Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K061716

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-08-14

    (56 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The following Prestige Ameritech Surgical Masks are single use disposable devices intended to be worn in the operating room as well as dental, isolation and other medical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material.

    Prestige Ameritech Earloop Surgical Mask-Blue, Pink, Green, White, Yellow, Peach Prestige Ameritech Earloop Surgical Mask with splash shield-Blue Green, White, Peach Prestige Ameritech Earloop Surgical Mask Tissue-Blue, Pink, White, Yellow Prestige Ameritech Tie on Surgical Mask-Blue, Pink, Green, White, Peach Prestige Ameritech Tie on Surgical Mask with splash shield-Blue, Green, White, Peach

    Device Description

    The Prestige Ameritech Surgical Mask device is composed of three layers of material flat folded and pleated to form the Mask. The inner layer is composed of cellulose or spunbonded polypropylene, the middle layer is a meltblown polypropylene filter material and the outer layer is a medical grade tissue or spunbonded polypropylene. Masks with splash shields have anti-fog plastic shields attached to masks. Masks are held in place on wearer with tie strips or latex free elastic loops and contain a malleable aluminum nosepiece strip. All of the materials used in this device are typical materials commonly used in the construction of Surgical Face Masks and are being used in current legally marketed devices.

    AI/ML Overview

    The provided 510(k) summary describes the Prestige Ameritech Face Mask, a surgical face mask. The acceptance criteria and device performance are based on bench testing of physical properties rather than clinical studies with human participants. Therefore, several of the requested sections (e.g., sample size for test set, number of experts, adjudication method, MRMC study, training set information) are not applicable or cannot be extracted from this document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the results of the predicate devices, as the Prestige Ameritech device aims to demonstrate substantial equivalence by performing similarly or better.

    Performance CharacteristicsTest MethodPredicate Device K022256 Results (Implied Acceptance)Predicate Device K060776 Results (Implied Acceptance)Prestige Ameritech Device Performance
    Fluid ResistanceASTM F1862No Visual Penetration (Implied: Fluid Resistant)Fluid ResistantFluid Resistant
    Particulate FiltrationASTM F22992.0 microns (Implicit: Efficiency at 2.0 microns)96.8% at 0.1 microns98.5% at 0.1 Microns
    Bacterial FiltrationASTM F210197.9%99.9%99.6%
    Differential PressureMil M36954C1.82.342.6
    Flammability Class16CFR 1610211

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in the document. These are bench tests, so sample sizes would refer to the number of mask samples tested for each characteristic.
    • Data Provenance: The data is generated from laboratory bench testing (non-clinical tests) of the device and predicate devices. No information about the country of origin or whether it's retrospective or prospective is provided, but it's understood to be prospective testing for the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The ground truth for these performance characteristics is established through standardized laboratory test methods (e.g., ASTM F1862, ASTM F2299, ASTM F2101, MIL M36954C, 16CFR 1610). Experts are involved in designing and executing these tests, but "ground truth" isn't established through expert consensus in the typical sense for medical imaging or diagnostic devices.

    4. Adjudication method for the test set

    Not applicable. This is not a study involving human interpretation or subjective assessments that would require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a surgical mask, not an AI-powered diagnostic device, so MRMC studies and AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a surgical mask, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the device's performance is objective measurement against established industry standards and test methodologies. For example:

    • Fluid Resistance: Visual observation of penetration after exposure to synthetic blood.
    • Particulate Filtration Efficiency: Quantitative measurement of particle removal at a specific micron size.
    • Bacterial Filtration Efficiency: Quantitative measurement of bacterial removal.
    • Differential Pressure: Measurement of airflow resistance.
    • Flammability: Classification based on burn rate.

    8. The sample size for the training set

    Not applicable. This is a physical device, and the submission does not describe a training set in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable. No training set is described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1