510(k) Data Aggregation

    K Number
    K063441

    Validate with FDA (Live)

    Date Cleared
    2007-02-12

    (90 days)

    Product Code
    Regulation Number
    876.5540
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spire Biomedical Inc's Alta Gold, Coated Twin Lumen Chronic Indications for Use: Hemodialysis Catheter with fixed tips is designed for chronic hemodialysis and apheresis. It is a radiopaque polyurethane with a heparin coating, designed for percutaneous insertion or insertion via cutdown. The ability of the Carmeda End point Bonded Heparin Coating to reduce clotting is supported by in-vitro testing. Catheters longer than 40cm are intended for femoral vein insertion.

    Device Description

    Spire Biomedical, Inc.'s 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter is processed with a proprietary Carmeda® BioActive Surface (CBAS®) coating technology that attaches a functionally active heparin to the surfaces of the device. The coating counteracts thrombus from forming on the catheter. Spire's 15.5Fr Alta Gold, Heparin Coated Catheter is fully coated with CBAS® on the internal surface and on the external surface of the catheter body (from 2cm distal to the ends of the distal tips; the cuff is not coated).

    The coating is essentially non-leaching. Additionally, the maximum amount of heparin on the surface is only 1mg. Therefore; the effects of the entire coating on a patient's coagulation status would be insignificant.

    AI/ML Overview

    Acceptance Criteria and Study for K063441: 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter

    This document describes the acceptance criteria and the study that proves the Spire Biomedical, Inc.'s 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceStudy Type
    Coating PerformanceHeparin bioactivity maintained for a minimum of 90 days.Heparin bioactivity remained essentially constant throughout the 90-day test period.In-vitro durability test
    Reduction in total thrombus accumulation.91% reduction in total thrombus accumulation compared to uncoated catheters.In-vitro thromboresistance study
    Effective in mitigating disturbed flow-mediated thrombosis (at catheter tip).Demonstrated effectiveness in mitigating disturbed flow-mediated thrombosis.In-vitro thromboresistance study
    Effective in mitigating fibrin sheath propagation (on catheter shaft).Demonstrated effectiveness in mitigating fibrin sheath propagation.In-vitro thromboresistance study
    Coating is essentially non-leaching.The coating is essentially non-leaching.Stability testing
    Maximum amount of heparin on the surface is 1mg.Maximum amount of heparin on the surface is only 1mg.(Implied by design, not a direct performance measure in studies)
    Physical/Mechanical PropertiesCatheter flowsMet established performance (specific values not provided but implied by passing).Bench testing
    Tensile strength (Catheter to hub, Extension to hub, Luer adapter to extensions, Cuff to catheter)Met established performance (specific values not provided but implied by passing).Mechanical testing
    Catheter extrusionMet established performance (specific values not provided but implied by passing).Mechanical testing
    BiocompatibilityMeets requirements for a permanent contact device.Met the requirements for a permanent contact device per ISO 10993.Biocompatibility testing

    2. Sample Size Used for the Test Set and Data Provenance

    The primary performance studies for the Alta Gold catheter leveraged data from previously cleared devices, particularly the 15.5Fr Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter (K060155), due to identical material, outside dimensions, and coating.

    • In-vitro durability test (90-day bioactivity): Not explicitly stated, but performed on "coated catheters in saline."
    • In-vitro thromboresistance study (2-hour blood loop): Involved "circulation of bovine blood" through an outer loop with "coated or uncoated catheter." The number of catheters tested is not specified but it's a comparative study. This study was performed on the Alta Gold catheters, while other tests were performed on the predicate Decathlon Gold catheters.
    • Biocompatibility Testing: Conducted on the 15.5Fr Decathlon coated catheter. Number of samples not explicitly stated but implied to be sufficient for ISO 10993 compliance.

    Data Provenance:
    All studies mentioned are in-vitro laboratory studies, and thus the concept of "country of origin of the data" or "retrospective/prospective" does not apply in the same way as for clinical trials with human subjects. The studies were conducted by Spire Biomedical, Inc. (or their contractors) in support of regulatory submission in the United States.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This submission does not involve a standard clinical study with expert interpretation of cases. The "ground truth" for the performance claims is established through objective, quantitative, in-vitro scientific measurements and validated international standards (e.g., ISO 10993). Therefore, there were no "experts" in the traditional sense establishing a diagnostic ground truth for a test set of medical images or patient data. The scientists and technicians conducting the laboratory tests provided the data.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) was used as this was not a study involving human interpretation of medical data requiring consensus or arbitration. The performance data was derived from laboratory measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. This device is a medical device (catheter) and not an AI or imaging diagnostic tool that would typically involve human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This device is a chronic hemodialysis catheter, not an algorithm or AI system. Its performance is inherent to the device's material properties and coating, not dependent on an algorithm.

    7. Type of Ground Truth Used

    The ground truth used for performance validation was based on:

    • Quantitative In-vitro Measurements: Direct measurements of heparin bioactivity (pmol/cm²), thrombus accumulation (radiolabeled measurements), pressure increase, and visual evaluation in controlled laboratory settings.
    • Adherence to Standards: Compliance with established physical and mechanical test methods, and international biocompatibility standards (ISO 10993).

    8. Sample Size for the Training Set

    Not applicable. This device does not use machine learning or AI, and therefore no "training set" was used. The device's design and performance are based on engineering principles and established biological interactions, not data training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set.