Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K073078

    Validate with FDA (Live)

    Device Name
    RAPID TOX CUP
    Date Cleared
    2008-05-30

    (212 days)

    Product Code
    Regulation Number
    862.3610
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rapid Tox Cup™ is a one-step, lateral flow immunoassay contained in a polypropylene cup for the simultaneous detection of abused drugs in urine. 'Rapid Tox Cup'- is intended for use in the qualitative detection of the following 14 drugs of abuse in human urine at the following levels:
    Amphetamine 1000 ng/mL
    Amphetamine 500 ng/mL
    Methamphetamine 1000 ng/mL
    Methamphetamine 500 ng/mL
    3,4-methylenedioxymethamphetamine (MDMA) 1000 ng/mL
    3,4-methylenedioxymethamphetamine (MDMA) 500 ng/mL
    Buprenorphine 12.5 ng/mL
    Benzodiazepines (Oxazepam) 300 ng/mL
    Barbiturates (Butalbital) 300 ng/mL
    Oxycodone 100 ng/mL
    Methadone 300 ng/mL
    Phencyclidine 25 ng/mL
    Propoxyphene 300 ng/mL
    Opiates 300 ng/mL
    Opiates 2000 ng/mL
    Cocaine (Benzoylecgonine) 300 ng/mL
    Cocaine (Benzoylecgonine) 150 ng/mL
    Tricyclic Antidepressants (Nortriptyiline) 1000 ng/mL
    THC/ Cannabinoids (11 nor△9-THC-9-carboxylic acid) 50 ng/mL

    'Rapid Tox Cup' is intended for professional use. It is not intended for over-the-counter sale to non-professionals. This assay is a simplified screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e. gas-chromatography/mass spectrometry (GC/MS).

    The barbiturate BAR, benzodiazepine BZO and tricyclic antidepressant TCA will yield preliminary positive results when BAR, BZO, and TCA is ingested at or above therapeutic doses. There are no uniformly recognized drug levels for barbiturate, benzodiazepine, or tricyclic antidepressant in urine. Certain foods or medicines may interfere with tests for Barbiturates, Benzodiazepines, and Tricyclic Antidepressants and may cause positive results.

    There is no calibration necessary and therefore no calibrator needed for this device.

    'Rapid Tox Cup' provides only a preliminary analytical result. A more specific alternate method must be used in order to obtain a more confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse result, particularly when preliminary positive results are used.

    Device Description

    The immunoassays employed in each test strip of the 'Rapid Tox Cup' are based on the same principle of the highly specific reaction between antigens and antibodies.

    Each assay consists of a membrane strip onto which up to five different drug conjugates have been immobilized. A colloidal gold-antibody complex consisting of up to five antibodies is dried at one end of the membrane. In the absence of any drug in the urine sample, the colloidal gold-multi-antibody complex moves with the urine sample by capillary action to contact the immobilized drug conjugate. Antibody-antigen reactions occur forming a visible line in all "test" areas. The formation of a visible line in the test areas occur when the test is negative for the adjacent labeled drug.

    When drug is present in the urine sample, the drug or metabolite will compete with the immobilized drug conjugate in the test area for the limited antibody binding sites on the colloidal gold-labeled antibody complex, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a colored line in any of the test areas is indicative of a presumptive positive result.

    A control line, comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence or absence of a drug in the urine, and therefore, should be present in all reactions.

    A negative urine will produce up to six colored lines, and a positive sample will produce a colored line in the control area and no colored line(s) in the test area corresponding to the individual analyte(s) that are present in the sample.

    The device is contained in a polypropylene cup.

    AI/ML Overview

    The provided document describes the "Rapid Tox Cup," a one-step, lateral flow immunoassay for the simultaneous detection of abused drugs in urine.

    1. Acceptance Criteria and Reported Device Performance

    The document presents targeted drug concentrations (cut-off levels) for detection. These can be considered the acceptance criteria for individual drug detection sensitivity. The study claims the device's performance is reproducible and equivalent to its predicate devices.

    Analyte (Drug)Acceptance Criteria (Cut-off level)Reported Device Performance (Specificity/Sensitivity)
    Amphetamine500 ng/mLReproducible, equivalent to predicate devices
    Amphetamine1000 ng/mLReproducible, equivalent to predicate devices
    Methamphetamine500 ng/mLReproducible, equivalent to predicate devices
    Methamphetamine1000 ng/mLReproducible, equivalent to predicate devices
    3,4-methylenedioxymethamphetamine (MDMA)500 ng/mLReproducible, equivalent to predicate devices
    3,4-methylenedioxymethamphetamine (MDMA)1000 ng/mLReproducible, equivalent to predicate devices
    Buprenorphine12.5 ng/mLReproducible, equivalent to predicate devices
    Benzodiazepines (Oxazepam)300 ng/mLReproducible, equivalent to predicate devices
    Barbiturates (Butalbital)300 ng/mLReproducible, equivalent to predicate devices
    Oxycodone100 ng/mLReproducible, equivalent to predicate devices
    Methadone300 ng/mLReproducible, equivalent to predicate devices
    Phencyclidine25 ng/mLReproducible, equivalent to predicate devices
    Propoxyphene300 ng/mLReproducible, equivalent to predicate devices
    Opiates300 ng/mLReproducible, equivalent to predicate devices
    Opiates2000 ng/mLReproducible, equivalent to predicate devices
    Cocaine (Benzoylecgonine)300 ng/mLReproducible, equivalent to predicate devices
    Cocaine (Benzoylecgonine)150 ng/mLReproducible, equivalent to predicate devices
    Tricyclic Antidepressants (Nortriptyline)1000 ng/mLReproducible, equivalent to predicate devices
    THC/ Cannabinoids (11 nor△9-THC-9-carboxylic)50 ng/mLReproducible, equivalent to predicate devices

    Note: The performance is described as "reproducible" and "equivalent to predicate devices," but specific accuracy metrics (e.g., sensitivity, specificity, PPV, NPV) from the study are not provided in this summary. The performance is inferred from the demonstration of reproducibility and equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The reproducibility study used commercially available control urines.

    • Sample Size: Each sample was tested 4 times, twice daily, for 5 days, for a total of 40 tests per sample. The specific number of distinct control urine samples (e.g., negative controls, above cut-off, below cut-off) used is not explicitly stated.
    • Data Provenance: The document does not specify the country of origin of the data. The data appears to be from a prospective study conducted for validation purposes, given the controlled testing conditions (commercially available control urines, specific testing regimen).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The ground truth for the test set (control urines) was established by GC/MS (Gas Chromatography/Mass Spectrometry), which is a laboratory analytical method, not by human experts. Therefore, the information about the number and qualifications of experts is not applicable in this context.

    4. Adjudication Method for the Test Set

    Since the ground truth was established by GC/MS and the device provides a direct qualitative result (visible line or no line), there was no adjudication method described for resolving discrepancies in the test set. The results from the device were compared against the GC/MS verified concentrations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The device is an immunoassay that provides a direct qualitative result, not an imaging device requiring human interpretation, so such a study would not be typically applicable. The comparison was the device's performance against GC/MS-verified samples.

    6. If a Standalone Study Was Done

    Yes, a standalone study was done. The reproducibility study evaluated the performance of the "Rapid Tox Cup" by testing control urines multiple times and verifying concentrations by GC/MS. This assesses the algorithm's (or, in this case, the immunoassay's) performance without human-in-the-loop interpretation beyond reading the visible lines. It also explicitly states that the device in cup form houses identical strips used in predicate devices, and testing showed "no difference relative to results from the Rapid Tec or Rapid Tox products."

    7. The Type of Ground Truth Used

    The ground truth used for the test set was laboratory analytical confirmation by GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly accurate and widely accepted method for drug concentration verification.

    8. The Sample Size for the Training Set

    The document does not specify a distinct training set sample size. The description of the device's technology (immunoassay with immobilized drug conjugates and colloidal gold-antibody complex) suggests that its "training" or development involved setting up the specific chemical reactions and cutoff concentrations. While this process would involve extensive testing and optimization, the document refers to validation studies rather than a formal "training set" in the machine learning sense. The performance characteristics were evaluated using commercially available control urines for reproducibility.

    9. How the Ground Truth for the Training Set Was Established

    As with the test set, the ground truth for optimizing or validating the device's detection capabilities (analogous to a training set's ground truth in AI) would have been established through laboratory analytical methods, most likely GC/MS, to accurately determine drug concentrations in samples used during the development and optimization phases. The summary states that "All concentrations were verified by GC/MS" for the reproducibility study, indicating this method for ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1