    Intended Use

    A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner against potentially infectious materials.

    Device Description

    Powder-Free Vinyl Examination Glove

    AI/ML Overview

    The provided document, Premarket Notification 510(k) K071022 for a Powder-Free Vinyl Examination Glove, does not contain any information regarding AI/ML device performance or studies involving human readers, training sets, or ground truth as typically associated with AI-powered medical devices.

    Instead, this submission is for a medical glove, and the acceptance criteria and supporting studies are based on traditional physical and biocompatibility testing according to established ASTM and ISO standards.

    Therefore, many of the requested categories in your prompt are not applicable to this type of medical device submission. I will answer the applicable questions based on the provided text, and explicitly state when a category is not relevant.


    Acceptance Criteria and Device Performance

    1. A table of acceptance criteria and the reported device performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D 5250-00Meets ASTM D 5250-00 (implied, no specific values listed for device)
    Width (Small)85 +/- 5 mm(Implicitly meets, specific value not reported for device)
    Width (Medium)95 +/- 5 mm(Implicitly meets, specific value not reported for device)
    Width (Large)105 +/- 5 mm(Implicitly meets, specific value not reported for device)
    Width (X-Large)115 +/- 5 mm(Implicitly meets, specific value not reported for device)
    Length (all sizes)250 +/- 10 mm(Implicitly meets, specific value not reported for device)
    Thickness (Finger)0.05 mm minimum(Implicitly meets, specific value not reported for device)
    Thickness (Palm)0.1 mm minimum(Implicitly meets, specific value not reported for device)
    Physical PropertiesMeets ASTM D 5250-00Meets ASTM D 5250-00 (implied, no specific values listed for device)
    Tensile Strength (Before Aging)9 MPa minimum9 MPa minimum (stated specification)
    Tensile Strength (After Aging)9 MPa minimum9 MPa minimum (stated specification)
    Ultimate Elongation300% minimum300% minimum (stated specification)
    Freedom from pinholesMeets ASTM D 5250-00 and ASTM D 5151-99Meets ASTM D 5250-00 and ASTM D 5151-99 (implied)
    Water Leak TestAQL 2.5, Level IAQL 2.5, Level I (stated specification)
    Powder FreeMeets ASTM D 6124-00 and ASTM D 5250-00Meets ASTM D 6124-00 and ASTM D 5250-00 (implied)
    BiocompatibilityBased on ISO 10993
    Cytotoxicity - Agar Diffusion(Acceptance criteria implied by standard)Passed
    Primary Skin Irritation(Acceptance criteria implied by standard)Passed
    Guinea Pig Sensitization(Acceptance criteria implied by standard)Passed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "The performance test data that support a determination of substantial equivalence are described above in Section 7" (which is the section outlining standards and specifications). However, it does not explicitly state the sample sizes used for the physical or biocompatibility testing, nor does it specify the data provenance (country of origin, retrospective/prospective). The testing would typically be prospective, conducted by the manufacturer (Glormed Colombia S.A.).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a physical product (medical glove). Ground truth in the context of expert review (e.g., for image interpretation) is not relevant here. Performance is measured against physical and chemical standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. See explanation for #3. Adjudication methods like 2+1 or 3+1 are used for expert disagreement in diagnostic tasks, which is not relevant for glove testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical glove, not an AI software. No MRMC study or AI assistance is involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical glove, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance is established by conformance to internationally recognized industry standards such as ASTM D 5250-00, ASTM D 5151-99, ASTM D 6124-00, and ISO 10993. These standards define the acceptable physical properties, non-toxicity, and integrity of medical examination gloves. There is no biological "ground truth" from patients or pathology in the sense of AI diagnostics.

    8. The sample size for the training set

    Not applicable. This device is a medical glove and does not involve AI/ML requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. See explanation for #8.