    Intended Use

    The Quest Diagnostics Urine Cocaine Metabolite EIA is intended for the qualitative detection of benzoylecgonine, the primary metabolite of cocaine, in human urine on automated clinical chemistry analyzers. It is a screen test with cutoffs of 300 ng/ml and 150 ng/ml of benzoylecgonine per ml of urine. This test is intended for laboratory use only. For in vitro diagnostic use.

    The Quest Diagnostics Urine Cocaine Metabolite EIA provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. To obtain confirmed analytical results a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.

    The Quest Diagnostics Urine Cocaine Metabolite EIA Calibrators are intended for medical purposes and for use only with the Quest Diagnostics Urine Cocaine Metabolite EIA to establish points of reference that are used in the determination of values in the measurement of benzoylecgonine in urine. The Quest Diagnostics Urine Cocaine Metabolite EIA Controls are intended for use as an assay quality control matrix to monitor the precision and accuracy of the laboratory testing procedures for benzoylecgonine.

    Device Description

    Quest Diagnostics Urine Cocaine Metabolite EIA is a competitive homogenous enzyme immunoassay for the determination of cocaine metabolites in urine specimens.

    AI/ML Overview

    The provided 510(k) summary describes the Quest Diagnostics Urine Cocaine Metabolite EIA, a competitive homogenous enzyme immunoassay for the qualitative detection of benzoylecgonine in urine specimens.

    Here's an analysis of the provided information, focusing on acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for specific performance metrics in a tabular format as one might see for digital health devices. Instead, it demonstrates substantial equivalence to a predicate device by comparing various performance characteristics. The implied acceptance criterion for this type of submission is usually that the new device performs as well as or substantially equivalent to the predicate device, especially when using a reference method like GC/MS.

    Performance CharacteristicQuest Diagnostics Urine Cocaine Metabolite EIA PerformancePredicate Device (DAT II Cocaine II) Performance
    PrecisionShown to be substantially equivalent(Not explicitly detailed, but implied as a benchmark)
    Analytical SensitivityShown to be substantially equivalent(Not explicitly detailed, but implied as a benchmark)
    InterferenceShown to be substantially equivalent(Not explicitly detailed, but implied as a benchmark)
    Antibody Cross-ReactivityShown to be substantially equivalent(Not explicitly detailed, but implied as a benchmark)
    Qualitative Results (Compared to GC/MS Reference)Substantially equivalentSubstantially equivalent (implied, as the standard for comparison)

    Notes on Cutoffs: The indications for use specify cutoffs of 300 ng/ml and 150 ng/ml for benzoylecgonine. The performance data would have been evaluated against these cutoffs.

    2. Sample Size Used for the Test Set and Data Provenance:

    The summary states: "Results screened from patient specimens with both the Quest Diagnostics Urine Cocaine Metabolite EIA and the DAT II Cocaine II assay also showed that the qualitative results from this two test systems are substantially equivalent when using GC/MS results as reference."

    • Sample Size: The exact number of "patient specimens" used for the test set is not specified in the provided document.
    • Data Provenance: The document refers to "patient specimens," implying real-world human samples. However, the country of origin is not specified, nor is it explicitly stated whether the data was retrospective or prospective. Given the context of a 510(k) submission for an in vitro diagnostic, it would typically involve prospective or retrospectively collected de-identified clinical samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • The ground truth for the test set was established using Gas Chromatography/Mass Spectroscopy (GC/MS).
    • GC/MS is an analytical chemical method, not reliant on human expert interpretation in the same way as imaging or clinical diagnosis relies on physicians. Therefore, the concept of "number of experts" or "qualifications of experts" as typically applied to clinical or imaging studies is not directly applicable here. The "experts" would be the technicians or chemists operating the GC/MS equipment, ensuring its proper calibration and interpretation of analytical results, and laboratories performing GC/MS are subject to stringent quality control and accreditation.

    4. Adjudication Method for the Test Set:

    • Not Applicable. Since GC/MS is an objective analytical method providing a definitive chemical concentration, there is no need for an adjudication method among multiple human readers for establishing ground truth. The GC/MS result itself serves as the unadjudicated "ground truth."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. This is an in vitro diagnostic device for chemical analysis, not a device requiring human interpretation of medical images or clinical data. Therefore, an MRMC comparative effectiveness study, which typically evaluates improvements in human reader performance with AI assistance, is not relevant or applicable to this type of device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    • Yes, in essence. The entire study described focuses on the standalone performance of the Quest Diagnostics Urine Cocaine Metabolite EIA (the "algorithm" or device) against a reference standard (GC/MS) and a predicate device. The output of the EIA is a qualitative result (positive/negative based on cutoffs), and its performance is evaluated directly without human intervention needed to interpret the EIA's direct output for the purpose of the study. The "human-in-the-loop" component comes later in clinical practice when a physician interprets the preliminary result in the context of the patient's overall clinical picture, but the device performance itself is standalone.

    7. Type of Ground Truth Used:

    • The type of ground truth used is analytical chemical confirmation via Gas Chromatography/Mass Spectroscopy (GC/MS). The summary explicitly states: "when using GC/MS results as reference."

    8. Sample Size for the Training Set:

    • The document does not provide information regarding a separate "training set" or its sample size. This is typical for a traditional immunoassay device. While method development and optimization would have involved extensive testing, the concept of a distinct, labeled "training set" (as in machine learning) is not usually described in 510(k) summaries for these types of IVDs. The "performance characteristic studies" and testing with patient specimens serve as the validation and verification of the final assay.

    9. How the Ground Truth for the Training Set Was Established:

    • Since a distinct "training set" with established ground truth, in the context of machine learning, is not described, this question is not applicable as phrased for this device. The development of the assay would have been based on established chemical and immunological principles, and calibration (using calibrators mentioned in the Indications for Use) would be performed using materials with known concentrations, verified by methods like GC/MS.