510(k) Data Aggregation

    K Number
    K060347

    Validate with FDA (Live)

    Date Cleared
    2006-06-05

    (115 days)

    Product Code
    Regulation Number
    872.4565
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by professionally qualified Licensed dentists, endodontists and clinicians to heat gutta percha and to place it into the previously prepared root canals of human teeth, in order to provide a quick and complete obturation of the canal.

    Device Description

    The B& L-Bcta(Model: WL-B1) is an electrically powered dental device used for heating gutta percha and placing the softened material in prepared root canals of teeth during root canal therapy. This model has a temperature memory controls. The user gently squeezes the trigger to express the desired amount of gutta percha in to the root canal through a soft silver applicator needle as the predicate, Obtura Heated Gutta Percha System(K042828)

    AI/ML Overview

    The B&L-Beta (Model: WL-B1) is an electrically powered dental device used for heating gutta percha and placing the softened material in prepared root canals of teeth during root canal therapy.

    Here's an analysis of the acceptance criteria and the study information provided:

    1. Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than defining and meeting specific analytical or clinical performance acceptance criteria in the traditional sense. The "acceptance criteria" here are implicitly that the new device is "as safe and effective as the predicate device," as stated in section 5 and 7.

    The performance comparison is based on the characteristics of the device and its intended use, aligning them with the predicate device.

    CharacteristicAcceptance Criteria (Predicate Device K042828)Reported Device Performance (B&L-Beta Model: WL-B1)Met?
    Intended UseHeat gutta percha and place it into previously prepared root canals of human teeth for quick and complete obturation.Heat gutta percha and place it into previously prepared root canals of human teeth for quick and complete obturation.Yes
    Power12V DC 1A3.7V Li-ion Battery (Difference noted, but deemed not to affect substantial equivalence)N/A
    AdapterDC 12V 1A powered from the AC lineDC 12V 850mA for charging, powered from AC line (Difference noted, but deemed not to affect substantial equivalence)N/A
    Setting Temp.140, 150, 170, 180, 200°C100, 150, 180, 200, 230°C (Different range but within typical clinical needs; deemed not to affect substantial equivalence)N/A
    Needle Size21 G, 23 G, 25G21, 23, 25GYes
    Material of NeedleAlloy of yellow copper, silver coatedAlloy of yellow copper, silver coatedYes
    Safety and EffectivenessDemonstrated through bench and clinical testing to be safe and effective for its intended use. (Implicitly matches predicate's safety/effectiveness)Bench and clinical testing indicates the new device is as safe and effective as the predicate device.Yes

    Conclusion from the Submission: The company concluded that the B&L-Beta Model: WL-B is as safe and effective as the predicate device, and the technological differences are minor, thus rendering them substantially equivalent.

    2. Sample Size and Data Provenance for Test Set

    • The document states "bench and clinical testing data" but does not specify the sample size used for the test set in either the bench or clinical testing.
    • Data Provenance: The document does not explicitly state the country of origin.
    • Retrospective or Prospective: The document does not specify whether the clinical testing was retrospective or prospective. It generally refers to "bench and clinical testing data" without further detail.

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    • The document does not provide information on the number of experts used or their qualifications for establishing ground truth for any test set. The evaluation focuses on comparison to a predicate device rather than an independent ground truth assessment by experts.

    4. Adjudication Method for the Test Set

    • The document does not describe any adjudication method as it does not detail a study involving expert readers or assessors forming a ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • The document does not indicate that an MRMC comparative effectiveness study was done. The comparison is device-to-device based on technical characteristics and a general assertion of safety and effectiveness, not a study of human readers' performance with and without AI assistance.

    6. Standalone Performance Study

    • The document does not describe a standalone (algorithm only without human-in-the-loop performance) study. This device is a physical instrument, not an AI algorithm.

    7. Type of Ground Truth Used

    • Given that this is a 510(k) for a physical dental instrument, the concept of "ground truth" in the context of diagnostic accuracy (like pathology or outcomes data) is not directly applicable.
    • The "ground truth" for demonstrating substantial equivalence relies on:
      • Technical specifications comparison to the predicate device.
      • Performance (safety and effectiveness) demonstrated through bench and clinical testing, implicitly showing it performs its intended function adequately and without undue risk.

    8. Sample Size for the Training Set

    • The document does not mention a training set sample size. This device is a physical instrument and does not involve machine learning algorithms that require training data.

    9. How Ground Truth for the Training Set was Established

    • As there is no training set for this type of device, this question is not applicable.