    Intended Use

    The FX Detachable Coils are intended for the endovascular embolization of intracranial aneurysms. The FX Detachable Coils are also intended for the embolization of other neuro vascular abnormalities, such as, arteriovenous malformations and arteriovenous fistulae.

    Device Description

    The FX Detachable Coil System consists of a platinum embolization coil attached to a composite implant delivery pusher with a radiopaque positioning marker and a handheld mechanical FX detachment actuator which facilities the release of the coil from the delivery pusher tip. The Nexus version of the FX Detachable coil is enlaced with absorbable polymer microfilaments. The FX Detachment Actuator is sold separately.

    AI/ML Overview

    The provided text describes a 510(k) submission for the FX Detachable Coil System, focusing on its substantial equivalence to predicate devices rather than a study demonstrating performance against specific acceptance criteria for standalone device performance. The study described is primarily a set of bench tests designed to show that the new device performs similarly to or meets the same engineering specifications as the predicate devices.

    Therefore, many of the requested categories, such as sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, and details on training sets for AI algorithms, are not applicable or cannot be extracted from this document, as it does not describe a clinical trial or an AI/ML-based device performance study.

    Here's the breakdown of the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various bench tests and states that the device "Met established criteria" for each. The specific numerical acceptance criteria themselves are not provided in this summary.

    Acceptance Criteria (Bench Testing)Reported Device Performance
    Dimensional & Visual AnalysisMet established criteria
    Force TransferMet established criteria
    Ease of Delivery/Coil Frictional CharacteristicsMet established criteria
    Reliability After Fatigue & Premature DetachmentMet established criteria
    Tensile Strength of Delivery PusherMet established criteria
    Tensile Strength at Detachment ZoneMet established criteria
    Particulate Generation - Adjusted Particles / 1 mLMet established criteria
    Delivery Pusher Distal Tip BucklingMet established criteria
    Delivery Pusher Distal Stiffness ProfileMet established criteria
    Packaging IntegrityMet established criteria
    Detachment Actuator Life Cycle and ReliabilityMet established criteria
    Time and Reliability of DetachmentMet established criteria
    RadiopacityMet established criteria

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. The document details bench tests, which often involve a specific number of units tested per criterion, but these numbers are not disclosed.
    • Data Provenance: Not applicable in the context of a clinical test set. The data originates from internal engineering bench testing performed by the submitting company (Micro Therapeutics, Inc. / ev3 Neurovascular). It is not retrospective or prospective clinical data, nor does it refer to country of origin in that sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable. The "ground truth" for bench tests is typically established by engineering specifications, validated test methods, and industry standards, not by clinical experts.

    4. Adjudication method for the test set

    • Not applicable. Bench tests are typically pass/fail against predefined criteria, not subject to expert adjudication in the same way clinical image reviews would be.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device, and no MRMC study was performed or described. This submission is for a physical medical device (neurovascular embolization coil).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML device. "Standalone performance" here refers to the physical device's performance as measured by the bench tests.

    7. The type of ground truth used

    • The "ground truth" for the bench tests is based on established engineering specifications, validated test methodologies, and material standards, presumably benchmarked against the predicate devices or industry requirements.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device, and no "training set" in that context is used. The development of the device would involve engineering design and verification, not machine learning training.

    9. How the ground truth for the training set was established

    • Not applicable. No training set for an AI/ML algorithm is involved.