Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K072157

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-08-27

    (24 days)

    Product Code
    Regulation Number
    880.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SPO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.
    For professional use only, by trained clinicians.

    Device Description

    The Ohmeda Medical Giraffe and Panda Warmers are devices with a radiant heating source intended to maintain the thermal balance of an infant patient by direct radiation of energy in the infrared region of the electromagnetic spectrum.
    The warmers operate similarly to warmers currently in use in hospitals. Radiant heat from an infrared heat source is focused onto the bed to warm the patient. The operator may select either the heater power or skin temperature control method. Depending on the control method selected, the heater is either regulated at the operator selected power level or the heater output is modulated to maintain the patient's temperature at the value selected by the operator.
    Infant radiant warmers are also used to provide thermal support during surgical procedures and during procedures such as extracorporeal membrane oxygenation, resuscitation, or other procedures requiring open access and thermal support.
    Both units also feature optional intearated SpO2 and Resuscitation Modules. The Resuscitation Module may feature either a traditional bag-and-mask technology or a T-piece technology. Both the SpO2 module and the Resuscitation Modules use existing technology.
    Infant radiant warmers may incorporate other features, such as tilting of the bed, elevating base, and data output to remote monitors or nurse call systems. Infant radiant warmers may also allow use with or attachment of an independent phototherapy device or other accessories.
    The Giraffe and Panda UPS is a stand-by power supply that will serve the purpose of temporarily maintaining the operation of the parent devices, the Giraffe and Panda Warmers.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study demonstrating the device's performance, but it focuses on performance testing rather than an AI/ML-driven diagnostic device. Therefore, several requested fields, such as MRMC studies, details about expert ground truth, and training set information, are not applicable.

    Here's the available information presented in the requested format:

    Acceptance Criteria and Device Performance Study

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Performance Specification)Reported Device Performance
    Compliance with consensus standards for infant radiant warmers.Conformance to performance specifications and multiple recognized performance standards for infant radiant warmers (established through bench testing).
    UPS Battery Life (nominal operating conditions, 60% heater power)20 minutes
    UPS Battery Life (worst case operating condition, 100% heater power)15 minutes
    UPS Battery Life (worst case operating condition, 100% heater power, with phototherapy accessory and all lights on)10 minutes
    UPS Recharge Time6 hours
    Thermal balance maintenance for neonates (by direct radiation of energy in the infrared region).The device operates similarly to warmers currently in use in hospitals, providing radiant heat from an infrared source focused onto the bed to warm the patient. The heater is regulated either at an operator-selected power level or modulated to maintain the patient's selected temperature.
    Provision of thermal support during surgical procedures, extracorporeal membrane oxygenation, resuscitation, or other procedures requiring open access and thermal support.Infant radiant warmers are explicitly stated to be used for these purposes.
    Continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (with optional integrated SpO2 monitoring feature).The optional integrated SpO2 module uses existing technology.
    Provision of basic equipment required for pulmonary resuscitation of infants through an optional integrated resuscitation system (including establishing clear airway, providing oxygen/air/oxygen mixtures, and/or manual ventilation).The optional integrated Resuscitation Module may feature either traditional bag-and-mask technology or T-piece technology, both using existing technology.
    Stand-by power supply function for temporary operation maintenance during power interruptions (e.g., power failure, unintentional disconnect, short-duration intra-hospital transport).The Giraffe and Panda UPS (medical grade battery and shelf) serves this purpose. "When disconnected from wall power, the parent device continues receiving the power that it needs to operate, but the energy is supplied by the UPS rather than the distribution line."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The study relies on "bench testing" against product specifications and recognized consensus standards.
    • Data Provenance: Not applicable, as no patient data was used. The testing was conducted in a laboratory setting ("bench testing").

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth was established by conformance to engineering specifications and recognized consensus standards for infant radiant warmers, not by expert interpretation of data.

    4. Adjudication method for the test set:

    • Not applicable. The study involved bench testing against predefined performance criteria and standards, not clinical adjudication of diagnostic findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (infant warmer) without AI or diagnostic functionality, so an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is not an algorithm, but a physical medical device. Performance was assessed through bench testing of its physical and electrical functions.

    7. The type of ground truth used:

    • Performance Specifications and Recognized Consensus Standards: The ground truth was based on the device meeting its predefined engineering performance specifications (e.g., battery life, recharge time) and conforming to established, recognized consensus standards for infant radiant warmers.

    8. The sample size for the training set:

    • Not applicable. This device does not involve machine learning and therefore has no "training set."

    9. How the ground truth for the training set was established:

    • Not applicable. This device does not involve machine learning and therefore has no "training set" or associated ground truth establishment process for training.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1