510(k) Data Aggregation

    K Number
    K063766

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-04-23

    (124 days)

    Product Code
    Regulation Number
    866.5880
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Good Biotech Corp. Transferrin test system is intended to be used for the quantitative determination of transferrin in human serum by turbidimetric immunoassay (TIA). Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

    Good Biotech Corp. Transferrin Calibrator Set is intended to be used with Transferrin TIA for the quantitative determination of transferrin in serum samples.

    Good Biotech Corp. Transferrin Controls are intended to be used as the assayed quality control material for transferrin analysis.

    For In Vitro Diagnostic Use.

    Device Description

    Good Biotech Corp. Transferrin TIA is a ready to use reagent for the quantitative determination of transferrin in human serum by turbidimetric immunoassay (TIA). When transferrin of the serum sample encounters with duck anti-transferrin antibody, the aqqlutination based on the antigen-antibody reaction increases the turbidity of the sample. The value of the absorbance change at 505 nm is proportional to the transferrin concentration of the sample and is recorded by a general chemistry autoanalyzer. Then, the actual transferrin concentration of the serum sample is determined by interpolation of the calibration curve obtained by standard samples with known transferrin concentrations.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state pre-defined acceptance criteria in terms of specific numeric thresholds for slope, intercept, and correlation coefficient. Instead, it relies on a "high correlation coefficient" as an indicator of substantial equivalence. The predicate device's performance characteristics (slope, intercept, correlation coefficient) likely serve as an implicit benchmark for "high correlation."

    MetricAcceptance Criteria (Implicit from Predicate/General Expectation for Substantial Equivalence)Reported Device Performance
    SlopeClose to 1.0 (indicating similar proportionality)0.927
    Intercept (mg/dL)Close to 0.0 (indicating similar baseline/bias)19.784
    Correlation CoefficientHigh (e.g., >0.95 or >0.90, indicating strong linear relationship)0.992

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 82 serum samples
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It only mentions "serum samples."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to this kind of diagnostic device (immunoassay). The "truth" for the test set is established by the measurements obtained from the predicate device, which is an already legally marketed and validated assay. No human experts are used to interpret the results or establish "ground truth" in the way they would for imaging or clinical diagnosis.

    4. Adjudication Method for the Test Set

    Not applicable. The comparison is between two quantitative assays, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices that involve human interpretation, such as imaging AI, where the impact of AI assistance on human reader performance is evaluated. This submission is for a quantitative immunoassay.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

    Yes, a standalone study was performed. The device's performance was compared directly against the predicate device without human intervention or interpretation as part of the primary measurement. The "algorithm" here refers to the immunoassay's ability to quantitatively determine transferrin levels.

    7. Type of Ground Truth Used

    The "ground truth" in this context is the quantitative determination of transferrin levels by the predicate device (Roche Tina-quant Transferrin ver.2). This is a common approach for establishing substantial equivalence for new diagnostic assays, where a well-established and legally marketed device serves as the reference standard.

    8. Sample Size for the Training Set

    The document does not specify a training set. This is a pre-market submission for an immunoassay, not a machine learning algorithm that typically requires a distinct training and test set in the same way. The development and optimization of the immunoassay reagents and methodology would have been an iterative process by the manufacturer, but the term "training set" as commonly applied to AI/ML is not relevant here. The studies described are for validation/testing.

    9. How the Ground Truth for the Training Set Was Established

    As no training set (in the AI/ML sense) is explicitly mentioned, this question is not directly applicable. The "ground truth" for the comparative performance study was established by the predicate device's measurements.