Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071565

    Validate with FDA (Live)

    Date Cleared
    2007-09-10

    (95 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biogel® Indicator™ Underglove is a disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

    Device Description

    The Biogel® Indicator™ Underglove is a sterile powder free, polymer coated latex surgical glove. The glove contains 50 micrograms or less of total water extractable protein per gram.

    AI/ML Overview

    This 510(k) summary describes the Biogel® Indicator™ Underglove, a sterile powder-free latex surgical glove. The device is classified as a Class I surgical glove (CFR 878.4461).

    Here's an analysis of the provided information covering the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D3577Meets ASTM D3577
    Physical PropertiesMeets ASTM D3577Meets ASTM D3577
    Freedom from HolesMeets ASTM D3577Meets ASTM D3577
    BiocompatibilityMeets ISO 10993-1Meets ISO 10993-1
    LAL Test - final endotoxin concentration< 0.25 EU/ml< 0.25 EU/ml
    Protein Content50 micrograms or less of total water extractable protein per gram50 micrograms or less of total water extractable protein per gram
    Non-pyrogenic ClaimImplied by "non-pyrogen statement" and "LAL Test" resultsDevice is Biogel® Indicator™ Underglove with non-pyrogen statement

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not explicitly state the sample size used for testing each characteristic. Given that the compliance is stated as "Meets ASTM D3577" and "Meets ISO 10993-1," it is understood that standard testing protocols for these specifications would have been followed, which inherently include defined sample sizes for each test. For example, ASTM D3577 specifies sampling plans for inspection of dimensions, physical properties, and freedom from holes.

    The data provenance is not explicitly mentioned (e.g., country of origin, retrospective or prospective). However, as these are technical performance tests against established standards (ASTM, ISO), the testing would typically be performed in a laboratory setting under controlled conditions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    For performance tests against established industry standards (ASTM D3577, ISO 10993-1), the "ground truth" is the standard itself. These standards define objective, measurable criteria. Therefore, the concept of "experts establishing ground truth" in the same way as, for example, diagnostic image interpretation, does not directly apply here.

    The "experts" involved would be the laboratory technicians and qualified personnel who conduct the tests according to the standard procedures and interpret the results against the specified limits. Their qualifications would include training and certification in performing these specific ASTM and ISO tests.

    4. Adjudication Method for the Test Set

    Adjudication methods like '2+1' or '3+1' are typically used for subjective assessments (e.g., medical image interpretation consensus). Since the device's performance is measured against objective, quantitative standards (ASTM D3577, ISO 10993-1, LAL test), an adjudication method in this context does not apply. The test results are either within the specified limits of the standard or they are not.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was done. The device is a surgical glove, and the evaluation focuses on its physical properties, biocompatibility, and barrier integrity as defined by objective standards, not on human interpretive tasks that would benefit from an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is a physical product (a surgical glove), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used

    The ground truth used for evaluating the device's performance is established industry and international standards (ASTM D3577, ISO 10993-1), along with the specific product claim for protein content (50 micrograms or less) and endotoxin concentration (<0.25 EU/ml). These are objective, measurable criteria.

    8. The Sample Size for the Training Set

    This question is not applicable. The device is a physical product, not an AI model or software algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable. As stated above, there is no "training set" for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1