510(k) Data Aggregation

    K Number
    K063006

    Validate with FDA (Live)

    Device Name
    GE LOGIQ WORKS
    Date Cleared
    2006-11-02

    (31 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To retrieve and display for reading, diagnostic review, analysis and measurement of digital medical images acquired from DICOM compliant acquisition systems such as CT, MR, PET, X-ray and Ultrasound. Also, for images acquired from GE LOGIQ ultrasound, to process, analyze and manipulate images with the same post acquisition capability as that provided on the GE LOGIQ ultrasound scanner.

    Device Description

    The GE LOGIQ Works Image Analysis and Review Workstation is optimized for ultrasound images that are acquired primarily via the GE LOGIQ family of diagnostic ultrasound systems. It is sold with or without the designated computer hardware.

    AI/ML Overview

    The GE LOGIQ Works Ultrasound Workstation is a medical image management device designed for retrieving, displaying, and processing diagnostic medical images.

    Here's an analysis of its acceptance criteria and the supporting study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceSection in Document
    System FunctionalityConformance to design specificationsDevice has been evaluated for conformance to its design specificationsSection b): 1. Non-clinical Tests
    Industry StandardsConformance to applicable industry standards for software developmentDevice has been evaluated for conformance to applicable industry standards for software developmentSection b): 1. Non-clinical Tests
    System CompatibilityVerification of system compatibility with communicating devicesDevice is further verified for system compatibility with the devices with which it communicatesSection b): 1. Non-clinical Tests
    Hardware SafetyCertification of computer hardware to applicable safety standardsComputer hardware is certified to applicable safety standardsSection b): 1. Non-clinical Tests
    Quality SystemsConformance to quality systemsThe design and development process conforms with 21 CFR 820, ISO 9001:2000 and ISO 13485:2000 quality systemsSection b): 3. Conclusion
    Medical Device SafetyConformance to applicable medical device safety standardsDevice conforms to applicable medical device safety standards and compliance is verified through internal and independent quality system auditSection b): 3. Conclusion
    Substantial EquivalenceEquivalence to cleared predicate devices regarding safety and effectivenessIt is the opinion of GE Healthcare that the GE LOGIQ is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.Section b): 3. Conclusion

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document does not mention a specific test set, sample size, or data provenance (country of origin, retrospective/prospective) for evaluating the functionality or performance of the GE LOGIQ Works Ultrasound Workstation. The submission primarily focuses on non-clinical tests related to design, standards, and compatibility.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable. No clinical test set or ground truth establishment by experts is mentioned for this device. The submission explicitly states "Clinical Tests: None required to confirm safety and effectiveness."

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical test set is described, there is no mention of an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not describe any MRMC comparative effectiveness study, nor does it discuss human reader improvement with or without AI assistance. The device is a workstation for image management and processing, not an AI-powered diagnostic tool.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Not applicable. This device is an image processing and review workstation, not an algorithm requiring standalone performance evaluation in a diagnostic context. Its performance is related to its ability to display, process, and manage images as intended, which is covered by non-clinical tests.

    7. Type of Ground Truth Used

    Not applicable. As clinical tests were not required and no diagnostic claims are being made that would necessitate comparison to a specific ground truth (like pathology or outcomes data), no "ground truth" as typically understood in diagnostic device evaluation was used. The ground for its "effectiveness" is its ability to perform its defined functions (display, process, analyze) in accordance with its specifications.

    8. Sample Size for the Training Set

    Not applicable. This device, a medical image management workstation, does not utilize machine learning or AI models that would require a "training set" in the conventional sense. Its functionality is based on established software engineering principles and image processing algorithms.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for this type of device, no ground truth for a training set was established.