Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K080695

    Validate with FDA (Live)

    Date Cleared
    2008-07-01

    (112 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated as a single use, sterile device for use in gravity fed I.V. therapy or with a pressure infuser, when an extended fluid path is required for administration.

    Indicated as a single use, sterile device for use in I.V. therapy when an extended fluid path is required for administration.

    Device Description

    Mobile I.V. Systems device, the Primary I.V. Fluid Administration Set consists of a tubing set with either a 0.2 um or a 1.2um GVS S.p.A filter, universal spike, all position drip chamber, two injection sites and a removable Easydrop flow regulator and is used as a fluid pathway for I.V. administration. Mobile I.V. will offer three models, which will allow the choice of 10drops/ml, 20 drops/ml, or 60 drops/ml. The Primary I.V. Fluid Administration Set was designed for Emergency Medical and Hospital settings and includes an all position drip chamber. The drip chamber allows the flow of liguid regardless of the position of the drip chamber.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an I.V. Administration Set, focusing on its substantial equivalence to a predicate device. It details various bench tests and biocompatibility assessments but does NOT describe an AI-powered device or associated performance studies involving AI. Therefore, I cannot provide details on AI acceptance criteria, specific AI performance metrics, or multi-reader multi-case studies.

    However, I can extract the acceptance criteria and the studies that prove the I.V. Administration Set meets these criteria based on the information provided, treating the entire product as the "device" in this context.

    Here's an analysis of the provided information, framed to answer the questions about acceptance criteria and study descriptions for the physical medical device:


    Acceptance Criteria and Study for MOBILE I.V. SYSTEMS, LLC Primary I.V. Fluid Administration Set

    The device in question is a physical medical device, an "I.V. Administration Set," and not an AI-powered diagnostic tool. Therefore, the "acceptance criteria" relate to its functional performance, safety, and substantial equivalence to a predicate device, as demonstrated through bench testing, biocompatibility testing, and sterilization validation.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestReported Device Performance / Study Finding
    Device MalfunctionBench Testing (ISO 8536-4, Infusion sets for single use, gravity feed):Performed and validated. The submission states that these tests were performed to validate critical parameters and reduce risk, implying successful completion. The device is found substantially equivalent based on these results.
    - Section 6.2 Leak Testing
    - Section 6.3 Tensile Strength Testing
    - Section 6.8 Drip Chamber Fluid Delivery Testing
    - Section 6.10 Flow Rate Testing
    Priming Volume TestingPerformed per internal test method. Implied successful, as part of the overall substantial equivalence claim.
    Adverse Tissue ReactionBiocompatibility Testing (ISO 10993 series):Testing performed on the predicate device by GVS. All materials of construction for the Mobile I.V. Systems device are stated to be the same as the predicate device. Therefore, the predicate's successful biocompatibility results are applied. - ISO 10993-10, Sensitization - ISO 10993-11, Acute Systemic Injection - ISO 10993-3, Haemocompatibility - ISO 10993-5, Cytotoxicity MEM Elution - ISO 10993-10, Intracutaneous Injection (Note: 10993-10 appears twice for different tests; this is as listed in the document for Sensitization and Intracutaneous Injection)
    InfectionSterility Testing (ISO 11137:2006)Validation will be done according to ISO 11137:2006, with a sterility assurance level of 10-6. (Note: "Will be done" implies this was a commitment for future validation but accepted for 510(k) submission based on the protocol).
    Sterile Packaging Testing (ISO 11607:2006)Validation will be done according to ISO 11607:2006. (Same note as above).
    Pyrogen Testing (USP 30:2007, <85> and <161>)Sets were validated using the Kinetic Turbidimetric method. Implied successful.
    Bacterial Retention Testing (on predicate device, GVS Speedflow)Performed on the predicate device (same filter used). - 0.2 um filter: removed ≥99.9% of Brevundimonas diminuta. - 1.2 um filter: removed >99.9% of Candida Albicans. (These results were accepted due to the use of the same filter).
    Improper UseLabelingMitigation measure is proper labeling. While no "performance" result is given, the implication is that the labeling will be compliant and adequate to mitigate this risk. This is typically reviewed by the FDA as part of the submission process and confirmed by the approval letter.
    Overall EquivalenceSubstantial Equivalence to Predicate Device (GVS Speedflow I.V. Administration Set with Easydrop Flow Regulator, K061115). The new device has the same materials, solvents, intended use, and similar design elements (e.g., all-position drip chamber).The FDA determined the device is substantially equivalent to the predicate device, allowing it to be marketed. The summary states that "The materials and solvents... are the same as, GVS Speedflow I.V. Set w/Easydrop Regulatory. K# 061115. The intended use and test results show or will show that the Mobile I.V. Systems Primary I.V. Fluid Administration Set is substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify exact sample sizes (e.g., number of sets tested) for each bench test conducted. It only mentions what tests were performed.

    • Test Set: Not explicitly stated as a number of devices, but rather specific tests according to ISO standards.
    • Data Provenance: The tests are standard bench tests, not clinical data from patients. They are performed in a laboratory setting. Biocompatibility and Bacterial Retention data were derived from testing performed on the predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This is a physical device, and its "ground truth" is established through engineering and scientific bench testing against established international standards (ISO, USP) for physical and chemical properties, sterilization, and biocompatibility. There are no "experts" establishing a "ground truth" in the way radiologists establish ground truth for diagnostic images. The test results themselves, when compliant with the specified standards, serve as the "ground truth" for the device's performance.

    4. Adjudication Method for the Test Set

    Not applicable. This concept ("2+1", "3+1") is typically relevant for expert consensus in clinical data labeling or interpretation, especially with AI. For a physical device, compliance is determined by meeting objective quantitative or qualitative criteria within the specified test methods.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is an I.V. administration set, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance and safety is based on:

    • Compliance with International Standards: ISO 8536-4 (Infusion sets), ISO 10993 (Biocompatibility), ISO 11137 (Sterilization), ISO 11607 (Packaging), and USP 30 (Pyrogen).
    • Predicate Device Data: Biocompatibility and bacterial retention were established by tests performed on the predicate device, given the identical materials and components.
    • Internal Test Methods: For priming volume testing.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1