Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K070295

    Validate with FDA (Live)

    Date Cleared
    2007-05-21

    (110 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single Planar Multi Axial (SPMA) Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, spinal stenosis, dislocation, scoliosis, kyphosis, lordosis, spinal turnor, pseudoarthrosis, and failed previous fusion.

    Device Description

    The SPMA Pedicle Screw System is an internal fixation device for spinal surgery consisting of rods and pedicle screw assemblies (screw, saddle, axle, locking washer and set screw). The axle that connects the screw to the saddle allows the saddle to be adjusted in a single plane to any angle, up to 60 ° from the midline. Grooves in the head of the screw and in the inferior face of the locking washer interdigitate with the splined surface of the rod to allow the assembly to be rigidly fixed in place when tightened.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Trinity Orthopedics, LLC Single Planar Multi Axial (SPMA) Pedicle Screw System. The submission focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a de novo clinical study with specific acceptance criteria and performance data for the device itself.

    Therefore, many of the requested categories related to clinical study design, ground truth establishment, expert adjudication, and detailed performance metrics are not applicable to this type of regulatory submission as they are typically found in clinical trials or performance studies for novel devices or those seeking a higher level of evidence.

    Here's a breakdown of the information that can be extracted or that is not present in the provided text:

    1. A table of acceptance criteria and the reported device performance

    Criteria CategoryAcceptance CriteriaReported Device Performance
    Mechanical PerformanceEquivalence to legally marketed predicate devices."Mechanical testing performed on the SPMA Pedicle Screw System using titanium rods demonstrated equivalence of the device to legally marketed predicate devices." (Specific quantitative performance targets or results are not provided in this summary.)
    Intended Use EquivalenceSame intended use as predicate devices.The SPMA Pedicle Screw System states the same intended use as its predicates: "to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, spinal stenosis, dislocation, scoliosis, kyphosis, lordosis, spinal tumor, pseudoarthrosis, and failed previous fusion."
    Design EquivalenceSame operating principle and basic design as predicates.The SPMA Pedicle Screw System "uses the same operating principle" and "incorporates the same basic design" as predicate devices (e.g., rods, pedicle screw assemblies including screw, saddle, axle, locking washer, and set screw; features like multi-axial adjustability in a single plane, interdigitating surfaces for fixation).
    Material EquivalenceSame materials as predicate devices.The SPMA Pedicle Screw System uses "the same materials" as predicate devices. Specifically, rods are titanium alloy and Co-Cr-Mo alloy, and pedicle screw assemblies are cannulated titanium alloy.
    Substantial EquivalenceDetermination of substantial equivalence to predicates.FDA reviewed the 510(k) and determined the device is "substantially equivalent... to legally marketed predicate devices" for the stated indications for use, thereby permitting the device to proceed to market. This is the overarching "acceptance criteria" for a 510(k) submission.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This document describes a 510(k) premarket notification for a medical device (pedicle screw system), which primarily relies on demonstrating substantial equivalence to existing predicate devices through design, material, and mechanical testing, rather than a clinical "test set" of patient data. The "testing" referred to is mechanical performance testing of the device itself.
    • The provenance of data for mechanical testing would typically be from the manufacturer's internal testing facilities, but this is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This submission does not involve a clinical test set requiring expert-established ground truth. The "ground truth", in a regulatory sense, is the performance of the predicate devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As no clinical test set for diagnostic or interpretative performance is described, no adjudication method is relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not a study assessing AI assistance or human reader performance. It is a submission for a surgical implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable / Refer to Predicate Device Performance. For this 510(k) submission, the "ground truth" equivalent is the established safety and effectiveness of the legally marketed predicate devices. The new device is deemed acceptable if it can demonstrate substantial equivalence to these predicates through its design, materials, and mechanical performance.

    8. The sample size for the training set

    • Not Applicable. This device does not use machine learning, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. There is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1