    K072687

    Validate with FDA (Live)

    Date Cleared
    2008-11-25

    (428 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Valleylab Microwave Ablation Generator is intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors.

    Device Description

    The Valleylab Microwave Ablation Generator is designed to provide microwave power for the ablation of soft tissue in medical procedures. It was cleared for commercial distribution (K011676) by demonstrating substantial equivalence to existing Radiofrequency and Microwave Ablation systems.

    The Microwave Ablation Generator is designed to be used with proprietary microwave antennas (cleared in K011676 and K032702) which are placed directly into the tissue to be ablated. The Valleylab Microwave Ablation Generator is a nonsterile electrical device that is not intended to contact the patient.

    The Microwave Ablation Generator is used with a pump to circulate cooling fluid (sterile saline) to the antennas, and an optional integrated cart. The device's power output and duration (time) are controlled by the user via manual controls on the front panel of the generator. The display on the front of the device shows power (watts), duration of power delivery (minutes and seconds), and elapsed time.

    The antenna is connected to the generator and radiates power to the tissue under treatment. The antennas are provided sterile, for single use only. The shaft of each antenna is made of insulated stainless steel. Three antennas of different lengths are available for use in percutaneous procedures, and a single length antenna fing surgical procedures. The active portion of the antenna is 3.7cm in length for all models. The percutaneous antennas are cooled by circulating sterile saline through the antenna. The surgical antennas are not internally cooled. The system pump circulates fluid from a reservoir to the antenna through sterile tubing. Sterile fluid is confined within the antenna and tubing, and does not contact the patient.

    AI/ML Overview

    This appears to be a 510(k) summary for the Valleylab Microwave Ablation Generator, seeking to expand its intended use to include the ablation of non-resectable liver tumors.

    Here's an analysis of the provided text in relation to your questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" for the expanded indication in a quantitative, tabular format. Instead, it describes comparative non-clinical testing and refers to existing clinical studies.

    Acceptance Criteria (Implied)Reported Device Performance
    Non-clinical Equivalence (Lesion Characteristics): Ablation size and geometry of lesions should be comparable to predicate devices, and lesions should have similar pathologic and imaging characteristics.Bench Testing (Bovine Livers) & In-vivo Testing (Porcine Livers): Compared ablation size and geometry with predicate Cool-tip™ RF System. "In some instances there was a difference in lesion size. However, the lesions that are generated by both systems have similar pathologic and imaging characteristics."
    Safety and Effectiveness for Liver Tumor AblationClinical Testing (Review of Existing Studies): "A review of the existing studies have shown that the use of the system to be clinically safe and effective for ablation of liver tumors."
    Compliance with Safety Standards"evaluation of the device to the current versions of the relevant safety standards" (No specific results provided, but implies compliance was met).
    Thermal Profile Performance"thermal profile testing" (No specific results provided).
    Performance in proximity to vasculature"evaluation of ablation characteristics next to vasculature" (No specific results provided).
    Software Functionality"software testing" (No specific results provided).

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-clinical Testing (Bench and In-vivo):

      • Sample Size: Not explicitly stated (e.g., number of livers, number of ablations). It mentions "single and multiple antenna comparisons."
      • Data Provenance:
        • Bench Testing: Bovine livers (ex-vivo, not human or patient data).
        • In-vivo Testing: Porcine livers (animal model, not human or patient data).
        • Retrospective/Prospective: These were likely prospective studies designed for this submission.
    • Clinical Testing:

      • Sample Size: Not explicitly stated. The document refers to "A review of the existing studies," implying a synthesis of previously conducted research rather than a new clinical trial undertaken for this submission.
      • Data Provenance: Not specified, but generally, human clinical studies would be prospective or retrospective patient data from various clinical sites. The text indicates "existing studies," so it would be entirely retrospective in relation to this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document for either the non-clinical or clinical testing.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. For the non-clinical testing, it mentions lesions were "evaluated histologically and physically," which implies some form of assessment, but not an adjudication method. For the clinical testing, as it's a "review of existing studies," the adjudication method would depend on the individual studies reviewed, which are not detailed here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This document describes a device for microwave ablation, not an AI/CAD (Computer-Aided Detection/Diagnosis) device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not conducted as described in this summary.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device is not an algorithm. It's a medical device (generator and antennas) for tissue ablation. Therefore, a standalone algorithm performance study is not applicable.

    7. The Type of Ground Truth Used

    • Non-clinical Testing (Bovine/Porcine Livers):

      • Histology: Lesions were "evaluated histologically." This implies detailed microscopic examination of tissue.
      • Physical Evaluation: Lesions were also evaluated "physically," likely referring to gross measurements and visual inspection.
      • This combines expert assessment (histology interpretation) with direct measurement.
    • Clinical Testing (Review of Existing Studies):

      • The document states the review "shown that the use of the system to be clinically safe and effective." This often implies ground truth established through a combination of:
        • Pathology: Confirmation of tumor type and complete ablation in excised specimens or biopsies.
        • Imaging Follow-up: Assessment of tumor recurrence or residual disease using modalities like CT, MRI, or ultrasound.
        • Clinical Outcomes Data: Patient survival, complications, quality of life, etc.
      • The specific methods for establishing ground truth within the "existing studies" are not detailed in this summary.

    8. The Sample Size for the Training Set

    This document does not describe the development of an algorithm that would require a "training set." It's a hardware device (microwave ablation generator). Therefore, the concept of a training set sample size is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, this concept is not applicable to the device described.