Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071202

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-11-30

    (213 days)

    Product Code
    Regulation Number
    876.5540
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DermaPort Percutaneous Vascular Access System (PVAS™) is indicated for long-term (greater than 30 days) vascular access for hemodialysis and apheresis. The system is inscrted percutaneously and the catheter is typically placed in the internal jugular vein of an adult patient. The subclavian vein is an alternate catheter insertion site.

    Device Description

    The DermaPort Percutaneous Vascular Access System (PVAS) is designed to facilitate catheter placement, reposition, and exchange procedures while maintaining the catheter attachment, bacterial barrier, and fixation functions of the predicate catheter fibrous cuff. The main component of the PVAS is a metal port which is implanted into the subcutaneous tunnel at the catheter exit site on the chest wall. The hemodialysis catheter passes through the metal port which acts as a percutaneous conduit, into the subcutaneous tunnel, and then into the central venous system in the usual fashion. The metal surface of the PVAS port has a porous, tissue integrating coating which allows ingrowth of tissue to anchor the PVAS port. The PVAS port holds the hemodialysis catheter in place.

    The DermaPort Percutaneous Vascular Access System (PVAS) consists of the following types of components:

      1. Implanted Hemodialysis 14.5 F Catheter (24 cm, 28 cm or 32 cm lengths)
    • Guidewire; 0.038 inch (70 cm or 100 cm lengths) 2.
      1. 16F Tcaraway Set Griplock Hub
      1. 12F Polyethylene Dilator
    • న్. 14F Polyethylene Dilator
      1. Clear Female Dust Cover
      1. Injection Caps
      1. 18 GA x 2.7" Cyrolite Introducer Needle
      1. Tunneler with Tri ball tip
      1. Tunneler Sleeve
      1. DermaPort Blade
      1. Commercially available alcohol pad
      1. Commercially available adhesive wound dressing
      1. Peel-away Sheath
      1. DermaPort Percutaneous Vascular Access System (PVAS) Port

    The catheter is identical to the Hemolilow catheter, with the exception that the fabric cuff on the HemoFlow catheter is omitted. The HemoFlow catheter is cleared to market by the FDA via 510(k) number K994105.

    The Percutancous Vascular Access System (PVAS™) has been developed to support central vascular access for hemodialysis and apheresis. The PVAS port consists of a percutaneous tubular conduit, through which a standard 14.5F polyurethane hemodialysis catheter enters the subcutaneous tunnel. An integral seal surrounds the catheter and prevents microbial migration along the catheter. The PVAS port is enclosed by a silicone anchor that braces the assembly to the skin, and an associated brake holds the catheter in place within the port. A tissue integrating biomaterial surrounds the port, providing anatomical fixation and prevention of microbial migration in a manner analogous to the fabric cuff of a tunneled catheter.

    AI/ML Overview

    The provided text is a 510(k) summary for the DermaPort Percutaneous Vascular Access System (PVAS). It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain information about specific acceptance criteria, a detailed study proving performance against those criteria, or the specific elements requested in your prompt (e.g., sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details).

    The document states: "The PVAS technical characteristics and construction are substantially equivalent to the predicate device. The difference in construction was qualified with bench and animal testing." This implies that the 'study' conducted was focused on demonstrating substantial equivalence through bench and animal testing to the predicate device (Medcomp Hemo-Flow Catheter, K994105, and Med-Conduit Inc. HemoCath II, K062901), rather than a specific clinical trial with defined acceptance criteria for the new device.

    Therefore, I cannot populate the table or provide the requested details based on the provided text. The document is primarily a regulatory submission demonstrating equivalence, not a detailed study report.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1