    Intended Use

    The Shanghai Little Doctor LD8 Digital Wrist Blood Pressure Monitor is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist.

    Device Description

    The LD8 Digital Wrist Blood Pressure Monitor, Model LD8 is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable wrist cuff is wrapped around the wrist. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic semiconductor sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well-known technique in the market called the "oscillometric method".

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the LD8 Digital Wrist Blood Pressure Monitor:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard Reference)Reported Device Performance
    Electrical, Mechanical and Environment Performance Requirement (FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND)"None of the testing demonstrated any design characteristics that violated the requirement of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the LD8 Digital Wrist Blood Pressure Monitor tested met all relevant requirements of the aforementioned tests."
    Clinical Data and Analysis (ANSI/AAMI-SP-10 Standard, Section 4.4.2)"All testing results met required parameters."

    The document implies that the device met all criteria set forth by these standards. Specific numerical acceptance thresholds for blood pressure accuracy (e.g., mean difference, standard deviation) as would typically be found in ANSI/AAMI SP-10 are not explicitly stated in this summary but are referred to as "required parameters".

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the clinical test set. It mentions "clinical testing" was performed according to "Clinical Data and Analysis ANSI/AAMI-SP-10 Standard, Section 4.4.2". This standard would typically dictate the minimum number of subjects required (e.g., 85 subjects for validation).

    The data provenance (country of origin, retrospective/prospective) is not specified. However, the manufacturer is "Little Doctor (Shanghai) Electronic Manufacture Co., Ltd" located in Shanghai, China, suggesting the clinical study may have been conducted there. The study is prospective, as it is clinical testing for device approval.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide details on the number or qualifications of experts used to establish a "ground truth" in the context of reference blood pressure measurements. For clinical validation of a blood pressure monitor, the ground truth is typically established by trained observers (often physicians or nurses) using a standardized mercury sphygmomanometer or an equivalent reference device, following a strict protocol. The document states their method is "similar to the auscultatory method," implying human observation as the reference.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method. In clinical validation of blood pressure monitors, if multiple observers are used for reference measurements, their readings are often averaged or specific protocols are followed to ensure consistency. However, this is not detailed here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically associated with imaging diagnostics where human readers interpret medical images with and without AI assistance to measure improvements in diagnostic accuracy. The LD8 Digital Wrist Blood Pressure Monitor is a standalone measurement device, not an interpretive diagnostic tool that requires human-in-the-loop assistance in the same way.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the blood pressure monitor itself is a standalone device. The clinical testing evaluates the performance of the device's algorithm and hardware in isolation (without human intervention in the measurement process, though a human initiates the measurement), as compared to reference measurements. The document describes the "oscillometric method" performed by the device to analyze signals and define blood pressure, indicating a standalone algorithm for measurement.

    7. The Type of Ground Truth Used

    The ground truth used for the clinical validation is based on a reference measurement method "similar to the auscultatory method" using an "electronic semiconductor sensor" rather than a stethoscope and mercury manometer, but its definition of systolic and diastolic pressure is aligned with this established manual method. For blood pressure device validation, the gold standard typically involves simultaneous readings by trained observers using a mercury sphygmomanometer (or a validated reference device like an expertly maintained oscillometric device) following specific protocols (like those outlined in ANSI/AAMI SP-10). The summary implies adherence to this standard, so it's reasonable to infer that a well-established, expert-derived reference measurement served as the ground truth.

    8. The Sample Size for the Training Set

    The document does not mention a separate "training set" or "training data" in the typical machine learning sense. The device utilizes an "oscillometric method" and a "software algorithm" for blood pressure measurement. While this algorithm would have been developed and likely optimized using data, the document does not specify the size or nature of such a dataset. For traditional medical devices like this, the algorithm is often based on established physiological principles and signal processing, rather than purely data-driven machine learning models that require distinct training and test sets as defined in AI product development.

    9. How the Ground Truth for the Training Set Was Established

    Since a dedicated "training set" is not explicitly mentioned in the context of machine learning, the process of establishing ground truth for any underlying algorithm development is not detailed. However, for a device based on the oscillometric method, the algorithm's parameters and logic would have been designed and refined based on extensive physiological knowledge and comparisons to reference blood pressure measurements, similar to how the clinical test set ground truth would be established. This likely involved engineering and medical expertise to correlate oscillometric waveforms with auscultatory cuff pressure readings.