Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K072191

    Validate with FDA (Live)

    Date Cleared
    2007-10-25

    (80 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClearVision nuclear medicine imaging system is intended for use as a diagnostic imaging device to acquire and process gated and non-gated Single Photon Emission Computed Tomography (SPECT) images.

    Used with appropriate radiopharmaceuticals, the ClearVision system produces images that depict the anatomical distribution of radioisotopes within the myocardium.

    Device Description

    The ClearVision Nuclear Medicine Imaging System acquires and processes cardiac data including gated and non-gated Single Photon Emission Computed Tomography (SPECT) studies. After completion of an acquisition, the operator can select the resulting acquisition data file to generate both qualitative and quantitative results for review by a physician. This includes processing using Release 5.6 of Segami Corporation's Mirage processing software that was previously cleared under 510(k) number K043441 dated 13-January-2005.

    The acquisition system consists of either a single or dual small field-of-view detectors with each mounted on top of a tower that contains system electronics. To support the acquisition of SPECT data, the patient chair rotates up to 360 degrees in either clockwise or counterclockwise direction.

    Prior to a patient scan, the following system features are used to ensure the myocardium is centered within each detector's field of view (FOV):

    • . Each tower can be moved horizontally along rails mounted to the floor plate.
    • . The patient chair seat pan can be moved side-to-side.
    • Vertical and a horizontal beam lasers are mounted to side of detector.

    The ClearVision system's compact footprint and small FOV detector are specifically designed for placement in a facility lacking adequate floor space for a typical nuclear medicine imaging system.

    AI/ML Overview

    The provided document is a 510(k) summary for the GVI Medical Devices ClearVision Nuclear Imaging System. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with acceptance criteria and performance metrics for the ClearVision system itself.

    Therefore, much of the requested information regarding acceptance criteria, specific study designs, sample sizes, expert qualifications, and ground truth establishment is not available in this document. The document describes a comparison of features and performance characteristics to a predicate device (Digirad Cardius 1 XPO and Cardius 2 XPO SPECT Imaging Systems, K070542) to establish substantial equivalence.

    Here's a breakdown of what can be extracted and what is not available based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" but rather presents a "Feature Comparison Summary" to demonstrate substantial equivalence to the predicate device. The performance characteristics listed are NEMA (National Electrical Manufacturers Association) standards, which are common for SPECT systems and implicitly serve as performance benchmarks.

    FeatureAcceptance Criteria (Predicate)Reported ClearVision PerformanceDoes it meet acceptance criteria?
    NEMA Reconstructed Spatial Resolution11.00 mm (for predicate)9.8 mm (central), 7.6 mm (tangential), 8.4 mm (radial)The ClearVision's spatial resolution values (smaller numbers indicate better resolution) are superior to the predicate's 11.00 mm, indicating it meets or exceeds this aspect.
    NEMA System Sensitivity160 cpm / uci (for predicate)147 cpm / uciClearVision's sensitivity is slightly lower than the predicate, but this is presented in the context of substantial equivalence, implying it is within an acceptable range for the intended use given other features. The document explicitly states "performs as well as the predicate".
    NEMA Energy Resolution< 10.5 % (for predicate)≤ 9.0 %The ClearVision's energy resolution is better (lower percentage) than the predicate, indicating it meets or exceeds this aspect.
    Energy Range50 - 170 keV (for predicate)90 – 160 keVThe ClearVision's energy range is narrower than the predicate, this is a difference in specification, but not framed as a failure to meet a criterion for substantial equivalence.
    Small Detector UFOVYes (6.2" x 8.3")Yes (8.5" x 8.5")The ClearVision has a larger UFOV, which is generally a benefit, so it meets or exceeds this.

    Note: The "Acceptance Criteria" column above is inferred from the predicate device's performance characteristics as presented for comparison in the document for demonstrating substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    Not Available. The document does not describe a clinical study with a specific test set, patient data, or data provenance (country of origin, retrospective/prospective). The comparison is based on technical specifications and performance characteristics from NEMA standards and device design.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not Available. There is no mention of a human expert review or ground truth establishment process in the context of a clinical study in this document.

    4. Adjudication Method

    Not Available. As no expert review or human-in-the-loop study is described, no adjudication method is mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention any MRMC comparative effectiveness study, nor any evaluation of human readers' improvement with or without AI assistance. The device is a SPECT imaging system, not an AI-assisted diagnostic tool in the sense of a software algorithm interpreting images.

    6. Standalone (Algorithm Only) Performance Study

    Yes, implicitly, for certain physical parameters. The performance characteristics listed (NEMA Reconstructed Spatial Resolution, NEMA System Sensitivity, NEMA Energy Resolution) are measures of the system's inherent physical performance, which can be considered "standalone" as they don't involve a human in the loop for the measurement itself. However, this is for the imaging system's hardware performance, not an "algorithm" in the sense of an AI model making diagnostic interpretations. The Segami Mirage processing software (Release 5.6) used for processing is mentioned as previously cleared (K043441), indicating its standalone processing capabilities were evaluated separately.

    7. Type of Ground Truth Used

    For the physical performance characteristics of the device, the "ground truth" is established through standardized phantom measurements according to NEMA (National Electrical Manufacturers Association) protocols. These are empirical measurements of physical performance metrics (e.g., resolution in mm, sensitivity in cpm/uci, energy resolution in %). For images generated, the ground truth for diagnostic interpretation would typically involve clinical data like pathology or patient outcomes, but this level of detail is not discussed for the ClearVision system in this document.

    8. Sample Size for the Training Set

    Not Applicable/Not Available. The ClearVision is a hardware device (SPECT scanner) combined with processing software. There is no mention of a "training set" in the context of machine learning. The system's design and engineering are based on established physics and medical imaging principles.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable/Not Available. As there is no "training set" in the machine learning sense, there is no discussion of how ground truth for such a set was established.


    Summary Explanation:

    This 510(k) submission primarily focuses on demonstrating substantial equivalence to an existing predicate device (Digirad Cardius 1 XPO and Cardius 2 XPO SPECT Imaging Systems), rather than presenting a new clinical study with novel acceptance criteria and extensive human-in-the-loop performance evaluations. The document highlights the ClearVision's technical specifications and physical performance characteristics (e.g., NEMA standards) in comparison to the predicate, arguing that despite some differences (like detector technology and collimator type), it performs "as well as" the predicate and does not introduce new safety risks. This approach is common for demonstrating equivalence of medical devices, especially imaging hardware, where performance is often benchmarked against established industry standards and predicate device capabilities.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1