Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071571

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-07-30

    (52 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SurgiQuest AirSeal™ Optical Trocar & Cannula System has applications in abdominal and thoracic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments. The trocar may be used with or without visualization for primary and secondary insertions.

    Device Description

    The subject is a surgical trocar and cannula composed of medical grade materials. The device is used to create and maintain a port of entry during endoscopic surgery. It incorporates a gas seal utilizing CO2, to maintain pneumoperitoneum during the course of surgery. It is supplied with a re-circulation and filtration pump designed-to maintain pneumoperitoneum and minimize CO2 consumption during minimally invasive surgery. The recirculation and filtration pump is reusable. The AirSeal™ Trocar & Cannula and Tube Set are fully disposable and are intended for single use only.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for SurgiQuest™ AirSeal™ Optical Trocar & Cannula System

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state numerical acceptance criteria with pass/fail thresholds. Instead, it describes functional requirements and the types of testing performed to demonstrate compliance. The "Reported Device Performance" column reflects what the document states the testing showed.

    Acceptance Criteria (Implied)Reported Device Performance
    Functional Performance for Surgical Use:
    Ability to create and maintain a port of entry during simulated laparoscopic surgery.The device has been tested to show its ability to create and maintain a port of entry during simulated laparoscopic surgery.
    Ability to maintain adequate pneumoperitoneum during simulated laparoscopic surgery.It has also been tested to show its ability to maintain adequate pneumoperitoneum during the course of simulated laparoscopic surgery.
    Minimize CO2 consumption during minimally invasive surgery.The recirculation and filtration pump is designed-to maintain pneumoperitoneum and minimize CO2 consumption during minimally invasive surgery. (Testing is implied for this design feature but not explicitly stated as a direct test of CO2 consumption rates in the summary).
    Safety and Sterility:
    Compliance with electrical safety and emissions standards (IEC60601-1, IEC60601-1-2).The unit will be tested for safety and emissions in accordance with IEC60601-1, General Requirements for Safety 1: Collateral Standard: Safety Requirements for Medical Electrical Systems and IEC60601-1-2, General Requirements for Safety -Collateral Standard: Electromagnetic Compatibility -Requirements and Tests. (Implies compliance was achieved).
    Achieves a Sterility Assurance Level (SAL) of 10-6.Sterility validation is in accordance with ISO 11137:2006 (Part 1) and AAMI TIR 27:2001. A Sterility Assurance Level (SAL) of 10-6 is achieved.
    Biocompatibility (implied for medical devices)Composed of medical grade materials. (Specific testing for biocompatibility is not detailed in this summary).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not specify the sample size used for the functional performance testing (e.g., number of simulated procedures or test subjects). It states "simulated laparoscopic surgery" for the functional tests.

    For sterility validation, ISO 11137:2006 and AAMI TIR 27:2001 are referenced, which are standards for radiation sterilization dose substantiation. These standards inherently involve a statistically significant number of samples, but the exact number for this specific device is not detailed in the summary.

    Data Provenance: The data provenance is internal to the manufacturer, SurgiQuest, Inc., as evidenced by the statement "The device has been tested to show..." and "The unit will be tested...". The testing appears to be primarily prospective as it's for pre-market notification to determine substantial equivalence. The country of origin for the data is implicitly the USA, where the company is based.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document does not describe the use of experts to establish a "ground truth" for the functional performance testing. The description indicates engineering and performance testing against industry standards and simulated conditions, rather than expert clinical evaluation for a specific medical outcome. For the safety and sterility testing, the "ground truth" is established by adherence to recognized international standards (IEC, ISO, AAMI).

    4. Adjudication Method for the Test Set

    Not applicable. There is no mention of an adjudication method as the testing described does not involve subjective interpretations or multiple readers requiring consensus.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The SurgiQuest™ AirSeal™ Optical Trocar & Cannula System is a physical surgical device, not an AI or imaging diagnostic tool. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study regarding "human readers improving with AI vs without AI assistance" is not relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not a software algorithm.

    7. The Type of Ground Truth Used

    For functional and safety testing, the ground truth is established by:

    • Performance against defined simulated conditions: For creating/maintaining port of entry and pneumoperitoneum.
    • Adherence to recognized international standards: IEC60601-1, IEC60601-1-2 for electrical safety and emissions; ISO 11137:2006 and AAMI TIR 27:2001 for sterility.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm, so there is no concept of a "training set" in the context of machine learning. The "training" for the device would be its manufacturing and design optimization processes.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1