Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K060092

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-03-09

    (56 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuickCat™ Extraction Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system.

    Device Description

    The QuickCat™ Extraction Catheter is a single use, disposable, dual lumen catheter with associated accessories consisting of a 30 ml. vacuum syringe, extension tubing with stopcock, and an independent 40-micron filter basket. The 145 cm working length and is compatible with 6F guiding catheters with an inner diameter (I.D.) ≥ 0.068" (1.73 mm) and 0.014" (0.36 mm) diameter guidewires. The extraction lumen of the catheter facilitates removal of emboli and thrombi via the attached tubing assembly, stopcock and vacuum syringe. The atraumatic distal tip, which incorporates a radiopaque marker for visibility under fluoroscopy, provides for smooth passage in the arterial system. The catheter consists of three segments. The stiffer proximal seqment and more flexible distal segments provide the required structural integrity and flexibility to navigate tortuous vasculature. The distal segment consists of a dual lumen to allow for "rapid exchange" attachment to the quidewire. The catheter's distal portion has a hydrophilic coating to enhance deliverability. A 40-micron pore filter basket is supplied to assist in filtering of blood and thrombotic material for visual or laboratory analysis.

    AI/ML Overview

    This document, a 510(k) summary for the QuickCat™ Extraction Catheter, details the device's characteristics and its substantial equivalence to predicate devices, but it does not provide specific acceptance criteria or an explicit study proving performance against those criteria in the way a clinical trial or performance study report would. It focuses on establishing equivalence for regulatory clearance.

    Here's a breakdown of the requested information based on the provided text, highlighting what is present and what is absent:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is NOT explicitly provided in the 510(k) summary. The document states that "Design verification testing was performed on the QuickCat™ Extraction Catheter to confirm that the design inputs meet the design outputs," and "Comparison bench testing was performed... regarding performance characteristics to demonstrate equivalency." However, it does not list specific quantitative acceptance criteria (e.g., "Device must extract X% of emboli") or the specific performance results against such criteria. It only asserts that equivalency was demonstrated.

    2. Sample Size for the Test Set and Data Provenance

    This information is NOT explicitly provided. The summary mentions "Performance Testing - Bench" and "Comparison bench testing," but does not specify the sample sizes (e.g., number of catheters tested, number of simulated emboli used) or the data provenance (e.g., country of origin, retrospective/prospective). This type of detail would typically be in a separate, more comprehensive test report, not usually in a 510(k) summary.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is NOT applicable/provided. The testing mentioned (biocompatibility and bench testing) does not involve human expert assessment for establishing "ground truth" in the way a clinical study using diagnostic imaging would. Ground truth in this context would likely refer to objective measurements of device performance (e.g., extraction force, flow rate, navigability).

    4. Adjudication Method for the Test Set

    This information is NOT applicable/provided. As the document describes bench testing and biocompatibility, there is no expert adjudication method like 2+1 or 3+1 typically used in clinical imaging studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a MRMC comparative effectiveness study was NOT done/reported in this document. The document discusses "Comparison bench testing" to demonstrate equivalency to predicate devices, but this refers to comparing physical performance characteristics, not comparing human reader performance with and without AI assistance. This device is a physical medical device (catheter), not an AI-powered diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    No, a standalone performance study was NOT done/reported in this document. This is a physical medical device, not an algorithm. Therefore, "algorithm only" performance is not relevant.

    7. Type of Ground Truth Used for the Test Set

    For the bench testing, the "ground truth" would likely be objective measurements and engineering specifications of the catheter's performance characteristics (e.g., successful extraction of simulated thrombi, navigability, guidewire compatibility, lumen integrity, material properties). For biocompatibility, the ground truth is established through adherence to ISO 10993 standards and FDA guidelines, indicating the absence of adverse biological reactions.

    8. Sample Size for the Training Set

    This information is NOT applicable/provided. As this is a physical medical device, not an AI or machine learning model, there is no "training set" in the computational sense. The design and validation of the catheter would involve engineering principles, material science, and physical testing, not data training.

    9. How the Ground Truth for the Training Set Was Established

    This information is NOT applicable/provided. No training set exists for this type of device. The "ground truth" for the device's design and manufacturing would be established through engineering design specifications, material standards, and quality control processes.


    In summary:

    The provided 510(k) summary for the QuickCat™ Extraction Catheter is for a physical medical device and primarily focuses on demonstrating "substantial equivalence" to predicate devices through biocompatibility and bench testing. It clearly lacks the specific details about acceptance criteria, quantifiable performance results against those criteria, sample sizes for test sets, and the methodologies related to expert review or AI model validation that would be found in a performance study for a diagnostic algorithm or a clinical trial report. The context of the questions (e.g., "number of experts," "MRMC study," "training set," "algorithm only performance") strongly suggests inquiry into an AI/ML powered device, which this catheter is not.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1