510(k) Data Aggregation

    K Number
    K080162

    Validate with FDA (Live)

    Date Cleared
    2008-04-22

    (90 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uno - One Piece Screw-Type 3.0 mm Dental Implant is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws, to provide support for prosthetic devices of partificial teeth, in order to restore the patient chewing function. The Uno - One Piece Screw-Type 3.0 mm Dental Implant is made of one piece solid material comprised of the implant and abutment one piose bend Mandibular central and lateral incisors must be splinted if using two or more 3.0 mm implants adjacent to one another. The If acting on a can factured of Titanium alloy Gr. 5 complying with standard ASTM F136-02.

    The Uno - One Piece Screw-Type 3.5 mm Dental Implants are indicated for use in surgical and restorative applications for placement in the premolar, cuspid and incisor regions of partially edentulous jaws, to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function. The Uno - One Piece Screw-Type 3.5 mm Dental Implants are made of one piece solid material comprised of the implant and abutment one piooo oone matting are manufactured of Titanium alloy Gr. 5 complying with standard ASTM F136-02.

    Device Description

    The Uno - One Piece Screw-Type Dental Implant is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, in order to restore the patient chewing function. The Uno - One Piece Screw-Type Dental Implant is made of one piece solid material comprised of the implant and abutment combination.

    The Uno implants are provided in two diameters. The Uno 3.0 mm dental implant is indicated for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions. The Uno 3.5 mm dental implant is indicated for placement in the premolar, cuspid and incisor regions.

    The Uno implants are manufactured of Titanium alloy Gr. 5 complying with standard ASTM F136-02. The Implants are provided in several lengths of 10mm, 13mm and 16mm with diameters of 3.0mm, 3.5mm. The implants surface is sand blasted and acid etched to improve the osseointegration. The implants are tapered with double thread (2mm pitch) for fast insertion.

    The Uno implants are supplied sterile and are intended for single use only.

    AI/ML Overview

    This 510(k) submission for the "Uno - One Piece Screw-Type Dental Implant" does not describe a study to prove the device meets acceptance criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device.

    Therefore, many of the requested sections (e.g., acceptance criteria, reported device performance, sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth type for test set, training set details) are not applicable or cannot be extracted from the provided text.

    The document primarily details the device description, intended use, and argues for its equivalence to a previously cleared device.

    Here's an analysis based on the information available:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. The submission does not define specific "acceptance criteria" for performance that would be met through testing. Instead, it relies on demonstrating that the device is manufactured from the same material and has similar technological characteristics to the predicate device, implying equivalent performance.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No test set data is described. The submission relies on a comparison to a predicate device, not new clinical or performance testing data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. No test set requiring ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a dental implant, not an AI-assisted diagnostic tool. Therefore, an MRMC study related to "human readers improving with AI" is irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device (dental implant), not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. No new clinical or performance data requiring ground truth is presented in this submission. The "ground truth" for the submission is the established safety and effectiveness of the predicate device based on its prior clearance.

    8. The sample size for the training set

    • Not Applicable. There is no training set mentioned as this is not a machine learning device.

    9. How the ground truth for the training set was established

    • Not Applicable. There is no training set mentioned.

    Summary of what the submission does present regarding equivalence:

    The submission argues for Substantial Equivalence based on the following:

    • Same Intended Use: The Uno implant has the same intended use as the ZIMMER DENTAL ONE-PIECE IMPLANT (K052997). Both are for placement in the jaw to support prosthetic devices and restore chewing function.
    • Equivalent Performance Characteristics: The submission states that
      • Both products are manufactured from the same material: Titanium alloy Gr. 5 complying with standard ASTM F136-02.
      • "All other technological characteristics are similar and show equivalent performance capabilities." (This is a general statement, and specific comparative data is not provided in the text.)
    • Material: Titanium alloy Gr. 5 complying with ASTM F136-02 for both the Uno implant and the predicate device.

    Conclusion stated in the document:
    "The evaluation of the Uno - One Piece Screw-Type Dental Implants does not raise any additional concerns regarding safety and effectiveness and may therefore be considered substantially equivalent to its predicate device."

    Essentially, the "study" proving the device meets criteria is the argument for substantial equivalence to a device already deemed safe and effective by the FDA, rather than new, independent testing against pre-defined acceptance criteria.