    Intended Use

    Ellipse Juvia is intended for use in dermatology and plastic surgery for treatment of:

    • Skin Resurfacing .
    • Wrinkles, Rhytides, and Furrows .
    • Acne Scars .
    Device Description

    The Ellipse Juvia system comprises the following major parts:

    • A laser console containing a CO2 laser module capable of providing a laser beam having a . wavelength of 10,600 nm.
    • A scanner that is intended to manipulate and place a pulsed beam received from the laser . console in a pre-specified pattern on the skin being treated.
    • An optical fiber providing a beam path from the laser to the scanner. .
    AI/ML Overview

    The Ellipse Juvia laser system received 510(k) clearance (K072023) based on substantial equivalence to predicate devices, not on a study demonstrating that it meets specific acceptance criteria. The submission focuses on comparing the new device's technical specifications and intended uses to those of already cleared devices to establish that the Ellipse Juvia does not raise new questions of safety or effectiveness.

    Therefore, the requested information regarding acceptance criteria and a study proving their fulfillment is not present in this 510(k submission, as it is not a requirement for substantial equivalence pathways where direct performance studies against pre-defined criteria might not be conducted.

    However, based on the provided text, we can infer the "acceptance criteria" as substantial equivalence to the predicate devices and the "study" as a detailed comparison table of technical specifications and indications for use.

    Here's an analysis based on what is provided:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred from Substantial Equivalence Basis)Reported Device Performance (Ellipse Juvia)
    Indications for Use comparable to predicates:Skin Resurfacing, Wrinkles, Rhytides, and Furrows, Acne Scars
    Technology comparable to predicates:CO2 laser module, scanner for pattern generation, fiber optic beam delivery system.
    Wavelength identical to predicates:10,600nm
    Power range comparable to predicates:0.11-20W
    Minimum scanner spot size comparable to predicates:Ø300µm (comparable or smaller than predicates)
    Max power density comparable to predicates:28.6kW/cm2 (comparable to predicates)
    Aiming beam specifications comparable to predicates:635nm, max 5mW
    Scanning speed comparable to predicates:0.3-100 Hz
    Beam activation method identical to predicates:Foot switch
    No new issues of safety or effectiveness:Concluded that the device "should not raise any new issues of safety and effectiveness."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable. The submission does not describe a clinical study with a "test set" in the traditional sense of evaluating device performance on a sample of patients or images. The "test" here is a technical comparison against predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No expert ground truth establishment for a test set is mentioned, as this was not a clinical performance study. The "truth" for substantial equivalence is based on the technical specifications and intended uses of legally marketed predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No adjudication method is mentioned, as there was no test set requiring expert review.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a medical device for dermatological and plastic surgery procedures, specifically a laser system, not an AI-powered diagnostic tool. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is irrelevant and not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a laser system, not an algorithm. Standalone performance as typically described for AI algorithms is not relevant here. The "standalone" performance would be its physical output (wavelength, power, spot size), which is characterized by its technical specifications but not presented as a "study" in the provided document.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this 510(k) submission is the pre-existing FDA clearance of the predicate devices (Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and SLIM Evolution Family of CO2 Lasers). The equivalence claim is based on the comparison of specifications and indications to these already-cleared devices, implying that their safety and effectiveness are established.

    8. The sample size for the training set

    • Not applicable. There is no "training set" mentioned, as this is a medical device clearance based on substantial equivalence, not a machine learning model.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set, there's no ground truth establishment for it.