    Intended Use

    Iontophoretic drug delivery electrodes are indicated for the administration of soluble salts or other drugs into the body for medical purposes as an alternative to hypodernic injections. They are also indicated for iontophoretic dermal administration of IONTOCAINE® (Lidocaine HCl 2% and Epinephrine 1:100,000 Topical Solution).

    Device Description

    An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body for medical purposes. Iontophoresis technology is based on the principle that an electric potential will cause ions in solution to migrate according to their electrical charges. The quantity and distribution of a drug delivered into and across the skin by iontophoresis is dependent on the charge and molecular weight of the ion, the strength of the electrical current applied, electrode composition, duration of current flow, and numerous other factors.

    The IOMED, Inc. MW-1000 iontophoresis system consists of an active delivery electrode and a return electrode. The electrodes are designed for a single-patient, oncapplication use and should be used with IOMED's approved iontophoretic dose controller. The active delivery electrode is attached to a handle which is used to guide the device over the area to be treated. The handle is designed to be reused.

    The IOMED, Inc. MW-1000 iontophoresis system is designed to attach to IOMED's approved Phoresor . These devices provide control of the current and therefore dosage delivered.

    AI/ML Overview

    Here's an analysis of the provided text regarding the IOMED MW-1000 Iontophoretic Drug Delivery Electrode, structured to address your specific questions about acceptance criteria and supporting studies:

    This document primarily describes an Iontophoresis device and its substantial equivalence to predicate devices, focusing on regulatory approval for marketing rather than detailed performance study results that would typically include specific quantitative acceptance criteria for a new AI/medical device. The "acceptance criteria" here are largely related to safety and biocompatibility and demonstration of drug delivery capability being equivalent or superior to predicate devices, rather than a quantifiable performance metric like sensitivity or specificity.


    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    Drug Delivery EffectivenessPositive: "This testing shows that both negatively and positively charged drugs can be effectively delivered using the drug electrode to be used with the MW-1000." (Implied acceptance criterion: Device must effectively deliver both negatively and positively charged drugs.)
    Biocompatibility / Primary Dermal IrritationMild Irritant (0.5 IDI): "rated as a mild irritant (0.5) during lidocaine administration." Non-Irritant: "a non-irritant during a Dexamethasone administration from a Dexamethasone/lidocaine (1:2) mixture, and a non-irritant during Dexamethasone administration alone." (Implied acceptance criterion: Device must demonstrate acceptable levels of biocompatibility/dermal irritation, ideally non-irritant or mild irritant, consistent with predicate devices and safety standards for skin contact devices.)
    Substantial Equivalence (Regulatory)Achieved: "determined the device is substantially equivalent... to legally marketed predicate devices." (Acceptance criterion: Device must demonstrate substantial equivalence to predicate devices to obtain 510(k) clearance.)

    Detailed breakdown of study information based on your questions:

    1. A table of acceptance criteria and the reported device performance:
    Provided above. The acceptance criteria are largely implied by the context of a 510(k) submission, focusing on safety, biocompatibility, and functional equivalence to existing devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Drug Delivery Study:
      • Test Set: "freshly excised hairless mouse skin." The number of mice or skin samples is not specified.
      • Data Provenance: Not explicitly stated, but an in-vitro study.
    • Safety and Biocompatibility (Primary Dermal Irritation) Study:
      • Test Set: "rabbits." The number of rabbits used is not specified.
      • Data Provenance: Not explicitly stated, but carried out "in accordance with FDA regulations for Good Laboratory Practices." This suggests a prospective study design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Drug Delivery Study: The "ground truth" here is the actual amount of drug delivered and quantified. This would have been established through a laboratory process (radioassay) by trained lab personnel, not necessarily "experts" in the clinical sense (like radiologists). No specific number or qualifications are given for these personnel in the document.
    • Safety and Biocompatibility Study: The "ground truth" (irritation scores) would have been established by trained personnel, likely toxicologists or veterinarians, following standard protocols. The document states "Using standard Primary Dermal Irritation Index scores," implying a standardized evaluation, but does not specify the number or qualifications of the evaluators.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable / Not Specified: For in-vitro drug delivery and animal irritation studies, the concept of "adjudication" as typically applied to human clinical image or diagnostic interpretation (e.g., 2+1 expert consensus for ground truth) does not apply. Results are typically based on quantitative measurements and standardized scoring protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No: This device is an iontophoresis electrode, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to human reader performance with or without AI assistance is not relevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable: This is a hardware device for drug delivery, not an algorithm. Therefore, "standalone algorithm performance" is not a relevant concept for this product.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Drug Delivery: Quantitative measurement of radiolabeled drugs in skin and receptor solutions (radioassay).
    • Biocompatibility: Standardized Primary Dermal Irritation Index scores based on visual assessment of skin reactions in rabbits.

    8. The sample size for the training set:

    • Not Applicable: This is a hardware device, not a machine learning algorithm that requires a "training set." The study data presented is for validation/testing of the device's physical properties and function.

    9. How the ground truth for the training set was established:

    • Not Applicable: As there is no training set, this question is not applicable.

    Summary Takeaway:

    The provided document is a 510(k) summary for a medical device (iontophoresis electrode). It focuses on demonstrating substantial equivalence to predicate devices through in-vitro and animal studies rather than presenting quantifiable clinical performance metrics typically associated with AI or diagnostic devices. The "acceptance criteria" revolve around regulatory compliance, demonstrable drug delivery function, and acceptable safety/biocompatibility profiles as measured by established laboratory methods.