 13.1% (at 5.0 ng/mL). DxC600 Analyzer: CV% ranged from 0.0% (at 0 ng/mL) to 16.4% (at 2.5 ng/mL). (Overall, CV% is generally very low for higher concentrations, reflecting good precision for quantitative results.)
       Run-to-Run CV% (various concentrations)Low coefficient of variation (CV%) to indicate good reproducibility across different runs. (e.g., typically <10-20% for low concentrations, <5-10% for higher concentrations)CX4CE Analyzer: CV% ranged from n/a (at 0 ng/mL) to 14.9% (at 2.5 ng/mL). LX20 Pro Analyzer: CV% ranged from 242.2% (at 0 ng/mL, likely due to very low mean value) to 22.6% (at 2.5 ng/mL). DxC600 Analyzer: CV% ranged from 162.1% (at 0 ng/mL, likely due to very low mean value) to 10.0% (at 2.5 ng/mL). (CV% generally improves significantly at concentrations further from 0, indicating better reproducibility away from the detection limit.)
    Precision (Qualitative - Positive/Negative)
       Within-Run Agreement (at cutoff and near cutoff)Consistent results (100% Negative below cutoff, 100% Positive above cutoff). Near cutoff, some variability is expected and defined by the "gray zone" around the cutoff.CX4CE Analyzer: 100% NEG at 0, 2.5, 5.0, 7.5 ng/mL. At 10 ng/mL (Cutoff): 13 POS/8 NEG (21 total). 100% POS at 12.5, 15.0, 17.5, 20.0 ng/mL. LX20 Pro Analyzer: 100% NEG at 0, 2.5, 5.0, 7.5 ng/mL. 100% POS at 10 ng/mL (Cutoff) and above. DxC600 Analyzer: 100% NEG at 0, 2.5, 5.0, 7.5 ng/mL. At 10 ng/mL (Cutoff): 18 POS/3 NEG (21 total). 100% POS at 12.5, 15.0, 17.5, 20.0 ng/mL.
       Run-to-Run Agreement (at cutoff and near cutoff)Consistent results (100% Negative below cutoff, 100% Positive above cutoff). Near cutoff, some variability is expected and defined by the "gray zone" around the cutoff.CX4CE Analyzer: 100% NEG at 0, 2.5, 5.0, 7.5 ng/mL. At 10 ng/mL (Cutoff): 12 POS/8 NEG (20 total). 100% POS at 12.5, 15.0, 17.5, 20.0 ng/mL. LX20 Pro Analyzer: 100% NEG at 0, 2.5, 5.0, 7.5 ng/mL. 100% POS at 10 ng/mL (Cutoff) and above. DxC600 Analyzer: 100% NEG at 0, 2.5, 5.0, 7.5 ng/mL. At 10 ng/mL (Cutoff): 13 POS/7 NEG (20 total). 100% POS at 12.5, 15.0, 17.5, 20.0 ng/mL.
    Method Comparison (vs. GC/MS)
       Positive AgreementHigh percentage of agreement for samples confirmed positive by GC/MS. (e.g., typically >90-95%)CX4CE Instrument: 93.0% agreement with positive results. LX20 Pro Instrument: 97.4% agreement with positive results. DxC Instrument: 97.4% agreement with positive results.
       Negative AgreementHigh percentage of agreement for samples confirmed negative by GC/MS. (e.g., typically >90-95%)CX4CE Instrument: 97.5% agreement with negative results. LX20 Pro Instrument: 95.3% agreement with negative results. DxC Instrument: 95.3% agreement with negative results.
    Detection LimitLow detection limit for clinical relevance and to ensure identification of drug presence at therapeutically relevant or abuse levels.Determined as 3 ng/mL on all three platforms of instruments with 95% confidence.
    Linearity RangeA broad and clinically relevant range over which the assay provides accurate quantitative results.CX4CE Instrument: 3-90 ng/mL. LX20 Pro Instrument: 3-70 ng/mL. DxC Instrument: 3-70 ng/mL.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Studies:
      • Semi-Quantitative and Qualitative Precision (Within-Run): n=21 determinations for each concentration level (0 ng/mL to 20 ng/mL) on each of the three analyzer types (CX4CE, LX20 Pro, DxC600).
      • Semi-Quantitative and Qualitative Precision (Run-to-Run): n=20 determinations over 2 weeks for each concentration level (0 ng/mL to 20 ng/mL) on each of the three analyzer types.
    • Method Comparison (against GC/MS):
      • CX4CE Instrument: 83 clinical unaltered samples.
      • LX20 Pro Instrument: 82 clinical unaltered samples.
      • DxC Instrument: 83 clinical unaltered samples.
    • Data Provenance: The document states "clinical unaltered samples" for the method comparison, implying human urine samples. There is no information provided about the country of origin or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of immunoassay device does not typically involve human "experts" establishing ground truth in the way medical imaging AI might.

    • For the precision studies, the ground truth (sample concentration) was established by preparing samples to specific known concentrations (e.g., 0, 2.5, 5.0 ng/mL, etc.).
    • For the method comparison study, the ground truth was established by a confirmatory analytical method: Chromatography/mass spectrometry (GC/MS). This is a highly accurate and widely accepted gold standard for drug detection and quantification in biological samples. The expertise implicitly lies in the validated GC/MS method and trained laboratory personnel performing these analyses, rather than individual "experts" reviewing cases.

    4. Adjudication Method for the Test Set

    Not applicable for this type of IVD device study. The ground truth (GC/MS results or known sample concentration) is determined analytically, not through human consensus or adjudication. Discrepancies between the device and GC/MS would be analyzed statistically rather than adjudicated by experts.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is an in vitro diagnostic device, not an AI or human-in-the-loop diagnostic imaging system. It performs a standalone analytical measurement.


    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this entire study is a standalone performance assessment of the LZI Buprenorphine Enzyme Immunoassay. It's an "algorithm only" in the sense that the device itself performs the chemical reaction and measurement, yielding a result without direct human intervention in the interpretation of that primary result for a given sample. Humans operate the instrument and interpret the final quantitative or qualitative output, but the core analytical performance is standalone.


    7. The Type of Ground Truth Used

    • Precision Studies: The ground truth was based on known, spiked concentrations of norbuprenorphine in samples.
    • Method Comparison Study: The ground truth was established using Chromatography/mass spectrometry (GC/MS), which is considered the gold standard for confirmatory drug testing.

    8. The Sample Size for the Training Set

    No information about a "training set" is provided. This is typical for traditional IVD devices. The device's performance characteristics (e.g., standard curves, reagent formulations) are developed and optimized by the manufacturer, but the regulatory submission focuses on the validation or test set performance. There isn't a "machine learning" training phase in the context of this immunoassay.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set (in the machine learning sense) is explicitly described for this traditional IVD device. The underlying chemical principles and instrument calibration guide its function, rather than data-driven machine learning.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1