510(k) Data Aggregation

    K Number
    K070841

    Validate with FDA (Live)

    Date Cleared
    2007-06-21

    (86 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NSI Implant System is intended to be implanted in the upper or lower jaw arches to provide support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or full arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.

    Device Description

    The NSI Implant System is intended to be implanted in the upper or lower jaw arches to provide support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or full arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.

    The 4.0 co-axis implant is not intended, nor should it be used, in conjunction with an angled abutment.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Endosseous Dental Implant System (K070841).

    It's important to note that this document is a 510(k) Premarket Notification Summary, which focuses on demonstrating substantial equivalence to a predicate device rather than strictly proving performance against pre-defined acceptance criteria through a clinical trial. Therefore, the "acceptance criteria" here are generally related to compliance with standards and demonstration of comparable performance to predicate devices, rather than specific sensitivity/specificity metrics.


    Acceptance Criteria and Reported Device Performance

    The submission relies on demonstrating substantial equivalence to previously cleared predicate devices and adherence to relevant standards for manufacturing, sterilization, and packaging. There are no explicit performance metrics in the format of sensitivity, specificity, accuracy, etc., as would be found for a diagnostic device.

    Acceptance Criterion TypeStandard/GuidanceReported Device Performance
    Material CompositionASTM F67-95 Grade III or Grade IV TitaniumCompliant: Northern Implant's screw-type implants and abutments are manufactured from ASTM F67-95 Grade III or Grade IV Titanium.
    Sterilization EfficacyISO 11737, ISO 11737-2, ISO 13409Compliant: Sterilization will be achieved using Co60 irradiation with a minimum dose of 25.0 kGy (2.5 mrads), creating a Sterility Assurance Level of 10⁻⁶. Validation done as per AAMI.
    Packaging IntegrityASTM D 4169-04, ASTM F 88-00, ASTM F 1929-98, EN 552, EN 556-1:1997, EN 868-1:1997, EN 868-5:1999, EN 868-9: 2000, EN 868-10:2000, ISO 11607Compliant: All Southern Implants packaging is validated following these standards.
    Mechanical Stability (Fatigue)FDA Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Abutments (May 12, 2004)Compliant: Fatigue testing conducted per the guidance document. Testing revealed a stable screw joint at the highest forces tested.
    Surface Modification (Biocompatibility/Adhesion)Not a specific standard, but addresses potential concernCompliant: Surface blasted with 100 micron alumina (Al₂O₃) particles (biocompatible). Relatively low pressure blasting to reduce embedment. Surface roughness (Sᵣ) of 1.43 microns (fraction of particle size). Each implant visually inspected; sample SEM testing quarterly.
    Technological CharacteristicsComparison to predicate devicesCompliant: Physical properties and designs of additional implants and accessories were comparable to legally marketed predicate devices. "Any differences in the technological characteristics did not raise new issues of safety or effectiveness."
    Intended UseSame as predicate devicesCompliant: Same intended use as predicate devices, with added option for immediate loading under specific conditions.

    Study Details

    This 510(k) submission states explicitly that clinical studies were not conducted or deemed necessary. The "study" here refers primarily to engineering bench tests and comparisons to predicate devices and standards.

    1. Sample size used for the test set and the data provenance:

      • Test Set (Engineering Studies): The document mentions "sample implant is sent for SEM testing four times a year" for surface evaluation. For fatigue testing, it states "Modification to the testing protocol was discussed with the FDA prior to conducting the test (Enclosure 8 Appendix B). Testing revealed a stable screw joint at the highest forces tested (Enclosure 8)." The exact number of implants tested for fatigue is not specified in the provided text, but it implies a representative sample.
      • Data Provenance: The engineering studies (fatigue, surface analysis) were conducted internally or by a contracted lab as part of the manufacturing and submission process. The submission does not specify a country of origin beyond the submitter's address (Fairfax, VA, USA). The studies are by nature prospective as they are conducted for the specific purpose of this submission or as part of ongoing quality control.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable in the typical sense. For engineering tests like fatigue, the "ground truth" is the established pass/fail criteria of the test method itself, which is based on an FDA guidance document. For material composition and sterilization, the ground truth is adherence to the specified ASTM/ISO/EN standards.
      • Visual Inspection/SEM: Manufacturing protocol includes visual inspection by trained personnel, and SEM testing is conducted by qualified lab personnel.
    3. Adjudication method for the test set:

      • Not applicable as this is not a study requiring human reader consensus for "ground truth". The "adjudication" is compliance with engineering and quality standards.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study is not relevant for an endosseous dental implant system, as it is not an imaging or diagnostic AI device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not an algorithm-based device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's evaluation is primarily compliance with established engineering standards (ASTM, ISO, EN), FDA guidance documents for mechanical testing, and material specifications. For surface characteristics, it involves direct physical measurement and visual/SEM verification against internal specifications and biocompatibility principles.
    7. The sample size for the training set:

      • Not applicable. There is no "training set" in the context of this device, as it is not an AI/machine learning product.
    8. How the ground truth for the training set was established:

      • Not applicable. See above.