    Validate with FDA (Live)

    Date Cleared
    2009-07-23

    (526 days)

    Product Code
    Regulation Number
    866.3328
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SAS™ FluAlert A & B Test is a visual and rapid assay for the presumptive in-vitro qualitative detection of influenza A and influenza B viral nucleoprotein antigens from nasal washes and nasal aspirates of symptomatic patients. Negative results should be confirmed via culture. This test is not intended for the detection of Influenza Type C viral antigen. This test is for professional use only. Negative results do not preclude infection with influenza A or B and should not be used as the sole basis for treatment or other patient management decisions. It is recommended that negative results be confirmed by cell culture.

    Device Description

    The SAS™ FluAlert A & B Test utilizes antibodies against influenza type A and influenza type B viral nucleoproteins. After the extraction has been completed, the sample is placed into two separate sample wells. The specimen is absorbed and migrates via capillary action through membranes that contain dried gold conjugated antibody, which is specific for either influenza A or influenza B viral nucleoproteins. If these nucleoproteins are present, a "half-sandwich" immunocomplex is formed. The membrane contains immobilized antibody to influenza A or influenza B nucleoproteins, respectively, which bind the "half sandwich" complex. Thus, in the presence of influenza nucleoproteins, a "whole sandwich" immunocomplex is formed and a visible, pink-colored line develops in the specimen zone of the test device. In the absence of an influenza antigen, a "sandwich" immunocomplex is not formed and a negative result is indicated. To serve as a procedural control, a pink-colored control line will always appear in the control zone of each strip regardless of the presence or absence of influenza A or influenza B nucleoproteins.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the SAS™ FluAlert A & B Test, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the document. Instead, the study aims to demonstrate "substantial equivalence" to predicate devices (SAS™ Influenza A Test and SAS™ Influenza B Test). The performance benchmarks are effectively the agreement rates with these predicate devices for positive and negative influenza cases.

    CharacteristicAcceptance Criteria (Implied by Substantial Equivalence to Predicate Device)Reported Device Performance (SAS™ FluAlert A & B Test)
    Influenza A
    Positive % AgreementHigh agreement with individual SAS™ Influenza A Test97.2% (95% CI 89-100%) (Combined Fresh NW & NA)
    Negative % AgreementHigh agreement with individual SAS™ Influenza A Test99.7% (95% CI 98-100%) (Combined Fresh NW & NA)
    Influenza B
    Positive % AgreementHigh agreement with individual SAS™ Influenza B Test98.7% (95% CI 92-100%) (Combined Fresh NW & NA)
    Negative % AgreementHigh agreement with individual SAS™ Influenza B Test99.7% (95% CI 98-100%) (Combined Fresh NW & NA)
    ReproducibilityHigh agreement with expected results across sites and users88.8% - 100% agreement depending on panel member
    Cross-ReactivityNo cross-reactivity with tested viral/bacterial strainsNo cross-reactivity observed
    InterferenceNo interference from tested viral/bacterial strains with influenza detectionNo interference observed (Influenza still detected as "Pos")

    2. Sample Size Used for the Test Set and Data Provenance

    • Prospective Clinical Study:
      • Sample Size: 461 nasal clinical specimens (nasal washes and nasal aspirates combined).
      • Data Provenance: Prospectively collected from four clinical trial sites in Texas and South Dakota, USA. An approximately equal mix of adult (>21 years) and pediatric patients (0-21 years) for nasal washes, and predominantly pediatric patients for nasal aspirates.
    • Retrospective Study:
      • Sample Size: 191 frozen, archived nasal wash samples.
      • Data Provenance: Retrospective, from a children's hospital in Texas, USA.

    In the prospective study, 13 samples yielded invalid results and were excluded from the totals.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth for the clinical specimens. The comparison is made against the predicate devices (SAS™ Influenza A Test and SAS™ Influenza B Test) and also mentions "cell culture or DFA" for those predicate devices in the "Results Summary" chart reference, but not explicitly for the new device's test set.

    4. Adjudication Method for the Test Set

    The document states that the clinical specimens were tested "blindly and prospectively comparing the SAS™ Influenza A Test and The SAS™ FluAlert A&B (combined) Test performance" and similarly for Influenza B. This implies that the results of the new device were compared against the established results of the predicate devices. There is no explicit mention of an adjudication method like 2+1 or 3+1 by human readers; the comparison is device-to-device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compares the new device's performance directly to the predicate devices, not human readers with and without AI assistance. The "reproducibility" section involved "non-professional users," but this was to assess consistent device results, not to evaluate human reader improvement with AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, this is essentially a standalone performance study. The SAS™ FluAlert A & B Test is a rapid immunoassay (visual and rapid assay) that produces a visible, pink-colored line. Its performance is evaluated directly against predicate devices based on its ability to detect influenza viral nucleoprotein antigens. There is no "human-in-the-loop" component mentioned for interpreting the device's fundamental output beyond simply reading the presence or absence of a line.

    7. The Type of Ground Truth Used

    The primary ground truth for evaluating the SAS™ FluAlert A & B Test's performance in the clinical studies is the results from the substantially equivalent predicate devices (SAS™ Influenza A Test and SAS™ Influenza B Test). For these predicate devices, the document refers to a comparison to "cell culture or DFA" in an older summary, implying that those methods serve as a "gold standard" for the predicate devices, and by extension, for this new device's evaluation against the predicates.

    8. The Sample Size for the Training Set

    The document describes performance studies for a diagnostic device, not a machine learning algorithm that typically requires a separate training set. Therefore, there is no explicit training set or sample size for training mentioned as it's not applicable in the context of this type of diagnostic device submission.

    9. How the Ground Truth for the Training Set was Established

    As there is no explicit training set for a machine learning algorithm, there is no description of how ground truth for a training set was established. The studies conducted are for performance evaluation against established methods and devices.