    Intended Use

    JOYKEY Surgical Gowns (Sterile) are single use article of surgical apparel worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

    Device Description

    JOYKEY Surgical Gowns (Sterile) are non-reinforced surgical Gowns manufactured from non-woven fabric. The surgical Gown is supplied sterile and for single use only.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for JOYKEY Surgical Gowns (Sterile). This type of device is classified as medical apparel, which falls under general controls and performance standards rather than requiring a complex clinical study with AI components. Therefore, many of the requested categories in your query (e.g., sample size for test set, number of experts for ground truth, MRMC study, AI improvement effect size, training set insights) are not applicable to this submission.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria StandardReported Device Performance (Conforms To)
    ASTM F1670-03Barrier properties against blood and body fluids
    ASTM D1424Physical specifications (related to tear resistance, flexibility)
    ASTM D5034Physical specifications (related to tensile strength, elongation)
    NFPA StandardsFlammability standards
    ISO 10993 seriesBiological standards (biocompatibility)
    ISO 11137Gamma Sterilization standard

    Note: The specific performance values or thresholds for each standard are not detailed in the provided summary, only that the device conforms to them.

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. The submission refers to "bench testing" which implies laboratory tests on samples of the gown material rather than a clinical "test set" with patient data. The provenance of the test materials is not specified, but they are manufactured by JOYKEY INDUSTRYIAL LTD. in Hong Kong. The testing is retrospective in the sense that the results were generated prior to submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. Ground truth for surgical gowns is established by adherence to recognized national and international performance standards (e.g., ASTM, ISO, NFPA) through laboratory testing, not expert consensus on clinical cases.

    4. Adjudication Method for the Test Set

    Not applicable. As ground truth is based on laboratory test results against established material standards, there is no need for an adjudication method as would be used in an imaging or diagnostic study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. A MRMC study is not relevant for surgical gowns. This type of study is typically performed for AI-assisted diagnostic devices to evaluate the impact of the AI on human readers' performance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not applicable. This device is not an algorithm or AI system. It is a physical medical device (surgical gown).

    7. The Type of Ground Truth Used

    The ground truth used for proving the device meets acceptance criteria is compliance with established national and international performance standards for physical characteristics, barrier properties, flammability, biocompatibility, and sterility. This is demonstrated through objective laboratory (bench) testing.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI component in this device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI component or training set, this information is not relevant.