    Intended Use

    The CUSA Selector Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery, Thoracic surgery, Laparoscopic surgery and Thoracoscopic surgery.

    Device Description

    The Selector Integra Ultrasonic Surgical Aspirator System (Selector) is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective dissection of target tissue while preserving vessels, ducts and other delicate structures. The Selector consists of a console which provides control and power functions, a surgical handpiece which provides ultrasonic mechanical energy, a titanium handpiece tip and flexible irrigation flue, and a suction/irrigation system (manifold tubing and cooling water canister).

    The Selector has been modified to enable it to fragment bone. This was accomplished by designing a new tip, which will be referred to as the Saber Tip, to work with existing system. Therefore no changes to the console, handpiece or suction/irrigation system were needed.

    AI/ML Overview

    This 510(k) summary describes a device modification, specifically a new bone tip for an existing ultrasonic surgical aspirator system. The submission focuses on demonstrating substantial equivalence to predicate devices for the expanded use case of fragmenting bone and does not include a study or acceptance criteria in the typical sense of evaluating a diagnostic or AI-driven device.

    Therefore, many of the requested categories in your prompt are not applicable to this type of regulatory submission. The document explicitly states that "no changes to the console, handpiece or suction/irrigation system were needed," only the tip was modified. This implies that the core performance of the aspirator system for soft tissue fragmentation remains established by the previous 510(k) clearances.

    Here's an analysis based on the provided document, addressing the applicable points and explaining why others are not relevant:

    1. Table of Acceptance Criteria and Reported Device Performance:

    There is no explicit table of acceptance criteria or reported device performance metrics in this 510(k) summary for the new bone tip. The submission relies on demonstrating substantial equivalence to predicate devices that are already cleared for fragmenting both soft and hard (bone) tissue. The acceptance is implied by the FDA's decision letter of "substantial equivalence" as detailed in the document.

    2. Sample Size Used for the Test Set and Data Provenance:

    Not applicable. This submission is for a device modification (new tip) and relies on demonstrating substantial equivalence to already-cleared predicate devices. There is no new "test set" of data or patient samples presented in this summary to evaluate the performance of this specific bone tip. The evaluation would have been based on engineering design, material compatibility, and potentially bench testing to confirm the tip's ability to fragment bone, as opposed to a clinical study with human data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. As noted above, there is no new "test set" for performance evaluation in the context of a diagnostic or AI study. The core claim is substantial equivalence to predicate devices already cleared for the indications of use.

    4. Adjudication Method for the Test Set:

    Not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    Not applicable. This is not a diagnostic or AI device that would typically involve a MRMC study.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done:

    Not applicable. This is a surgical instrument, not an algorithm or AI.

    7. The Type of Ground Truth Used:

    The "ground truth" in this context is the established safety and efficacy of the predicate devices for fragmenting both soft and hard tissue. The new bone tip is designed to achieve the same functional outcome as the bone tips of the predicate devices. The assessment relies on a comparison of technological characteristics and intended use.

    8. The Sample Size for the Training Set:

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable.

    Summary of what the document does provide regarding acceptance:

    • Acceptance Criteria (Implicit): The primary acceptance criterion for this 510(k) submission is to demonstrate substantial equivalence to legally marketed predicate devices. This means proving that the modified device is as safe and effective as the predicate device(s) for its intended use, which now includes bone fragmentation.
    • Study Proving Acceptance: The "study" is a comparison and justification presented in the 510(k) submission itself, stating:
      • The device (Selector Ultrasonic Surgical Aspirator System with Bone Tip) is substantially equivalent to:
        • Selector Integra Ultrasonic Surgical Aspirator System (K021989) - cleared for general ultrasonic aspiration.
        • Radionics CUSA Excel Ultrasonic Surgical Aspirator System with Bone Tip (K051947) - a predicate device that already includes a bone tip, directly supporting the expanded indication.
        • Synergetics Sonotome Ultrasonic Aspirator Tips (K020220) - another predicate for aspirator tips.
      • The modification involves designing a new bone tip (Saber Tip) to work with the existing system.
      • No changes were needed for the console, handpiece, or suction/irrigation system.
      • The technological characteristics are the same or similar to those found with the predicate devices.
      • The indications for use are aligned with the predicate devices, specifically mentioning fragmentation, emulsification, and aspiration of "soft and hard (e.g., bone) tissue."

    The FDA's decision letter (starting on page 2 of the input) confirms that the FDA reviewed the submission and determined that the device is "substantially equivalent" to the legally marketed predicate devices, thereby allowing it to proceed to market. This determination is the documented acceptance.