510(k) Data Aggregation

    K Number
    K061293

    Validate with FDA (Live)

    Date Cleared
    2006-08-25

    (108 days)

    Product Code
    Regulation Number
    870.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Telaheart Digital Recorder can be used in Holter mode and Event Recorder mode.

    Holter Mode

    Detection of Arrhythmias, Efficacy of Pharmacological Treatment, and Pacemaker Evaluation.

    Event Recorder

    The Telaheart event recorder module is a patient activated device designed to record and for diagnostic evaluation of transient symptoms (such as dizziness, palpitations, syncope, and chest pain). Once data is recorded, the data is transmitted via telephone for evaluation.

    Device Description

    An ambulatory monitor, sometimes called a Holter, is a painless method to monitor the heart for a period of time (such as 24 hours, or 72 hours). The Holter is a small recording device that records the heart beat while being worn by the patient.

    The physician or technician places electrodes and wires on the patient. The wires are connected to the Holter or digital recorder.

    The Telaheart Digital Recorder has two modes that allow it to be used either as a standard Holter monitor or a looping Event recorder. The device is designed to facilitate ambulatory cardiac monitoring on the order of a physician, of those patients (including infants weighing less than 10 kg) who may herefit from such monitoring.

    The data obtained during monitoring is not analyzed at the time of recording. After the recording is complete, the data must later be downloaded to a compatible NorthEast Monitoring Holtec or Event analysis system (TelePro Software) to be analyzed. The Holter Analysis Software was cleared by FDA under K930564.

    The device is not intended to replace real time telemetry monitoring for patient suspected of having life threatening arrhythmias.

    The Telaheart Digital Recorder package includes:

    • . Telaheart Digital Recorder
    • . Operation Manual
    • SD Card .
    • Patient Cable

    The Telaheart digital recorder is powered by one 1.5 volt AA alkaline battery (MN1500 or the equivalent), one AA rechargeable NiMH (nickel metal hydride) battery, or one AA Eveready Lithium Lqurbattery. Batteries should not be re-used for a second patient. The batteries are not included, users are instructed to purchase 2 AA batteries.

    The device is compatible with standard silver / silver chloride ECG electrodes are not provided with the subject device. The user is instructed to purchase standard silver / silver chloride ECG electrodes.

    The Telaheart digital recorder uses NorthEast Monitoring Telaheart patient cables with either 2, 5, or 7 patient leads. The patient cable connects to the recorder via a connector on the recorder. A patient cable is provided with the Telaheart Digital Recorder.

    The Telaheart has an LCD that is used to display the time of day (during the recording), prompts and error messages (during the hookup procedure or during recording), or lead quality (during the hookup procedure).

    The data collected by the Telaheart digital recorder is stored on a removable SD Card.

    The Telaheart is packaged in a plastic bag in a cardboard shipping carton. The shipping carton will also include a patient cable and an SD Card.

    AI/ML Overview

    This 510(k) submission (K061293) for the Telaheart Digital Recorder does not contain a study that establishes performance against acceptance criteria in the way a typical clinical performance study would.

    This submission is for a medical device (an ambulatory ECG recorder, also known as a Holter or Event Recorder) that does not perform automated analysis. Its primary function is to record physiological data (ECG signals) for later analysis by compatible software (TelePro Software, cleared under K930564). Therefore, the "performance data" and "acceptance criteria" presented here relate to the device's functional and safety aspects, not its ability to diagnose conditions or assist human readers.

    Here's the breakdown of the information based on your request, focusing on what is available in the provided text:


    Acceptance Criteria and Reported Device Performance (Table)

    Since the device's function is data acquisition and not analysis, the "acceptance criteria" are related to its physical and electrical specifications and compliance with safety standards rather than diagnostic performance metrics.

    Acceptance Criteria CategorySpecificsReported Device Performance (or Compliance)
    Physical SpecificationsDimensions8.6cm (length) x 6.0cm (width) x 2cm (depth)
    Weight70.9 grams (2.5 oz) without battery, 99.3 grams (3.5 oz) with battery
    Electrical SpecificationsRecording Bandwidth0.05 to 70 hertz in 180 samples/sec mode
    Prefilter Sampling Rate720 samples/sec
    Data Storage Rate180 samples/sec (4 sample average), 360 samples/sec (2 sample average), or 720 samples/sec
    Pacemaker Sensitivity2 millivolts
    Pacemaker Pulse Duration150 to 2500 microseconds
    Safety and EM CompatibilityElectrical Safety StandardsIEC 60601-1, IEC 60601-1-2-47, EN 60601-1-2 (All passed)
    Software PerformanceValidation Testing (Minor Level of Concern Software)Successfully passed all validation testing

    Study Details for Performance Evaluation:

    As mentioned, this 510(k) does not describe a clinical performance study in the traditional sense (e.g., comparing diagnostic accuracy). Instead, it focuses on the device's engineering, safety, and software validation.

    2. Sample Size for Test Set and Data Provenance:
    * N/A. The document does not describe a clinical test set with patient data for diagnostic performance evaluation. The "tests" mentioned are for electrical safety, EMC, and software validation, which typically use engineering samples, test signals, and software testing methodologies rather than a clinical test set of patients.

    3. Number of Experts and Qualifications for Ground Truth:
    * N/A. There is no mention of experts establishing a ground truth for a clinical test set, as no such study is presented.

    4. Adjudication Method:
    * N/A. Not applicable, as there is no clinical test set requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
    * No. This type of study would compare human readers with and without AI assistance to assess diagnostic improvement. The Telaheart Digital Recorder is a data acquisition device, not an AI-powered diagnostic tool, and it does not perform automated analysis. Therefore, an MRMC study is not relevant to this submission.

    6. Standalone (Algorithm Only) Performance Study:
    * No. The device itself does not contain an algorithm for standalone diagnostic performance. It records raw ECG data. The separate "TelePro Software" (cleared under K930564) is responsible for analysis, but its standalone performance is not detailed in this 510(k) for the recorder.

    7. Type of Ground Truth Used:
    * N/A. For the device itself, "ground truth" would relate to accurate signal acquisition and storage (e.g., comparing recorded waveforms to known input signals for electrical performance, or verifying software functionality against specifications). It does not involve "expert consensus," "pathology," or "outcomes data" for diagnostic purposes.

    8. Sample Size for Training Set:
    * N/A. There is no mention of a training set for an AI algorithm, as this device does not utilize AI for diagnostic analysis.

    9. How Ground Truth for Training Set was Established:
    * N/A. Not applicable, as there is no training set mentioned.