510(k) Data Aggregation

    K Number
    K070188

    Validate with FDA (Live)

    Date Cleared
    2007-03-19

    (56 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

    A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power-Free Vinyl Patient Examination Gloves, 80 LYZ, and meets all requirements of ASTM standard D-5250-06.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Powder Free Vinyl Patient Examination Glove, Yellow:

    This document is a 510(k) summary for a medical device (patient examination glove) and focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel efficacy through detailed clinical studies. Therefore, many of the requested points related to advanced AI/ML studies (like MRMC, effect size, training sets, and detailed ground truth establishment for a complex algorithm) are not applicable in this context. The "acceptance criteria" here mainly refer to meeting established physical, chemical, and biocompatibility standards for gloves.


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Physical DimensionsASTM D-5250-06 (all requirements)"All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0."
    Pinholes (Barrier Integrity)FDA 1000 ml. Water Fill Test, AQL 2.5, inspection level I"The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements."
    BiocompatibilityPrimary Skin Irritation and Skin Sensitization testing"Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions."
    Powder ContentASTM D-6124-06 (no more than 2 mg powder per glove)"A Residual Powder Test that based on ASTM D-6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our 'powder-free' claims (contain no more than 2 mg powder per glove)."
    General ComplianceApplicable 21 CFR references"Hong Di Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, Yellow conform fully to ASTM D-5250-06 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7."

    Study Details (as applicable for this type of device submission)

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size:
        • For Physical/Dimensions Testing: Inspection level S-2, AQL 4.0 (specific sample size from a lot is not explicitly stated but is derivable from the AQL tables).
        • For Pinhole (Water Fill) Test: AQL 2.5, inspection level I (specific sample size from a lot is not explicitly stated but is derivable from the AQL tables).
        • For Biocompatibility Testing: Not specified, but standard animal or human patch test panels are typically used for irritation/sensitization.
        • For Residual Powder Test: Not specified, but samples from batches would be tested.
      • Data Provenance: The tests were conducted by the manufacturer, Hong Di Plastic Products Co., Ltd., in China. The data would be prospective from their manufacturing and quality control processes.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. For this type of device, "ground truth" is established by adherence to recognized industrial standards (ASTM) and FDA test methods. There are no "experts" in the sense of clinical reviewers establishing a diagnostic "ground truth" for the device's performance, but rather qualified lab personnel performing standardized tests.
    3. Adjudication method for the test set:

      • Not Applicable. The "adjudication" is essentially the pass/fail criteria defined by the ASTM standards and FDA test methods (e.g., AQL levels for defects, negative results for irritation/sensitization). There's no consensus-based adjudication as would be seen in reading medical images.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This device is a patient examination glove, not an AI-powered diagnostic or assistive tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This is a physical medical device (glove), not an algorithm.
    6. The type of ground truth used:

      • Standardized Test Results & Compliance to Specifications: The "ground truth" for this device is its performance against established, quantitative specifications outlined in ASTM standards (D-5250-06, D-6124-06) and FDA-mandated tests (1000 mL Water Fill, biocompatibility assays).
    7. The sample size for the training set:

      • Not Applicable. This is not an AI/ML device that requires a training set. The "training" for manufacturing involves process control and adherence to quality systems for all produced gloves.
    8. How the ground truth for the training set was established:

      • Not Applicable. As above, no training set in the AI/ML sense. "Ground truth" for manufacturing process quality is established through ongoing quality control and adherence to Good Manufacturing Practices (GMPs) and the specified ASTM standards throughout the production lifecycle.