510(k) Data Aggregation

    K Number
    K063121

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-11

    (91 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PORT-X II Portable X-Ray system is intended to be used by trained dentists and dental technicians as an extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

    Device Description

    PORT-X II is a portable dental X-ray system that operates on 22.2VDC supplied by a rechargeable Li-Polymer battery pack. The X-ray tubehead, X-ray controls, and power source are assembled into a single hand-held enclosure. The package includes spare batteries, a battery charger. This equipment generates and controls X-ray in order to diagnose of hand, tooth and jaw. It is composed of X-ray generator, controller and beam limiting device.

    AI/ML Overview

    The provided 510(k) summary for the GENORAY Co., Ltd. Portable X-Ray System (PORT-X II) primarily focuses on establishing substantial equivalence to a predicate device and demonstrating compliance with safety and performance standards rather than presenting a detailed study with specific acceptance criteria related to diagnostic accuracy or clinical effectiveness. Therefore, many of the requested criteria for a study proving device performance are not explicitly available in the provided text.

    Here's an analysis based on the information given, and where information is missing, it will be noted.

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, accuracy) or device performance related to image quality or clinical outcomes. Instead, it focuses on compliance with established safety and electrical standards.

    Acceptance Criteria (Derived from Compliance)Reported Device Performance
    Compliance with UL/IEC 60601-1 (General medical electrical equipment safety)Will comply (stated intention)
    Compliance with IEC 60601-2-7 (Medical electrical equipment - X-ray generators for diagnostic use)Will comply (stated intention)
    Compliance with IEC 60601-2-28 (Medical electrical equipment - Diagnostic X-ray source assemblies)Will comply (stated intention)
    Compliance with IEC 60601-2-32 (Medical electrical equipment - X-ray equipment for general use)Will comply (stated intention)
    Compliance with EN/IEC 60601-1-2(2001) (EMC for medical electrical equipment)All test results were satisfactory (by EMC Compliance Co., Ltd.)
    Leakage Current of External Body meets IEC 60601-1 requirementShould be satisfied (stated design feature)
    X-ray Shooting Controller: Dead Man Type functionShooting of X-Ray and continuous shooting should not be operated without opening Dead Man Controller (stated design feature)
    Buzzer sound during radiation shooting, off when stoppedBuzzer makes sound during shooting, turned off when stopped (stated design feature)

    2. Sample sized used for the test set and the data provenance

    The document does not describe a clinical test set for evaluating the diagnostic performance of the device or human reader performance. The testing mentioned relates to electrical safety and EMC compliance, not clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set with ground truth established by experts is described.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set with ground truth requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a portable X-ray system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a portable X-ray system, not an algorithm.

    7. The type of ground truth used

    For the safety and performance evaluations mentioned, the "ground truth" is adherence to established international standards (e.g., IEC 60601 series, EN/IEC 60601-1-2). These are engineering and electrical standards, not medical ground truth like pathology or outcomes data.

    8. The sample size for the training set

    Not applicable. This device is a hardware X-ray system, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.