Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K072523

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-01-22

    (137 days)

    Product Code
    Regulation Number
    862.1475
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diazyme LDL-Cholesterol Assay is intended for the in vitro quantitative determination of Low Density Lipoprotein Cholesterol in human serum or plasma. The reagents can assist in the diagnosis and treatment of patients at risk of developing coronary heart disease. Elevated LDL cholesterol is the primary target of cholesterol-lowering therapy.

    Device Description

    The assay is based on a modified polyvinyl sulfonic acid (PVS) and polyethylene-glycol methyl ether (PEGME) coupled classic precipitation method with the improvements in using optimized quantities of PVS/PEGME and selected detergents. LDL, VLDL. and chylomicron (CM) react with PVS and PEGME and the reaction results in inaccessibility of LDL, VLDL and CM by cholesterol oxidase (CHOD) and cholesterol esterase (CHER), whereas HDL reacts with the enzymes. Addition of R2 containing a specific detergent releases LDL from the PVS/PEGME complex. The released LDL reacts with the enzymes to produce H2O2 which is quantified by the Trinder reaction.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Diazyme LDL-Cholesterol Reagent, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state "acceptance criteria" as distinct, quantifiable thresholds that the device must meet on its own for approval. Instead, the performance of the Diazyme LDL-Cholesterol Reagent is presented and compared primarily to a legally marketed predicate device (Genzyme N-Geneous LDL Cholesterol Reagent, K971573) to demonstrate substantial equivalence. The implication is that performance comparable to the predicate (and within generally accepted clinical laboratory standards for such assays) constitutes the acceptance.

    Therefore, the table below reflects the reported performance of the Diazyme device and, for comparison, the predicate device as presented in the submission. The "acceptance criteria" are inferred as comparable performance to the predicate and generally good laboratory practice for quantitative assays.

    Performance MetricDiazyme LDL-Cholesterol Reagent Reported PerformanceInferred Acceptance Criteria (based on predicate & good lab practice)
    Reportable RangeSerum: 2.04 - 250 mg/dLComparable to predicate (6.6-992 mg/dL) and clinically relevant. Diazyme's range is narrower on the high end but acceptable.
    Linearity1.64 - 830 mg/dL (for serum samples)Sufficiently wide to cover clinically relevant range.
    Precision (Within Run)0.7% - 1.0% ($C_v$%) at different LDL levels ($95, 146, 210 mg/dL$)Comparable to predicate (0.62% - 0.73%) and within expected analytic variability for clinical assays.
    Precision (Total)1.4% - 1.6% ($C_v$%) at different LDL levels ($95, 146, 210 mg/dL$)Comparable to predicate (1.73% - 2.27%) and within expected analytic variability for clinical assays.
    Accuracy (Correlation Coefficient vs. Predicate)0.996 (for serum samples)High correlation (e.g., > 0.95) with a legally marketed device.
    Accuracy (Slope/Intercept vs. Predicate)y = 1.0883x + 0.6078 mg/dLSlope close to 1 and intercept close to 0 when compared to a legally marketed device. (Predicate: y = 0.95x + 3.02mg/dL)
    InterferenceLess than 10% interference with:Minimal interference from common endogenous substances at specified concentrations.
    - Triglycerides (1000 mg/dL)
    - Ascorbic acid (10 mmol/L)
    - Bilirubin (40 mg/dL)
    - Bilirubin Conjugated (40 mg/dL)
    - Hemoglobin (1000 mg/dL)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document repeatedly mentions "samples tested with Diazyme LDL-Cholesterol Reagent showed good correlation with Genzyme N-geneous LDL Cholesterol Reagent (K971573) with correlation coefficients of 0.996 for serum samples" and refers to "clinical patient samples." However, the exact number of samples used for the method comparison (test set) is not explicitly stated in the provided text.
    • Data Provenance: The text does not specify the country of origin of the data. It is implied that the data is retrospective, as it involves "samples tested" and a comparison was made using these samples. The term "clinical patient samples" also suggests that these were real-world samples collected from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For an in vitro diagnostic reagent like this, the "ground truth" for the test set is established by testing the same samples using a well-established, often reference, method or a legally marketed predicate device. The expertise lies in the laboratory performing the tests and ensuring proper execution, not typically in a panel of experts reviewing the results in the same way as, for example, image-based diagnostics.

    In this case, the predicate device (Genzyme N-geneous LDL Cholesterol Reagent) serves as the comparator or "reference" for demonstrating substantial equivalence. The ground truth for the predicate device itself is stated to be traceable to the CDC HDL reference method, which is a highly standardized and rigorous method.

    4. Adjudication Method for the Test Set

    This is not applicable in the context of this type of quantitative diagnostic assay. Adjudication methods (like 2+1 or 3+1) are typically used in studies involving subjective interpretation, such as image analysis where multiple readers might disagree, and a consensus needs to be formed for the ground truth. Here, the "truth" is a quantitative value obtained from a chemical reaction. The comparison is statistical, between two quantitative results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, especially in imaging, where human readers interpret results with and without AI assistance. The Diazyme LDL-Cholesterol Reagent is a laboratory reagent for quantitative chemical analysis, not an AI-assisted diagnostic tool that aids human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The entire performance evaluation described for the Diazyme LDL-Cholesterol Reagent is a standalone (algorithm/reagent only) performance. The device is a reagent system that produces a quantitative result. There is no "human-in-the-loop" in the interpretation of the final quantitative value (though humans operate the instruments and handle the samples). The accuracy and precision data presented are reflective of the reagent's performance on its own.

    7. The Type of Ground Truth Used

    The ground truth used for comparison (and thus, implicitly, for evaluating the Diazyme device) is derived from the results obtained from a legally marketed predicate device (Genzyme N-geneous LDL Cholesterol Reagent, K971573). The submission states: "Genzyme N-geneous LDL Cholesterol Reagent (K971573) was selected for comparing serum samples with to the results generated by Diazyme LDL-Cholesterol Reagent." The predicate device itself traces its calibrator to the NIST SRM 1915b and its method to the CDC HDL reference method, indicating a robust underlying ground truth reference.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided in the document. The Diazyme LDL-Cholesterol Reagent is a chemical reagent assay, not an AI/machine learning model that requires a "training set." Its operating parameters (e.g., reagent concentrations, reaction times) would be optimized during its development process, but this is a different concept from training an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" in the context of this chemical reagent device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1