Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K060846

    Validate with FDA (Live)

    Date Cleared
    2007-12-03

    (615 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To be used for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household work activities choose Manual modes 1, 2, 3, 4, 5, 6 or Auto 4

    To be used for temporary relief of pain associated with sore and aching muscles in the upper extremities (arm) due to strain from exercise or normal household work activities choose Manual modes 1, 2, 3, 4, 5, 6 or Auto 1 or Auto 3

    To be used for temporary relief of pain associated with sore and aching muscles in the lower extremities (leg) due to strain from exercise or normal household work activities choose Manual modes 1, 2, 6 or Auto1 or Auto 4

    Used to stimulate healthy muscles in order to improve and facilitate muscle performance choose Manual Mode 1 or Auto Mode 2

    Device Description

    The T1040 is a portable; battery powered (4.5 VDC) multi function device offering both Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) qualities in one device.

    Independent channel (two pads) that effectively transfers your desired choice of preprogrammed electrical pulses directly through electrode adhesive pads to the suggested area of the body where the electrodes are placed, causing minimal muscle contractions. A garment belt is used to hold the electrode pads to treat the lower back muscles. There are 10 modes of operation, 6 manual and 4 automatic.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the T1040 device, extracted from the provided 510(k) submission:

    Acceptance Criteria and Device Performance

    The submission relies on demonstrating substantial equivalence to existing predicate devices rather than setting specific numerical acceptance criteria for clinical performance. The acceptance criteria essentially revolve around the T1040's technical specifications falling within the range of, or being comparable to, the predicate devices, and its safety being established.

    Acceptance Criteria CategorySpecific Criteria (Implicit or Explicit in Submission)Reported Device Performance (T1040)
    Electrical ParametersMax. Voltage, Max. Current, Pulse Width, Pulse Period, Max. Pulse Frequency, Max. Charge per Phase, Max. Current Density, Max. Average Power DensitySee Table 14 - Comparison Table (Provided in the original text)
    Intended UseRelief of pain associated with sore/aching muscles (lower back, upper & lower extremities); Stimulation of healthy muscles to improve performance.Matches the stated intended uses of the predicate devices.
    SafetyNo new safety issues compared to predicates; demonstrated safe for OTC use; compliance with electrical safety standards.No adverse events reported with over 300,000 units sold globally since 2002; compliance with AAMI NS-4, EN 60601-1, EN 60601-1-2.
    EffectivenessEffectiveness for stated indications, generally supported by peer-reviewed literature on electrical stimulation.Effectiveness supported by peer-reviewed articles demonstrating electrical stimulation improves muscle performance and reduces pain.
    Technological CharacteristicsSubstantially equivalent features, specifications, materials."Technological characteristics, features, specifications, materials and intended uses of the T1040 are substantially equivalent to the quoted predicate devices."
    Regulatory ComplianceCompliance with applicable voluntary standards and FDA guidance for software.AAMI NS-4 1985, EN 60601-1, EN 60601-1-2, ISO 9001;2000, ISO 13485;2003; FDA Guidance for Software Contained in Medical Devices.

    Table 14 - Comparison Table (from the original text, showing reported device performance and predicate performance for comparison)

    QuantityT1040Bio-StimCompex5000Z
    Max. Voltage over 10kΩ, V154.1132126.8/103.3226
    Max. Current over 10kΩ, mA15.413.212.7/10.322.6
    Max. Voltage over 2.2kΩ, V105.190167.8/153.5218
    Max. Current over 2.2kΩ, mA47.84176.3/69.899
    Max. Voltage over 500Ω, V40.729.548208
    Max. Current over 500Ω, mA81.45996.1416
    Pulse Width, µseconds21030 - 225270100
    Pulse Period, msec4.1 - 5009 – 12.512510
    Max. Pulse Frequency, Hz245110118120
    Max. Charge per Phase over 500Ω,μC16.97.632.33.4
    Max. Current Density over 500Ω,mA/cm²2.713.933.8416.64
    Max. Average Power Density over500Ω, mW/cm²5.351.110.2

    Study Details for T1040

    The submission for T1040 primarily relies on demonstrating substantial equivalence to predicate devices through technical comparisons and a post-market safety record, rather than a specific clinical study with a defined test set, ground truth, or expert review for the T1040 itself.

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Test Set Sample Size: Not applicable in the traditional sense of a clinical trial's "test set" for performance evaluation. The submission refers to a post-market safety record as evidence of safety and effectiveness.
      • Data Provenance: The device has been sold in 20 countries since 2002. This is retrospective, real-world usage data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. No "ground truth" was established by experts for a specific test set. The safety claim is based on the absence of reported adverse events from general public use over an extended period.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. No formal adjudication of a test set occurred. The safety claim is based on passive surveillance (absence of adverse event reports).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study (MRMC, AI-assisted) is not relevant to this device submission, which is for a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) that does not involve "readers" or "AI assistance" in its operation or interpretation.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is not an algorithm, but a physical electrical stimulation device. Its performance is inherent in its electrical output and user interaction.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The primary "ground truth" for safety and effectiveness is implicit:
        • Safety: The absence of reported adverse events from over 300,000 units sold over 5 years. This acts as a real-world outcome indicator for safety.
        • Effectiveness: Reference to "a number of articles in peer-reviewed publications" demonstrating the general effectiveness of electrical stimulation for muscle performance and pain reduction. This is a scientific literature-based "ground truth" for the underlying technology, not the specific device.
    7. The sample size for the training set:

      • Not applicable. This device does not use machine learning or AI models that require a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable. As no training set was used, no ground truth for it was established.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1