510(k) Data Aggregation

    K Number
    K063689

    Validate with FDA (Live)

    Date Cleared
    2007-03-07

    (85 days)

    Product Code
    Regulation Number
    866.3328
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD Directigen™ EZ Flu A+B test is a rapid chromatographic immunoassay for the direct and qualitative detection of influenza A and B viral antigens from nasopharyngeal washes/aspirates, nasopharyngeal swabs and throat swabs of symptomatic patients. The BD Directigen EZ Flu A+B test is a differentiated test, such that influenza A viral antigens can be distinguished from influenza B viral antigens from a single processed sample using a single device. The test is to be used as an aid in the diagnosis of influenza A and B viral infections. All negative test results should be confirmed by cell culture because negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.

    Device Description

    The BD Directigen EZ Flu A+B test is a chromatographic assay to qualitatively detect influenza A and B viral antigens in samples processed from respiratory specimens. When specimens are processed and added to the test device, influenza A or B viral antigens bind to anti-influenza antibodies conjugated to visualizing particles in the corresponding A and B test strips. The antigen-conjugate complex migrates across the test strip to the reaction area and is captured by the line of antibody on the membrane. A positive result for influenza A is visualized as a reddish-purple line at the Test "T" position and the Control "C" position in the BD Directigen EZ device Flu A read window. A positive result for influenza B is visualized as a reddish-purple line at the Test "T" position and the Control "C" position in the BD Directigen EZ device Flu B read window.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity and specificity. Instead, the study aims to demonstrate substantial equivalence to viral cell culture. The reported performance is as follows:

    Target Performance (Implicit Acceptance Criteria: Substantial Equivalence to Viral Cell Culture)Reported Device Performance (BD Directigen™ EZ Flu A+B Test)
    Influenza A
    Sensitivity (Prospective Nasopharyngeal Swabs)90.7% (146/161)
    Specificity (Prospective Nasopharyngeal Swabs)93.2% (275/295)
    Positive Agreement (Retrospective Nasopharyngeal Swabs)85.7% (6/7)
    Negative Agreement (Retrospective Nasopharyngeal Swabs)96.2% (50/52)
    Influenza B
    Sensitivity (Prospective Nasopharyngeal Swabs)100.0% (1/1)
    Specificity (Prospective Nasopharyngeal Swabs)100.0% (455/455)
    Positive Agreement (Retrospective Nasopharyngeal Swabs)74.4% (32/43)
    Negative Agreement (Retrospective Nasopharyngeal Swabs)100.0% (16/16)

    2. Sample Size Used for the Test Set and Data Provenance

    • Prospective Test Set: 456 nasopharyngeal swab specimens. The study was a "geographically diverse multi-center study."
    • Retrospective Test Set: 59 nasopharyngeal swab specimens. These were "frozen specimens."
    • Provenance: The document doesn't specify the exact countries of origin but states it was a "geographically diverse multi-center study," implying multiple locations within potentially different regions. The retrospective samples are specifically noted as "frozen specimens."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish ground truth. The comparison is made against viral cell culture, which serves as the reference standard. Therefore, there's no information about the number or qualifications of human experts involved in ground truth establishment for this specific study.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by viral cell culture, not via human expert consensus requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test that produces a qualitative result (positive/negative for influenza A or B). Its performance is evaluated directly against a reference method (viral cell culture), not in terms of how it assists human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this is a standalone performance study of the device. The BD Directigen™ EZ Flu A+B test is a rapid chromatographic immunoassay, meaning it is a self-contained test that provides results without human interpretation beyond reading the built-in visual lines (reddish-purple lines at "T" and "C" positions). There is no "human-in-the-loop" component in the sense of a diagnostic aid to an interpreting clinician or a software algorithm; it's a direct diagnostic device.

    7. The Type of Ground Truth Used

    The ground truth used was viral cell culture. For Influenza B, retrospective, frozen specimens were also used, with viral cell culture again serving as the reference standard.

    8. The Sample Size for the Training Set

    The document does not provide information about a separate "training set" in the context of machine learning or algorithm development. This is an immunoassay device, and its performance characteristics are established through clinical validation studies as described, not through training data in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As noted above, there's no mention of a traditional training set as would be used for an AI/ML algorithm. The performance of the immunoassay is intrinsically defined by its chemical and biological components.