Found 1 results

510(k) Data Aggregation

    K Number
    K063503

    Validate with FDA (Live)

    Date Cleared
    2007-04-27

    (158 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OptiVantage DH Injector System with Enhanced Communication is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used with computed tomography (i.e. "CT") equipment.

    Device Description

    The OptiVantage DH Injector System with Enhanced Communication delivers radiographic contrast media and/or saline at a controlled flow rate and volume into a patient's vascular system for the purpose of obtaining enhanced diagnostic images. The OptiVantage DH Injector System with Enhanced Communication consists of 4 main components, just like the predicate device:

    • Power Head- Contains two electromechanical syringe drive systems, the syringe holding mechanisms, the main microprocessor, control electronics, control keypad for programming and initiating injection protocols, a status display, and two purge/retract manual knobs. The OptiVantage uses a drive system that is similar to existing Mallinckrodt injectors in the marketplace. It uses a motor and gearbox coupled to a ball screw. The ball screw drives a ram that attaches to the syringe plunger in order to fill or expel the contents of the syringe.
    • Power Supply- The power supply converts the line voltage to the working voltage for the power head and console (approximately 24-vdc).
    • Console Communicates with the Power Head to program and initiate injection protocols, displays the injection status, and displays a timer.
    • Syringes The OptiVantage DH Injector System with Enhanced Communication accommodates the Syringes - The Oper rainage styles as well as a 200 ml front loading empty syringe. These Minimonrol. I 22 million place on the market and are currently used with the predicate device.
    AI/ML Overview

    The provided text is a 510(k) summary for the OptiVantage DH Injector System with Enhanced Communication. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria, a study proving device performance against those criteria, or details regarding sample sizes, ground truth establishment, or expert involvement for a performance study.

    The document is a submission to the FDA for market clearance, focusing on demonstrating that the new device is substantially equivalent to a previously cleared predicate device (K042744). The comparison table highlights features that are identical between the new device and the predicate. The "enhancement" mentioned is an additional layer of enforcement for existing safety features, rather than a new performance characteristic requiring a new clinical performance study with acceptance criteria.

    Therefore, many of the requested details about performance studies, data provenance, expert ground truth, etc., are not present in this 510(k) summary document.

    Here's a breakdown of what is and is not available in the provided text:

    1. Table of acceptance criteria and reported device performance:

    • Not available. The document does not specify quantitative acceptance criteria (e.g., specific thresholds for flow rate accuracy or pressure limits) against which a study's results would be measured. It rather lists specifications and features that are identical to the predicate device.

    2. Sample size used for the test set and the data provenance:

    • Not available. No performance study or test set is described.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not available. Not applicable as no performance study with ground truth establishment is described.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not available. Not applicable.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:

    • Not applicable. This device is a power injector for contrast media, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not available. No standalone performance study is mentioned. The device's performance is implicit in its specifications, which are deemed substantially equivalent to the predicate device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not available. Not applicable.

    8. The sample size for the training set:

    • Not available. Not applicable as no machine learning algorithm development is described.

    9. How the ground truth for the training set was established:

    • Not available. Not applicable.

    What can be extracted from the document related to "performance" and "criteria":

    The document relies on substantial equivalence to a predicate device (OptiVantage DH Power Injection System, K042744). The "acceptance criteria" are implicitly that the new device performs identically or acceptably compared to the predicate for all relevant functions.

    Reported Device Performance (based on comparison to predicate):

    FeatureOptiVantage DH Injector System with Enhanced Communication (New Device)
    Intended UseIdentical to predicate
    Multi-phasic Injections6 phases per protocol (Identical to predicate)
    OptiBolusYes (Identical to predicate)
    Protocol Storage40 protocols (Identical to predicate)
    Syringe SizesAll Mallinckrodt 125-ml, Liebel-Flarsheim 200 ml (Identical to predicate)
    Single or Dual Syringe ModalitySingle syringe and Dual syringe (Identical to predicate)
    Syringe Drive SystemElectromechanical (Identical to predicate)
    Syringe HeaterYes (Built-in, Powerhead controlled) (Identical to predicate)
    Syringe Fill Rate2 to 15 ml/sec (Identical to predicate)
    Flow Rate0.1 to 10 ml/sec (Identical to predicate)
    Max Pressure Limit325 psi (Identical to predicate)
    Pressure Limit ControlUser-settable or automatic 25 psi increments (Identical to predicate)
    Flushing SystemManual or via injector (Identical to predicate)
    Remote StartYes (Identical to predicate)
    Scan Delay0-600 seconds (Identical to predicate)
    Phase Delays0-600 seconds plus Pause (Identical to predicate)
    Console ControlsTouch screen Color LCD (Identical to predicate)
    Power head ControlsTouch screen Color LCD (Identical to predicate)
    Post Injection ReadoutYes (Identical to predicate)
    OEM InterfaceRelays & Optical Couplings/Serial (CAN) (Identical to predicate)
    Safety Stop MechanismElectrical Stop when injection parameters are out of specifications (Identical to predicate)
    Remote Check for AirYes (Identical to predicate)
    Volume Remaining DisplayDisplayed on Power Head and Console (Identical to predicate)
    Used Syringe DetectionYes (with Enhanced Communication) (New enhancement, assumed to meet safety/function criteria)